Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric) to the CRLF2 gene
Recall: Z-1099-2021
·
Reported February 17, 2021
Enforcement
- Recall Number
- Z-1099-2021
- Event ID
- 87090
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cytocell Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- February 17, 2021
- Initiation Date
- December 22, 2020
- Classification Date
- February 8, 2021
- Termination Date
- June 7, 2022
- Address
- 418 Cambridge Science Park, Milton Road, Oxford Gene Technology, Cambridge, N/A, N/A, United Kingdom
Description
Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric) to the CRLF2 gene
Reason
Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.
Code Info
Lots 072240, 071390, and 072693
Distribution
Domestic distribution to California, Illinois, Michigan, Minnesota, New Jersey, Texas, and Washington. Foreign distribution to Canada.
Quantity
72