FDA Enforcement Class II Ongoing

Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable

Recall: Z-1092-2026 · Reported January 21, 2026

Enforcement

Recall Number
Z-1092-2026
Event ID
97872
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
C-RAD POSITIONING AB
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 21, 2026
Initiation Date
October 9, 2025
Classification Date
January 14, 2026
Address
Bredgrand 18, Uppsala, N/A, Sweden

Description

Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable

Reason

Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.

Code Info

Lot Code: Note: The error is not related to any serial numbers or LOT numbers. The error is limited to the software module cSRS in combination with cMotion module

Distribution

The software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Australia, Austria, Belarus, Belgium, Benin, Canada, China, Colombia, Egypt, El Salvador, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Mexico, Morocco, Netherlands, New Zealand, Norway, Portugal, Romania, Russia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Ukraine, and Vietnam.

Quantity

189