FDA Enforcement Class II Terminated

Telescopic Smoke Evac Pencil, PB, Coated, Product No. SHKTSPCL. The product is a Bovie pencil that is 75 inches long compressed, with a telescoping shaft. UDI GTIN 37613327466332

Recall: Z-1098-2020 · Reported February 12, 2020

Enforcement

Recall Number
Z-1098-2020
Event ID
84721
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Instruments Div. of Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 12, 2020
Initiation Date
January 13, 2020
Classification Date
February 6, 2020
Termination Date
August 10, 2021
Address
4100 E Milham Ave, N/A, Portage, MI, 49002-9704, United States

Description

Telescopic Smoke Evac Pencil, PB, Coated, Product No. SHKTSPCL. The product is a Bovie pencil that is 75 inches long compressed, with a telescoping shaft. UDI GTIN 37613327466332

Reason

Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.

Code Info

1923011, 1923050, 1924004, 1924005, 1924015, 1929038, 1929039, 1929055

Distribution

Domestic distribution nationwide. Foreign distribution to Netherlands and Australia.

Quantity

15150