FDA Enforcement Class II Ongoing

Azurion 7 M20 System code: (1) 722224 (2) 722079 (3) 722234 (4) 722282 System Code: (1) 722078 (2) 722223 (3) 722233

Recall: Z-1094-2025 · Reported February 12, 2025

Enforcement

Recall Number
Z-1094-2025
Event ID
96120
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 12, 2025
Initiation Date
January 13, 2025
Classification Date
February 4, 2025
Address
Veenpluis 4-6, Best, N/A, Netherlands

Description

Azurion 7 M20 System code: (1) 722224 (2) 722079 (3) 722234 (4) 722282 System Code: (1) 722078 (2) 722223 (3) 722233

Reason

Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

Code Info

UDI: (1) 722224 00884838099258(21); (2) 722079 00884838085268(21); (3) 722234 00884838116771(21); (4) 722282 00884838110571(21)

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, JAMAICA, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen.

Quantity

1346 units US; 2582 units