FDA Enforcement
Class II
Terminated
SafeAir Telescopic Smoke Evacuation Pencil, Product No. 0703 046 004. The product is a Bovie pencil that is 75 inches long compressed, with a telescoping shaft. UDI GTIN 37613327466332
Recall: Z-1097-2020
·
Reported February 12, 2020
Enforcement
- Recall Number
- Z-1097-2020
- Event ID
- 84721
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Instruments Div. of Stryker Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 12, 2020
- Initiation Date
- January 13, 2020
- Classification Date
- February 6, 2020
- Termination Date
- August 10, 2021
- Address
- 4100 E Milham Ave, N/A, Portage, MI, 49002-9704, United States
Description
SafeAir Telescopic Smoke Evacuation Pencil, Product No. 0703 046 004. The product is a Bovie pencil that is 75 inches long compressed, with a telescoping shaft. UDI GTIN 37613327466332
Reason
Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.
Code Info
1921032
Distribution
Domestic distribution nationwide. Foreign distribution to Netherlands and Australia.
Quantity
15150