FDA Enforcement Class II Terminated

Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric) to the CRLF2 gene

Recall: Z-1099-2021 · Reported February 17, 2021

Enforcement

Recall Number
Z-1099-2021
Event ID
87090
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cytocell Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 17, 2021
Initiation Date
December 22, 2020
Classification Date
February 8, 2021
Termination Date
June 7, 2022
Address
418 Cambridge Science Park, Milton Road, Oxford Gene Technology, Cambridge, N/A, N/A, United Kingdom

Description

Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric) to the CRLF2 gene

Reason

Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.

Code Info

Lots 072240, 071390, and 072693

Distribution

Domestic distribution to California, Illinois, Michigan, Minnesota, New Jersey, Texas, and Washington. Foreign distribution to Canada.

Quantity

72