CareStream Dental Kodak 2200 Intraoral X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures.

Recall: Z-1095-2013 · Reported May 8, 2013

Enforcement

Recall Number
Z-1095-2013
Event ID
64589
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Trophy Sas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 8, 2013
Initiation Date
September 18, 2012
Classification Date
May 2, 2013
Termination Date
May 2, 2017
Address
4 Rue Pelloutier, Croissy Beaubourg, N/A, France

Description

CareStream Dental Kodak 2200 Intraoral X-ray system Product Usage: The expected usage (intended use) for all of these products is for dental radiographic examination and diagnosis of the teeth, jaw and oral structures.

Reason

There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator.

Code Info

ABYJ011 ABYJ012 ABYJ020 through ABYJ026 ACJY522 ACYJ004 through ACYJ007 ACYJ009 through ACYJ012 ACYJ014 ACYJ015 ACYJ017 through ACYJ023 ACYJ500 through ACYJ509 ACYJ517 through ACYJ521 ACYJ523 through ACYJ531 ACYJ533 through ACYJ536 ADYJ001 through ADYJ005 ADYJ007 ADYJ008 ADYJ011 through ADYJ013 ADYJ015 through ADYJ018 ADYJ020 through ADYJ032 ADYJ034 through ADYJ037 ADYJ039 through ADYJ049 ADYJ051 ADYJ054 ADYJ056 ADYJ061 ADYJ062 ADYJ064 through ADYJ067 ADYJ069 through ADYJ073 ADYJ075 through ADYJ082 AEYJ001 through AEYJ003 AEYJ005 AEYJ006 AEYJ008 through AEYJ010 AEYP003 through AEYP011 AEYP024 through AEYP026 AEYP028 through AEYP030 AFYP005 through AFYP008 AFYP012 AFYP014 through AFYP018 AFYP021 AFYP022 AGYP010 AGYP012 AGYP016 AGYP018 through AGYP020 AGYP022 AGYP025 AGYP027 AGYP028 AGYP030 AGYP031 AGYP033 through AGYP041 AGYP045 AGYP046 AGYP068 AGYP069 AGYP073 AGYP086 AGYP088 AGYP089 AGYP092 AJYP025 AJYP027 AKYP001 AKYP007 AKYP009

Distribution

USA Nationwide Distribution

Quantity

167 installed in the US