FDA Enforcement Class II Terminated

Accu-Chek Connect Diabetes Management App

Recall: Z-1099-2017 · Reported February 1, 2017

Enforcement

Recall Number
Z-1099-2017
Event ID
76147
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Roche Diabetes Care, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Other
Report Date
February 1, 2017
Initiation Date
December 20, 2016
Classification Date
January 26, 2017
Termination Date
September 5, 2017
Address
9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States

Description

Accu-Chek Connect Diabetes Management App

Reason

iOS and Android: Under certain conditions the affected app versions may disregard historical bolus data potentially leading to an incorrect bolus insulin recommendation being provided to the user. iOS only: Pairing and using multiple meters with the Accu-Chek Connect app can under rare circumstances cause the bolus advisor to fail to offer a correction bolus recommendation within the eligible time window following a blood glucose measurement (10  15 minutes). Depending on the individual metabolic situation potentially incorrect bolus advice could lead to serious health consequences such as hypoglycemia. Both software issues may also cause the amount of active insulin displayed during the bolus calculation process to be incorrect and should not be used to manually calculate a bolus.

Code Info

Catalog number 07562462001 / GTIN number 00365702700000 Catalog number 07250452001 / GTIN number 00365702700017

Distribution

Worldwide Distribution - Nationwide Distribution The Accu-Chek Connect App for iOS is distributed directly to end users via the Apple App Store. The Accu-Chek Connect App for Android is distributed directly to end users via the Google Play App Store.

Quantity

27243