CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System
Enforcement
- Recall Number
- Z-1099-2026
- Event ID
- 97784
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Thoratec LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 21, 2026
- Initiation Date
- October 10, 2025
- Classification Date
- January 15, 2026
- Address
- 6035 Stoneridge Dr, Pleasanton, CA, 94588-3270, United States
Description
CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System
Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.
UDI-DI: 07640135140627, 07640135140603, 05415067036414 IFU: Part(Name) Revision: For US: 1. ARTEN600254920 (IFU CentriMag Blood Pump ABT) Rev C 2. ARTEN600254958 (IFU PediMag Blood Pump US ABT) Rev C 3. ARTEN100174794 (CentriMag Adult Pre-connected Pack IFU, US) Rev B 4. ARTEN600230157 (IFU CentriMag Blood Pump ECMO US) Rev D FOR OUS: ARTEN600254935 (IFU CentriMag Blood Pump CAN ABT) Rev C ARTEN600254936 (IFU PediVas Blood Pump Canada ABT) Rev C ARTEN600332205 (IFU CMag VAS-ECMO Pump OUS) Rev A ARTEN600332207 (IFU PVAS VAS-ECMO Pump OUS) Rev A REF 201-20003, 104318, 201-51006, 201-51007, 201-90001 201-90001T don't have associated un-expired batches/serial numbers. REF/Batch(Serial): 102953/10007334, 10007335, 10007336, 10012012, 10012013, 10012014, 10012015, 10012016, 10036179, 10036180, 10036181, 10037990, 10037992, 10037993, 10037995, 10037996, 10046028, 10046029, 10046030, 10046031, 10046032, 10046033, 10058754, 10058755, 10058756, 10061673, 10061675, 10061676, 10061677, 10065946, 10065947, 10065954, 10065955, 10065956, 10071358, 10071359, 10071361, 10073780, 10073781, 10073783, 10082053, 10082055, 10103208, 10103209, 10106345, 10106346, 10106347, 10106348, 10120074, 10120075, 10120076, 10120077, 10122157, 10122158, 10122159, 10125562, 10125563, 10125564, 10125565, 10138336, 10138337, 10138338, 10138339, 10145124, 10145125, 10145126, 10145127, 10164040, 10164041, 10164044, 10168193, 10168194, 10168195, 10168196, 10168197, 10168199, 10168200, 10168201, 10168202, 10168203, 10173214, 10173215, 10206751, 10206752, 10206754, 10211194, 10211195, 10211196, 10211198, 10211199, 10214967, 10214968, 10214969, 10214970, 10214971, 10235009, 10235014, 10235019, 10235024, 10245539, 10245540, 10245541, 10245542, 10245543, 10256365, 10256366, 10289217, 10289218, 10291590, 10291591, 10301036, 10301037, 10301038, 10301039, 10301040, 10301041, 10301042, 10301043, 10301044, 10303473, 10303476, 10303477, 10303478, 10325561, 10341677, 10342238 10343635, 10343636, 10343637, 10343638, 10343639, 10343640, 10351700, 10351701, 10351702, 10351703, 10351704, 10351707, 10359524, 10359525, 10359526, 10359527, 10359528, 10359529, 10359530, 10359531, 10359532, 10371222, 10371223, 10371224, 10371225, 10371226, 10373595, 10373596, 10373597, 10373598, 10373599, 10391563, 10391564, 10391565, 10391566, 10398007, 10398008, 10398009, 10398010, 10398011, 10399853, 10399854, 10399855, 10399856, 10399857, 10399858, 10399859, 10399860, 10399861, 10399862, 10399866, 10399867, 10410026, 10410027, 10410028, 10410197, 10410198, 10410199, 10411641, 10411642, 10414322, 10414323, 10414324, 10414325, 10414326, 10416065, 10416066, 10416067, 10418871, 10419212, 10424524, 10428901, 10428902, 10428904, 10428905, 10428906, 10456505, 10456506, 10456507, 10456508, 10456509, 10457499, 10457500, 10457502, 10464083, 10464085, 10464086, 10464092, 10498116, 10498117, 10498118, 10517215, 10518978, 10518979, 10518980, 10518981, 10518982, 10528585, 10528587, 10528588, 10530704, 10530705, 10530706, 10535290, 10535294, 10535297, 10535299, 10535300, 10535302, 10541032, 10541033, 10541034, 10541035, 10543873, 10543874, 10550324, 10550325, 10550326, 10550328, 10550329, 10560628, 10560629, 10560630, 10567320, 10567321, 10567322, 10567323, 10571084, 10571085, 10571086, 10571087, 10571088, 10571090, 10588824, 10588825, 10588826, 10591390, 10591391, 10591392, 10591393, 10602157, 10602160, 10602161, 10602162, 10602163, 10602164, 10623578, 10623579, 10623581, 10623582, 10623583, 10632619, 10632625, 10632630, 10639955, 10639956, 10639957, 10639959, 10639960, 10646446, 10646447, 10646449, 10646450, 10646451, 10646452, 10650250, 10650251, 10652470, 10652471, 10652472, 10652473, 10652474, 10652475, 10652477, 10652479, 10659763, 10659764, 10663359, 10663361, 10663363, 10670118, 10670122, 10670123, 10670124, 10671920, 10671921, 10671922, 10671923, 10671925, 10680535, 10680536, 10680538, 10680539, 10680540, 10683369, 10683370, 10683371, 10683372, 10688155, 10688156, 10688157, 10688158, 10688159, 10688160, 10694256, 10694258, 10694260, 10694262, 10694264, 10694266, 10694268, 10694270, 10694272, 10694273, 10694274, 10694275, 10694296, 10712284, 10712285, 10716549, 10716551, 10716552, 10716553, 10716554, 10727906, 10727907, 10727908, 10727909, 10727910, 10727911, 10727912, 10748553, 10748554, 10748556, 10748557, 10771409, 10771411, 10771412, 10773388, 10773389, 10773390, 10773391, 10773392, 10773393, 10773394, 10776304, 10776308, 10776310, 10776316, 10776317, 10783737, 10783738, 10783739, 10783740, 10783741, 10788706, 10788707, 10788708, 10788709, 10788710, 10788711, 10816789, 10816790, 10816791, 10817064, 10817065, 10821308, 10821309, 10821310, 10828333, 10828334, 10828335, 10828336, 10828337, 10832541, 10832543, 10832544, 10832545, 10832547, 10832548, 10834958, 10834959, 10834960, 10834961, 10834962, 10834963, 10834964, 10834965, 10834966, 10834967, 10834968, 10834969, 10875109, 10875110, 10875111, 10875112, 10875113, 10877738, 10877739, 10877740, 10877742, 10877743, 10877744, 10879671, 10879672, 10879673, 10879674, 10879676, 10879677, 10879678, 10879679, 10879680, 10879682, 10879683, 10879684, 10879685, 10879686, 10879687, 10896069, 10896070, 10896071, 10911663, 10911665, 10911667, 10911669, 10911671, 10911675, 10941952, 10941953, 10941954, 10941955, 10941956, 10947110, 10947111, 10947112, 10947113, 10947114, 10947603, 10947604, 10947605, 10947607, 10947608, 10952744, 10952746, 10952747, 10952749, 10952750, 10956143, 10956144, 10956145, 10960853, 10985404, 10985408, 10985409, 10985410, 10989308, 10989309, 10989311, 11013097, 11013099, 11013100, 11013101, 9103442, 9143416, 9143417, 9174168, 9237175, 9237176, 9237177; 201-90010/10012019, 10012020, 10012021, 10012022, 10012023, 10012024, 10012025, 10012026, 10026482, 10026483, 10026484, 10026485, 10026486, 10026487, 10026488, 10026489, 10061666, 10061668, 10061669, 10061671, 10075586, 10075587, 10075588, 10082037, 10082039, 10082040, 10082041, 10106341, 10106342, 10106343, 10106344, 10114497, 10114499, 10114500, 10118096, 10118097, 10118098, 10118099, 10118100, 10118101, 10118102, 10118103, 10122160, 10122161, 10122162, 10122163, 10157219, 10157220, 10157221, 10184547, 10184548, 10184549, 10184551, 10184553, 10184554, 10184555, 10190508, 10190509, 10194010, 10214972, 10214973, 10234422, 10234423, 10234424, 10235052, 10250573, 10250574, 10250576, 10250577, 10256367, 10256368, 10256369, 10256370, 10256371, 10256372, 10256373, 10261919, 10261920, 10261921, 10261923, 10279769, 10333758, 10333759, 10333760, 10339372, 10339373, 10339374, 10343663, 10343664, 10343665, 10351697, 10351698, 10351699, 10379489, 10379490, 10379491, 10383826, 10383827, 10383828, 10383829, 10391724, 10391725, 10391726, 10391727, 10391730, 10391732, 10391733, 10413131, 10413132, 10413133, 10413135, 10413561, 10413565, 10413566, 10418562, 10462022, 10462023, 10462024, 10470479, 10470521, 10470524, 10472542, 10472550, 10474373, 10474374, 10474376, 10474377, 10498175, 10498176, 10501172, 10501173, 10503762, 10528582, 10543891, 10543892, 10552629, 10552630, 10552631, 10560591, 10560620, 10560621, 10560622, 10560623, 10560624, 10567338, 10567342, 10567343, 10567344, 10571082, 10571083, 10586726, 10586727, 10586728, 10591394, 10591395, 10591396, 10598036, 10598037, 10598038, 10598039, 10598040, 10602152, 10602153, 10602154, 10602155, 10602156, 10619564, 10619565, 10619566, 10619567, 10619568, 10621605, 10621606, 10621607, 10621608, 10621611, 10632450, 10632451, 10636575, 10640045, 10640050, 10640051, 10650671, 10650677, 10650680, 10650681, 10659755, 10659756, 10659757, 10659758, 10659759, 10659760, 10659761, 10663352, 10663353, 10663354, 10727899, 10727900, 10727901, 10727902, 10727903, 10727904, 10727905, 10735527, 10735528, 10735530, 10738501, 10739056, 10739065, 10739066, 10739067, 10739068, 10739069, 10739070, 10739071, 10739072, 10748571, 10748573, 10748574, 10778162, 10778164, 10778165, 10797739, 10797742, 10797746, 10815359, 10815362, 10815363, 10815364, 10816785, 10816786, 10816787, 10821305, 10821306, 10821307, 10828329, 10828330, 10828331, 10828332, 10828529, 10832540, 10835410, 10847295, 10847297, 10847298, 10847299, 10847300, 10847301, 10847302, 10847303, 10858660, 10858661, 10859392, 10859400, 10859403, 10859406, 10859407, 10872106, 10872108, 10872109, 10872111, 10872113, 10875119, 10875123, 10875126, 10875128, 10877758, 10877759, 10877760, 10893762, 10893763, 10893764, 10893765, 10893768, 10893769, 11001344, 11001345, 11001347, 9088894, 9088895, 9088897, 9088898, 9088899, 9088900, 9188312, 9230175, 9239396, 9239397, 9239398, 9239879, 9239880, 9239883, 9251990, 9251991, 9251992, 9251993; 201-90016/10104959, 10109004, 10148485, 10148517, 10195550, 10395893, 10477785, 10528583, 10560573, 10579395, 10579396, 10630119, 10665672, 10916780, 10916781, 10969131, 9163847, 9239394; CMAEK00/CMAEK00(20236575, 20236576, 20236577, 20236578, 20236579, 20236580, 20236581, 20236582, 20236583, 20236584, 20236585, 20236586, 20236587, 20236588, 20236589, 20236590, 20236591, 20236592), 10438711(20236593, 20236594, 20236595, 20236596, 20236597, 20236598, 20236599, 20236600, 20236601, 20236602, 20236603, 20236604, 20236605, 20236607, 20236608, 20236609, 20236610), 10451609(20236611, 20236612, 20236613, 20236614, 20236615, 20236616, 20236617, 20236619, 20236620, 20236621, 20236622, 20236623, 20236624, 20236625, 20236626, 20236627, 20236628), 10451610(20256947, 20256948, 20256949, 20256950, 20256951, 20256952, 20256953, 20256955, 20256956, 20256957, 20256958, 20256959, 20256960, 20256961, 20256962, 20256963, 20256964), 10451615(20256965, 20256966, 20256967, 20256968, 20256969, 20256970, 20256971, 20256972, 20256973, 20256974, 20256975, 20256976, 20256977, 20256978, 20256979, 20256980, 20256982), 10451617(20256983, 20256984, 20256985, 20256986, 20256987, 20256988, 20256989, 20256990, 20256991, 20256992, 20256993, 20256994, 20256995, 20256996, 20256997, 20256998, 20256999, 20257000), 10463759(20278309, 20278310, 20278311, 20278312, 20278313, 20278314, 20278315, 20278316, 20278317, 20278318, 20278319, 20278320, 20278321, 20278322, 20278323, 20278324, 20278325, 20278326), 10463760(20278327, 20278328, 20278329, 20278330, 20278331, 20278332, 20278333, 20278334, 20278335, 20278336, 20278337, 20278338, 20278339, 20278340, 20278341, 20278342, 20278343, 20278344), 10552471(20278363 20278365, 20278366, 20278367, 20278368, 20278369, 20278370, 20278371, 20278372, 20278373, 20278374, 20278375, 20278376, 20278377, 20278378, 20278379, 20278380), 10552472(20278345, 20278346, 20278348, 20278349, 20278350, 20278351, 20278352, 20278353, 20278354, 20278355, 20278356, 20278357, 20278358, 20278359, 20278360, 20278361, 20278362), 10552475(20278381, 20278382, 20278383, 20278384, 20278385, 20278386, 20278387, 20278388, 20278389, 20278390, 20278391, 20278392, 20278394, 20278395, 20278396, 20278397, 20278398), 10552484(20284132, 20284133, 20284134, 20284136, 20284137, 20284138, 20284139, 20284141, 20284142, 20284143, 20284144, 20284145, 20284146, 20284147, 20284148, 20284149), 10552486(20284114, 20284115, 20284116, 20284117, 20284118, 20284119, 20284120, 20284121, 20284122, 20284123, 20284124, 20284125, 20284126, 20284127, 20284128, 20284129, 20284130, 20284131), 10552487(20342251, 20342252, 20342253, 20342254, 20342255, 20342256, 20342257, 20342258, 20342259, 20342260, 20342261, 20342262, 20342263, 20342264, 20342265, 20342266, 20342267, 20342268), 10574868(20302808, 20302810, 20302811, 20302812, 20302813, 20302814, 20302815, 20302816, 20302817, 20302818, 20302819, 20302820, 20302821, 20302822, 20302823, 20302824, 20302825), 10574869(20302714, 20302715, 20302716, 20302717, 20302718, 20302719, 20302720, 20302721, 20302722, 20302723, 20302724, 20302725, 20302726, 20302727, 20302728, 20302729, 20302730, 20302731), 10574872(20302732, 20302733, 20302734, 20302735, 20302736, 20302737, 20302738, 20302739, 20302740, 20302741, 20302742, 20302743, 20302744, 20302745, 20302746, 20302747, 20302748, 20302749), 10574873(20302750, 20302751, 20302752, 20302753, 20302754, 20302755, 20302756, 20302757, 20302758, 20302759, 20302760, 20302761, 20302762, 20302763, 20302764, 20302765, 20302766), 10574875(20302790, 20302791, 20302792, 20302793, 20302794, 20302795, 20302796, 20302797, 20302798, 20302799, 20302800, 20302801, 20302802, 20302804, 20302805, 20302806, 20302807), 10574876(20302768, 20302770, 20302771, 20302772, 20302773, 20302774, 20302775, 20302776, 20302777, 20302778, 20302779, 20302780, 20302781, 20302783, 20302784, 20302785), 10623548(20356887, 20356888, 20356889, 20356890, 20356891, 20356892, 20356893, 20356894, 20356895, 20356896, 20356897, 20356898, 20356899, 20356900, 20356901, 20356902, 20356903, 20356904), 10623549(20356959, 20356960, 20356961, 20356962, 20356963, 20356964, 20356965, 20356966, 20356967, 20356969, 20356971, 20356973, 20356974, 20356975), 10623550(20361617, 20361618, 20361619, 20361620, 20361621, 20361622, 20361623, 20361624, 20361625, 20361626, 20361627, 20361628, 20361629, 20361630, 20361631, 20361632, 20361633, 20361634), 10623551(20361635, 20361636, 20361637, 20361638, 20361639, 20361640, 20361641, 20361642, 20361643, 20361644, 20361645, 20361646, 20361647, 20361648, 20361649, 20361650, 20361651, 20361652), 10623552(20356941, 20356942, 20356943, 20356944, 20356945, 20356946, 20356947, 20356948, 20356949, 20356950, 20356951, 20356952, 20356953, 20356954, 20356955, 20356956, 20356957), 10623553(20356923, 20356924, 20356925, 20356926, 20356927, 20356928, 20356929, 20356930, 20356931, 20356932, 20356933, 20356934, 20356935, 20356936, 20356937, 20356938, 20356940), 10623554(20361582, 20361583, 20361584, 20361585, 20361586, 20361587, 20361588, 20361589, 20361590, 20361591, 20361592, 20361593, 20361594, 20361595, 20361596, 20361597, 20361598), 10649991(20356905, 20356906, 20356907, 20356908, 20356909, 20356910, 20356911, 20356912, 20356913, 20356914, 20356915, 20356916, 20356917, 20356918, 20356919, 20356920, 20356922), 10649992(20361563, 20361564, 20361565, 20361566, 20361567, 20361569, 20361570, 20361571, 20361572, 20361573, 20361574, 20361575, 20361576, 20361577, 20361578, 20361579, 20361580), 10649993(20361599, 20361600, 20361601, 20361602, 20361603, 20361604, 20361605, 20361607, 20361611, 20361612, 20361613, 20361614), 10649994(20375956, 20375957, 20375958, 20375959, 20375960, 20375961, 20375962, 20375963, 20375964, 20375965, 20375966, 20375968, 20375970, 20375971, 20375972, 20375973), 10649995(20375974, 20375975, 20375976, 20375977, 20375978, 20375980, 20375981, 20375982, 20375983, 20375984, 20375985, 20375986, 20375987, 20375988, 20375989, 20375990, 20375991), 10671740(20409885, 20409886, 20409887, 20409888, 20409889, 20409890, 20409891, 20409893, 20409894, 20409895, 20409896, 20409897, 20409898, 20409899, 20409900, 20409901, 20409902), 10671741(20409903, 20409904, 20409905, 20409906, 20409907, 20409908, 20409910, 20409911, 20409912, 20409913, 20409914, 20409915, 20409916, 20409918, 20409919, 20409920), 10671743(20409921, 20409922, 20409923, 20409924, 20409925, 20409926, 20409928, 20409929, 20409930, 20409931, 20409932, 20409933, 20409934, 20409935, 20409936, 20409938), 10671745(20409939, 20409940, 20409941, 20409942, 20409943, 20409944, 20409945, 20409946, 20409947, 20409948, 20409949, 20409950, 20409952, 20409953, 20409954, 20409955, 20409956), 10715960(20409957, 20409958, 20409959, 20409960, 20409961, 20409962, 20409963, 20409964, 20409965, 20409966, 20409967, 20409968, 20409970, 20409971, 20409972, 20409973, 20409974), 10715961(20412892, 20412893, 20412894, 20412895, 20412896, 20412897, 20412898, 20412899, 20412900, 20412901, 20412902, 20412903, 20412904, 20412905, 20412906, 20412907, 20412908, 20412909), 10715962(20412910, 20412911, 20412912, 20412913, 20412914, 20412915, 20412916, 20412917, 20412918, 20412919, 20412920, 20412921, 20412922, 20412923, 20412924, 20412925, 20412926, 20412927), 10779655(20412928, 20412929, 20412930, 20412931, 20412932, 20412933, 20412934, 20412935, 20412936, 20412937, 20412938, 20412939, 20412940, 20412941, 20412942, 20412943, 20412944, 20412945), 10779656(20412946, 20412947, 20412948, 20412950, 20412952, 20412953, 20412954, 20412956, 20412957, 20412958, 20412959, 20412960, 20412961, 20412962, 20412963), 10794262(20430144, 20430145, 20430146, 20430147, 20430148, 20430149, 20430150, 20430151, 20430152, 20430153, 20430154, 20430155, 20430156, 20430157, 20430158, 20430159, 20430161), 10794264(20430180, 20430182, 20430183, 20430184, 20430185, 20430186, 20430187, 20430188, 20430189, 20430190, 20430191, 20430192, 20430193, 20430194, 20430195, 20430196, 20430197), 10794266(20430198, 20430200, 20430201, 20430202, 20430203, 20430204, 20430205, 20430206, 20430207, 20430208, 20430209, 20430210, 20430211, 20430212, 20430213, 20430214, 20430215), 10794268(,20430162, 20430163, 20430164, 20430165, 20430166, 20430168, 20430169, 20430170, 20430171, 20430172, 20430173, 20430174, 20430175, 20430176, 20430177, 20430179), 10794269(20432844, 20432845, 20432846, 20432847, 20432848, 20432849, 20432850, 20432851, 20432853, 20432854, 20432855, 20432856, 20432857, 20432858, 20432859, 20432860, 20432861), 10805521(20430222, 20430223, 20430224, 20430225, 20430226, 20430227, 20430228, 20430229, 20430230, 20430231, 20430232, 20430233, 20430234, 20430235, 20430236, 20430237, 20430238, 20430239), 10805522(20430244, 20430245, 20430246, 20430247, 20430248, 20430249, 20430250, 20430251, 20430252, 20430253, 20430254, 20430255, 20430256, 20430257, 20430258, 20430259, 20430260, 20430261), 10805523(20430262, 20430263, 20430264, 20430265, 20430266, 20430267, 20430268, 20430269, 20430270, 20430271, 20430272, 20430273, 20430274, 20430275, 20430276, 20430277, 20430278, 20430279), 10805524(20457844, 20457845, 20457846, 20457847, 20457848, 20457849, 20457850, 20457851, 20457852, 20457853, 20457854, 20457855, 20457856, 20457857, 20457858, 20457859, 20457860), 10805525(20457862, 20457864, 20457865, 20457867, 20457868, 20457869, 20457870, 20457871, 20457872, 20457873, 20457874, 20457875, 20457876, 20457877, 20457878, 20457879), 10834379(20457880, 20457881, 20457882, 20457883, 20457884, 20457885, 20457886, 20457887, 20457888, 20457889, 20457890, 20457891, 20457892, 20457893, 20457894, 20457895, 20457896, 20457897), 10834385(20481824, 20481826, 20481829, 20481832, 20481834, 20481835, 20481839, 20481841), 10834397(20481807, 20481809, 20481812, 20481814, 20481815, 20481821, 20481822, 20481823); CMAEK01/10485765(20284196, 20284197, 20284198, 20284199, 20284200, 20284201, 20284202, 20284203, 20284204, 20284205, 20284206, 20284207, 20284208, 20284209, 20284210, 20284211, 20284212, 20284213), 10485769(20284214, 20284216, 20284217, 20284218, 20284219, 20284220, 20284221, 20284222, 20284223, 20284224, 20284225, 20284226, 20284227, 20284228, 20284229, 20284230, 20284231), 10506836(20261873, 20261874, 20261875, 20261876, 20261877, 20261878, 20261879, 20261880, 20261881, 20261882, 20261883, 20261884, 20261885, 20261886, 20261887, 20261889, 20261890), 10506837(20261891, 20261892, 20261893, 20261894, 20261895, 20261896, 20261897, 20261898, 20261899, 20261900, 20261901, 20261902, 20261903, 20261904, 20261905, 20261906, 20261907, 20261908), 10506839(20261945, 20261946, 20261947, 20261949, 20261950, 20261951, 20261952, 20261953, 20261955, 20261957, 20261958, 20261959, 20261962), 10506840(20261909, 20261911, 20261912, 20261913, 20261914, 20261918, 20261919, 20261920, 20261921, 20261922, 20261924, 20261925, 20261926), 10506841(20261927, 20261928, 20261929, 20261930, 20261931, 20261932, 20261933, 20261934, 20261935, 20261936, 20261937, 20261938, 20261940, 20261943), 10547650(20284269, 20284277), 10547655(20284294, 20284296, 20284297, 20284298, 20284299, 20284300, 20284302, 20284303, 20284304, 20284306, 20284308, 20284310, 20284311), 10605980(20361662, 20361663, 20361664, 20361665, 20361668, 20361669, 20361670), 10605982(20375884, 20375885, 20375887, 20375888, 20375889, 20375890, 20375891, 20375892, 20375895, 20375896, 20375898, 20375899, 20375900, 20375901), 10605987(20375938, 20375939, 20375940, 20375941, 20375942, 20375943, 20375944, 20375945, 20375946, 20375947, 20375948, 20375949, 20375950, 20375951, 20375952, 20375953, 20375954), 10605988(20375902, 20375903, 20375904, 20375905, 20375906, 20375907, 20375908, 20375909, 20375910, 20375911, 20375912, 20375913, 20375914, 20375915, 20375917, 20375918, 20375919), 10605989(20375921, 20375922, 20375924, 20375925, 20375926, 20375927, 20375929, 20375930, 20375931, 20375932, 20375933, 20375934, 20375935, 20375936, 20375937), 10671735(20422760, 20422761, 20422762, 20422763, 20422764, 20422765, 20422766, 20422767, 20422768, 20422769, 20422770, 20422771, 20422772, 20422773, 20422774, 20422775, 20422776), 10671737(20422777, 20422778, 20422779, 20422780, 20422781, 20422782, 20422783, 20422784, 20422785, 20422786, 20422787, 20422789, 20422790, 20422791, 20422792, 20422794), 10671738(20422795, 20422796, 20422797, 20422798, 20422799, 20422800, 20422801, 20422802, 20422803, 20422805, 20422807, 20422808, 20422809, 20422810, 20422811, 20422812), 10715950(20422705, 20422706, 20422707, 20422708, 20422709, 20422710, 20422711, 20422713, 20422715, 20422716, 20422717, 20422718, 20422719, 20422720, 20422721, 20422722), 10715951(20422725, 20422726, 20422728, 20422733, 20422734, 20422735, 20422736, 20422739, 20422740), 10715951(20422725, 20422726, 20422728, 20422733, 20422734, 20422735, 20422736, 20422739, 20422740), 10715952(20422742, 20422743, 20422745, 20422746, 20422747, 20422748, 20422754, 20422755, 20422756, 20422757, 20422758), 10715956(20422814, 20422816, 20422818, 20422819, 20422821, 20422822, 20422823, 20422824, 20422825, 20422826), 10715957(20443495, 20443496, 20443497, 20443498, 20443499, 20443500, 20443501, 20443502, 20443503, 20443504, 20443505, 20443506, 20443507, 20443508, 20443512), 10715958(20443513, 20443514, 20443516, 20443517, 20443518, 20443522, 20443524, 20443525, 20443527, 20443528, 20443530), 10757855(20443533, 20443535, 20443536, 20443537, 20443538, 20443539, 20443540, 20443545, 20443546, 20443547, 20443548), 10757856(20443549, 20443551, 20443552, 20443553, 20443555, 20443556, 20443557, 20443558, 20443560, 20443562, 20443563, 20443564, 20443565, 20443566), 10757858(20446755, 20446761), 10757859(20446726, 20446727, 20446728, 20446729, 20446730, 20446731, 20446732, 20446733, 20446734, 20446735, 20446736, 20446737, 20446738, 20446739, 20446740, 20446741, 20446742, 20446743), 10794274(20457826, 20457827, 20457828, 20457829, 20457830, 20457831, 20457835), 10805518(20481842, 20481846, 20481849, 20481850, 20481851, 20481852, 20481853, 20481854, 20481856, 20481857), 10805519(20481869, 20481871, 20481874, 20481876), 10844705(20513099, 20513100, 20513102, 20513104, 20513105, 20513107, 20513110, 20513111, 20513115), 10844706(20513118, 20513122, 20513123, 20513124, 20513131, 20513133), 10944546(20570873, 20570874, 20570881, 20570882, 20570883, 20570884, 20570885, 20570886, 20570887, 20570888)
Worldwide - US Nationwide distribution including in the states of MN, PA, OK, FL, CA, OR, IL, NY, DE, GA, NJ, NC, WI, OH, AZ, AR, TX, TN, MO, MA, MT, KY, NE, WV, VA, ME, ND, AL, CO, DC, MS, IN, MI, CT, SC, WA, NM, MD, HI, ID, IA, LA, UT, RI, NV and the countries of Algeria, Australia, Austria, Bahrain, CANADA, Cayman Islands, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Hong Kong, Italy, Jordan, Kuwait, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, Oman, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
8399