Sterile Procedural Trays, labeled as the following: a. C-SECTION OR 30 CDS b. C-SECTION CDS c. C-SECTION CDS d. C-SECTION CDS e. MAJOR ORTHO CDS-LF f. C-SECTION CDS g. C-SECTION CDS h. C-SECTION PROCEDURE i. C-SECTION CDS-LF j. C-SECTION CDS k. C-SECTION CDS l. C-SECTION m. C-SECTION CDS n. C-SECTION PACK-LF o. C-SECTION PACK-LF p. C-SECTION PACK-LF q. C-SECTION PACK-LF r. C-BIRTH PACK-LF s. C SECTION PACK-LF t. C-SECTION PACK-LF u. C SECTION PACK-LF v. C-SECTION PACK-LF w. C SECTION PACK-LF x. C-SECTION PACK-LF y. C-SECTION PACK z. C-SECTION PACK-LF aa. C-SECTION PACK-LF bb. C-SECTION PACK-LF cc. C-SECTION PACK-LF dd. C-BIRTH SETUP PACK-LF ee. C-SECTION PACK-LF ff. C-SECTION PACK-LF gg. C-SECTION PACK-LF hh. C SECTION PACK-LF ii. C-SECTION PACK-LF jj. C-SECTION TRAY kk. C-SECTION PACK ll. C-SECTION PACK-LF mm. C-SECTION PACK nn. C-SECTION PACK oo. C-SECTION pp. C-SECTION PACK qq. SVMMC C-SECTION PACK rr. C-SECTION PACK-LF ss. C-SECTION SETUP PACK tt. C-SECTION PACK uu. C-SECTION PACK vv. C-SECTION PACK *STANDARD* ww. C-SECTION PACK xx. C-SECTION PACK yy. C-SECTION PACK zz. C-SECTION 83305, MERCY REG aaa. CUH C-SECTION PACK-LF bbb. C-SECTION PACK-LF ccc. C-SECTION PACK ddd. C-SECTION PACK eee. C-SECTION PACK fff. C-SECTION PK ggg. C-SECTION hhh. C-SECTION PACK iii. ST JO WEST C-SECTION BASIN-LF jjj. C SECTION PACK kkk. C-SECTION PACK lll. C SECTION PACK mmm. C SECTION DELIVERY nnn. C-SECTION PACK ooo. C-SECTION PACK ppp. C SECTION qqq. WR C SECTION PACK-LF rrr. C-SECTION PACK sss. PK, L&D-C SECTION ttt. C-SECTION PACK HIGHLAND-LF uuu. C- SECTION vvv. C SECTION PACK-LF www. C-SECTION PACK xxx. C-SECTION PACK yyy. C SECTION PACK zzz. PK, L&D-C SECTION aaaa. C SECTION PACK bbbb. C SECTION PACK cccc. C-SECTION PACK dddd. SECTION PACK (CSSJE)642-LF eeee. C-SECTION PACK ffff. C-SECTION PACK-RFD gggg. C-SECTION DELIVERY PACK-LF hhhh. C-SECTION PACK iiii. C SECTION PACK jjjj. C-SECTION PACK kkkk. C SECTION llll. C SECTION PLUS PACK mmmm. C SECTION PACK-6220 nnnn. C-SECTION PACK oooo. C-SECTION PACK pppp. C-SECTION PACK - WCH qqqq. C SECTION WITH BASIN PACK rrrr. PHP C-SECTION PACK-LF ssss. C-SECTION KIT tttt. C SECTION PACK uuuu. C-SECTION vvvv. C-SECTION I PACK-LF wwww. C-SECTION PACK 2 xxxx. C-SECTION PACK-LF yyyy. C-SECTION PACK zzzz. C-SECTION PACK aaaaa. C SECTION PACK bbbbb. C-SECTION PACK ccccc. C SECTION ddddd. C SECTION PACK-LF eeeee. C SECTION PACK-LF fffff. BSMI CSECTION BIRTHING PACK ggggg. C-SECTION PACK hhhhh. C SECTION PACK-LF iiiii. C-SECTION PACK jjjjj. C SECTION PACK kkkkk. C SECTION lllll. PK, L&D-C SECTION mmmmm. C-SECTION PACK nnnnn. C SECTION ooooo. C-SECTION PACK ppppp. C-SECTION SMMC qqqqq. C SECTION PACK rrrrr. C SECTION PACK sssss. C-SECTION PACK ttttt. C SECTION #2 SHARED uuuuu. C SECTION #3 SHARED vvvvv. C SECTION wwwww. C-SECTION PACK xxxxx. CESAREAN BIRTH PACK yyyyy. C SECTION PACK-LF zzzzz. C-SECTION PACK aaaaaa. C-SECTION PACK-LF bbbbbb. C SECTION PACK cccccc. C-SECTION PACK dddddd. C-SECTION eeeeee. C-SECTION ffffff. C SECTION gggggg. C-SECTION hhhhhh. C SECTION iiiiii. C-SECTION jjjjjj. C-SECTION kkkkkk. PROHEALTH C-SECTION llllll. C SECTION mmmmmm. C SECTION nnnnnn. C-SECTION oooooo. C-SECTION pppppp. C-SECTION PACK qqqqqq. C SECTION rrrrrr. C SECTION ssssss. C-SECTION tttttt. C SECTION uuuuuu. C-SECTION - SMH vvvvvv. C-SECTION MRMC wwwwww. SCHED C SECTION xxxxxx. C-SECTION yyyyyy. C-SECTION -LF zzzzzz. ADVENTIST C-SECTION PACK aaaaaaa. C-SECTION PK-LF bbbbbbb. C-SECTION PROCEDURE ccccccc. C-SECTION II-LF ddddddd. C-SECTION I-LF eeeeeee. C-SECTION PACK fffffff. C-BIRTH PACK-LF ggggggg. C-SECTION PACK hhhhhhh. THEDACARE C-SECTION iiiiiii. C-SECTION BASIN PACK jjjjjjj. C-SECTION PACK kkkkkkk. C-SECTION PACK-LF lllllll. PPH C-SECTION PACK-LF mmmmmmm. C SECTION PACK-LF nnnnnnn. C-SECTION PACK ooooooo. C-SECTION PACK-LF
Enforcement
- Recall Number
- Z-1092-2023
- Event ID
- 91386
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- MEDLINE INDUSTRIES, LP - Northfield
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- February 15, 2023
- Initiation Date
- November 17, 2022
- Classification Date
- February 6, 2023
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Sterile Procedural Trays, labeled as the following: a. C-SECTION OR 30 CDS b. C-SECTION CDS c. C-SECTION CDS d. C-SECTION CDS e. MAJOR ORTHO CDS-LF f. C-SECTION CDS g. C-SECTION CDS h. C-SECTION PROCEDURE i. C-SECTION CDS-LF j. C-SECTION CDS k. C-SECTION CDS l. C-SECTION m. C-SECTION CDS n. C-SECTION PACK-LF o. C-SECTION PACK-LF p. C-SECTION PACK-LF q. C-SECTION PACK-LF r. C-BIRTH PACK-LF s. C SECTION PACK-LF t. C-SECTION PACK-LF u. C SECTION PACK-LF v. C-SECTION PACK-LF w. C SECTION PACK-LF x. C-SECTION PACK-LF y. C-SECTION PACK z. C-SECTION PACK-LF aa. C-SECTION PACK-LF bb. C-SECTION PACK-LF cc. C-SECTION PACK-LF dd. C-BIRTH SETUP PACK-LF ee. C-SECTION PACK-LF ff. C-SECTION PACK-LF gg. C-SECTION PACK-LF hh. C SECTION PACK-LF ii. C-SECTION PACK-LF jj. C-SECTION TRAY kk. C-SECTION PACK ll. C-SECTION PACK-LF mm. C-SECTION PACK nn. C-SECTION PACK oo. C-SECTION pp. C-SECTION PACK qq. SVMMC C-SECTION PACK rr. C-SECTION PACK-LF ss. C-SECTION SETUP PACK tt. C-SECTION PACK uu. C-SECTION PACK vv. C-SECTION PACK *STANDARD* ww. C-SECTION PACK xx. C-SECTION PACK yy. C-SECTION PACK zz. C-SECTION 83305, MERCY REG aaa. CUH C-SECTION PACK-LF bbb. C-SECTION PACK-LF ccc. C-SECTION PACK ddd. C-SECTION PACK eee. C-SECTION PACK fff. C-SECTION PK ggg. C-SECTION hhh. C-SECTION PACK iii. ST JO WEST C-SECTION BASIN-LF jjj. C SECTION PACK kkk. C-SECTION PACK lll. C SECTION PACK mmm. C SECTION DELIVERY nnn. C-SECTION PACK ooo. C-SECTION PACK ppp. C SECTION qqq. WR C SECTION PACK-LF rrr. C-SECTION PACK sss. PK, L&D-C SECTION ttt. C-SECTION PACK HIGHLAND-LF uuu. C- SECTION vvv. C SECTION PACK-LF www. C-SECTION PACK xxx. C-SECTION PACK yyy. C SECTION PACK zzz. PK, L&D-C SECTION aaaa. C SECTION PACK bbbb. C SECTION PACK cccc. C-SECTION PACK dddd. SECTION PACK (CSSJE)642-LF eeee. C-SECTION PACK ffff. C-SECTION PACK-RFD gggg. C-SECTION DELIVERY PACK-LF hhhh. C-SECTION PACK iiii. C SECTION PACK jjjj. C-SECTION PACK kkkk. C SECTION llll. C SECTION PLUS PACK mmmm. C SECTION PACK-6220 nnnn. C-SECTION PACK oooo. C-SECTION PACK pppp. C-SECTION PACK - WCH qqqq. C SECTION WITH BASIN PACK rrrr. PHP C-SECTION PACK-LF ssss. C-SECTION KIT tttt. C SECTION PACK uuuu. C-SECTION vvvv. C-SECTION I PACK-LF wwww. C-SECTION PACK 2 xxxx. C-SECTION PACK-LF yyyy. C-SECTION PACK zzzz. C-SECTION PACK aaaaa. C SECTION PACK bbbbb. C-SECTION PACK ccccc. C SECTION ddddd. C SECTION PACK-LF eeeee. C SECTION PACK-LF fffff. BSMI CSECTION BIRTHING PACK ggggg. C-SECTION PACK hhhhh. C SECTION PACK-LF iiiii. C-SECTION PACK jjjjj. C SECTION PACK kkkkk. C SECTION lllll. PK, L&D-C SECTION mmmmm. C-SECTION PACK nnnnn. C SECTION ooooo. C-SECTION PACK ppppp. C-SECTION SMMC qqqqq. C SECTION PACK rrrrr. C SECTION PACK sssss. C-SECTION PACK ttttt. C SECTION #2 SHARED uuuuu. C SECTION #3 SHARED vvvvv. C SECTION wwwww. C-SECTION PACK xxxxx. CESAREAN BIRTH PACK yyyyy. C SECTION PACK-LF zzzzz. C-SECTION PACK aaaaaa. C-SECTION PACK-LF bbbbbb. C SECTION PACK cccccc. C-SECTION PACK dddddd. C-SECTION eeeeee. C-SECTION ffffff. C SECTION gggggg. C-SECTION hhhhhh. C SECTION iiiiii. C-SECTION jjjjjj. C-SECTION kkkkkk. PROHEALTH C-SECTION llllll. C SECTION mmmmmm. C SECTION nnnnnn. C-SECTION oooooo. C-SECTION pppppp. C-SECTION PACK qqqqqq. C SECTION rrrrrr. C SECTION ssssss. C-SECTION tttttt. C SECTION uuuuuu. C-SECTION - SMH vvvvvv. C-SECTION MRMC wwwwww. SCHED C SECTION xxxxxx. C-SECTION yyyyyy. C-SECTION -LF zzzzzz. ADVENTIST C-SECTION PACK aaaaaaa. C-SECTION PK-LF bbbbbbb. C-SECTION PROCEDURE ccccccc. C-SECTION II-LF ddddddd. C-SECTION I-LF eeeeeee. C-SECTION PACK fffffff. C-BIRTH PACK-LF ggggggg. C-SECTION PACK hhhhhhh. THEDACARE C-SECTION iiiiiii. C-SECTION BASIN PACK jjjjjjj. C-SECTION PACK kkkkkkk. C-SECTION PACK-LF lllllll. PPH C-SECTION PACK-LF mmmmmmm. C SECTION PACK-LF nnnnnnn. C-SECTION PACK ooooooo. C-SECTION PACK-LF
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
a. C-SECTION OR 30 CDS, Item Number: CDS820075W, Case UDI/GTIN: 40195327209798, Unit UDI/GTIN: 10195327209797, Lot Number: 22JBX603,22JBX483; b. C-SECTION CDS, Item Number: CDS830218J, Case UDI/GTIN: 40888277530172, Unit UDI/GTIN: 10888277530171, Lot Number: 22JBV584; c. C-SECTION CDS, Item Number: CDS980747J, Case UDI/GTIN: 40193489923613, Unit UDI/GTIN: 10193489923612, Lot Number: 22JBO006; d. C-SECTION CDS, Item Number: CDS981430C, Case UDI/GTIN: 40193489260879, Unit UDI/GTIN: 10193489260878, Lot Number: 22JMG179; e. MAJOR ORTHO CDS-LF, Item Number: CDS982233G, Case UDI/GTIN: 40193489986212, Unit UDI/GTIN: 10193489986211, Lot Number: 22JMG573; f. C-SECTION CDS, Item Number: CDS982367I, Case UDI/GTIN: 40193489312349, Unit UDI/GTIN: 10193489312348, Lot Number: 22KBE753,22KBC690; g. C-SECTION CDS, Item Number: CDS983117G, Case UDI/GTIN: 40195327192113, Unit UDI/GTIN: 10195327192112, Lot Number: 22IBR943; h. C-SECTION PROCEDURE, Item Number: CDS983450L, Case UDI/GTIN: 40193489372336, Unit UDI/GTIN: 10193489372335, Lot Number: 22KBE444,22IBD865,22IBE112; i. C-SECTION CDS-LF, Item Number: CDS983869I, Case UDI/GTIN: 40888277752734, Unit UDI/GTIN: 10888277752733, Lot Number: 22JDB259; j. C-SECTION CDS, Item Number: CDS983917M, Case UDI/GTIN: 40193489402712, Unit UDI/GTIN: 10193489402711, Lot Number: 22JBO248; k. C-SECTION CDS, Item Number: CDS984354C, Case UDI/GTIN: 40193489738187, Unit UDI/GTIN: 10193489738186, Lot Number: 22JBC335; l. C-SECTION, Item Number: CDS984902K, Case UDI/GTIN: 40195327156924, Unit UDI/GTIN: 10195327156923, Lot Number: 22IBQ711,22IBL648; m. C-SECTION CDS, Item Number: CDS985206B, Case UDI/GTIN: 40889942942078, Unit UDI/GTIN: 10889942942077, Lot Number: 22JBL398; n. C-SECTION PACK-LF, Item Number: DYNJ0173422AF, Case UDI/GTIN: 40195327097579, Unit UDI/GTIN: 10195327097578, Lot Number: 22JMA807; o. C-SECTION PACK-LF, Item Number: DYNJ0188507C, Case UDI/GTIN: 40889942053491, Unit UDI/GTIN: 10889942053490, Lot Number: 22JBY657,22IBI286; p. C-SECTION PACK-LF, Item Number: DYNJ0200274P, Case UDI/GTIN: 40889942054115, Unit UDI/GTIN: 10889942054114, Lot Number: 22IMG322; q. C-SECTION PACK-LF, Item Number: DYNJ0365888N, Case UDI/GTIN: 40195327045273, Unit UDI/GTIN: 10195327045272, Lot Number: 22IMF444,22HMF014; r. C-BIRTH PACK-LF, Item Number: DYNJ0366291M, Case UDI/GTIN: 40889942059615, Unit UDI/GTIN: 10889942059614, Lot Number: 22IME718; s. C SECTION PACK-LF, Item Number: DYNJ0367189M, Case UDI/GTIN: 40193489788298, Unit UDI/GTIN: 10193489788297, Lot Number: 22IMG638,22HMC273; t. C-SECTION PACK-LF, Item Number: DYNJ0367679W, Case UDI/GTIN: 40193489617369, Unit UDI/GTIN: 10193489617368, Lot Number: 22JMI197,22IMD016; u. C SECTION PACK-LF, Item Number: DYNJ0376534Q, Case UDI/GTIN: 40195327237173, Unit UDI/GTIN: 10195327237172, Lot Number: 22IBE468; v. C-SECTION PACK-LF, Item Number: DYNJ0382235G, Case UDI/GTIN: 40889942435020, Unit UDI/GTIN: 10889942435029, Lot Number: 22IBC549,22IBC247,22HBG107; w. C SECTION PACK-LF, Item Number: DYNJ0397841N, Case UDI/GTIN: 40193489824064, Unit UDI/GTIN: 10193489824063, Lot Number: 22IBL780; x. C-SECTION PACK-LF, Item Number: DYNJ04609I, Case UDI/GTIN: 40193489552707, Unit UDI/GTIN: 10193489552706, Lot Number: 22IMA202; y. C-SECTION PACK, Item Number: DYNJ04620S, Case UDI/GTIN: 40195327135851, Unit UDI/GTIN: 10195327135850, Lot Number: 22IMB266; z. C-SECTION PACK-LF, Item Number: DYNJ04724Q, Case UDI/GTIN: 40193489554015, Unit UDI/GTIN: 10193489554014, Lot Number: 22IMF982; aa. C-SECTION PACK-LF, Item Number: DYNJ0508830M, Case UDI/GTIN: 40193489579247, Unit UDI/GTIN: 10193489579246, Lot Number: 22IMD130; bb. C-SECTION PACK-LF, Item Number: DYNJ0551773O, Case UDI/GTIN: 40193489268622, Unit UDI/GTIN: 10193489268621, Lot Number: 22JBS505; cc. C-SECTION PACK-LF, Item Number: DYNJ06584K, Case UDI/GTIN: 40193489420525, Unit UDI/GTIN: 10193489420524, Lot Number: 22JBY658,22IBU471,22HBL261; dd. C-BIRTH SETUP PACK-LF, Item Number: DYNJ0781722T, Case UDI/GTIN: 40193489226356, Unit UDI/GTIN: 10193489226355, Lot Number: 22JMG062; ee. C-SECTION PACK-LF, Item Number: DYNJ07836Q, Case UDI/GTIN: 40195327209552, Unit UDI/GTIN: 10195327209551, Lot Number: 22IBO410; ff. C-SECTION PACK-LF, Item Number: DYNJ0842739K, Case UDI/GTIN: 40193489359634, Unit UDI/GTIN: 10193489359633, Lot Number: 22JDB302; gg. C-SECTION PACK-LF, Item Number: DYNJ0843654G, Case UDI/GTIN: 40889942402855, Unit UDI/GTIN: 10889942402854, Lot Number: 22IDB683,22IDA305; hh. C SECTION PACK-LF, Item Number: DYNJ0923021, Case UDI/GTIN: 40889942099642, Unit UDI/GTIN: 10889942099641, Lot Number: 22JMH549,22IMF460; ii. C-SECTION PACK-LF, Item Number: DYNJ15232M, Case UDI/GTIN: 40195327231744, Unit UDI/GTIN: 10195327231743, Lot Number: 22IMH405; jj. C-SECTION TRAY, Item Number: DYNJ15494G, Case UDI/GTIN: 40193489961196, Unit UDI/GTIN: 10193489961195, Lot Number: 22IBI076; kk. C-SECTION PACK, Item Number: DYNJ15726I, Case UDI/GTIN: 40193489832403, Unit UDI/GTIN: 10193489832402, Lot Number: 22JBU168; ll. C-SECTION PACK-LF, Item Number: DYNJ16672I, Case UDI/GTIN: 40193489990752, Unit UDI/GTIN: 10193489990751, Lot Number: 22JBK965,22IBL997,22HBU474,22HBV462; mm. C-SECTION PACK, Item Number: DYNJ19958P, Case UDI/GTIN: 40193489867733, Unit UDI/GTIN: 10193489867732, Lot Number: 22JMI344,22IMG005,22HME230; nn. C-SECTION PACK, Item Number: DYNJ20135M, Case UDI/GTIN: 40193489905534, Unit UDI/GTIN: 10193489905533, Lot Number: 22JBS400; oo. C-SECTION, Item Number: DYNJ22368I, Case UDI/GTIN: 40193489780513, Unit UDI/GTIN: 10193489780512, Lot Number: 22IBN332; pp. C-SECTION PACK, Item Number: DYNJ24120C, Case UDI/GTIN: 40884389396507, Unit UDI/GTIN: 10884389396506, Lot Number: 22IBF425; qq. SVMMC C-SECTION PACK, Item Number: DYNJ24632K, Case UDI/GTIN: 40195327058624, Unit UDI/GTIN: 10195327058623, Lot Number: 22JLB138,22ILA591; rr. C-SECTION PACK-LF, Item Number: DYNJ26665J, Case UDI/GTIN: 40889942112136, Unit UDI/GTIN: 10889942112135, Lot Number: 22JBW002; ss. C-SECTION SETUP PACK, Item Number: DYNJ27751F, Case UDI/GTIN: 40195327228706, Unit UDI/GTIN: 10195327228705, Lot Number: 22JDA003,22HDC395,22HDA918; tt. C-SECTION PACK, Item Number: DYNJ29034F, Case UDI/GTIN: 40193489461146, Unit UDI/GTIN: 10193489461145, Lot Number: 22IME948; uu. C-SECTION PACK, Item Number: DYNJ30483F, Case UDI/GTIN: 40193489912419, Unit UDI/GTIN: 10193489912418, Lot Number: 22JMB299,22HMB825; vv. C-SECTION PACK *STANDARD*, Item Number: DYNJ30609G, Case UDI/GTIN: 40193489301541, Unit UDI/GTIN: 10193489301540, Lot Number: 22IMG006; ww. C-SECTION PACK, Item Number: DYNJ30610D, Case UDI/GTIN: 40889942491972, Unit UDI/GTIN: 10889942491971, Lot Number: 22HDC051; xx. C-SECTION PACK, Item Number: DYNJ31011D, Case UDI/GTIN: 40889942383413, Unit UDI/GTIN: 10889942383412, Lot Number: 22JBR171; yy. C-SECTION PACK, Item Number: DYNJ32011G, Case UDI/GTIN: 40195327067046, Unit UDI/GTIN: 10195327067045, Lot Number: 22JMH079; zz. C-SECTION 83305, MERCY REG, Item Number: DYNJ32059F, Case UDI/GTIN: 40193489822497, Unit UDI/GTIN: 10193489822496, Lot Number: 22IBV950; aaa. CUH C-SECTION PACK-LF, Item Number: DYNJ32253G, Case UDI/GTIN: 40195327066537, Unit UDI/GTIN: 10195327066536, Lot Number: 22IBJ049; bbb. C-SECTION PACK-LF, Item Number: DYNJ33059B, Case UDI/GTIN: 40193489950947, Unit UDI/GTIN: 10193489950946, Lot Number: 22JBO407,22IBG828,22HBR300; ccc. C-SECTION PACK, Item Number: DYNJ33188B, Case UDI/GTIN: 40888277746283, Unit UDI/GTIN: 10888277746282, Lot Number: 22JMH212,22IMI213; ddd. C-SECTION PACK, Item Number: DYNJ33550B, Case UDI/GTIN: 40889942435686, Unit UDI/GTIN: 10889942435685, Lot Number: 22JMF403; eee. C-SECTION PACK, Item Number: DYNJ33799C, Case UDI/GTIN: 40889942447153, Unit UDI/GTIN: 10889942447152, Lot Number: 22JBI238,22HBG189; fff. C-SECTION PK, Item Number: DYNJ34266B, Case UDI/GTIN: 40888277246516, Unit UDI/GTIN: 10888277246515, Lot Number: 22IBK679; ggg. C-SECTION, Item Number: DYNJ34431G, Case UDI/GTIN: 40193489834957, Unit UDI/GTIN: 10193489834956, Lot Number: 22JBN841; hhh. C-SECTION PACK, Item Number: DYNJ36310K, Case UDI/GTIN: 40193489443456, Unit UDI/GTIN: 10193489443455, Lot Number: 22JMG326; iii. ST JO WEST C-SECTION BASIN-LF, Item Number: DYNJ37027A, Case UDI/GTIN: 40889942418320, Unit UDI/GTIN: 10889942418329, Lot Number: 22JBR330,22HBT598; jjj. C SECTION PACK, Item Number: DYNJ37622C, Case UDI/GTIN: 40884389864235, Unit UDI/GTIN: 10884389864234, Lot Number: 22IBT225; kkk. C-SECTION PACK, Item Number: DYNJ38350J, Case UDI/GTIN: 40195327129959, Unit UDI/GTIN: 10195327129958, Lot Number: 22JMC173,22IME739,22HMD564; lll. C SECTION PACK, Item Number: DYNJ38440C, Case UDI/GTIN: 40195327061792, Unit UDI/GTIN: 10195327061791, Lot Number: 22JMC367,22IMF067; mmm. C SECTION DELIVERY, Item Number: DYNJ39375F, Case UDI/GTIN: 40193489418980, Unit UDI/GTIN: 10193489418989, Lot Number: 22IBG340; nnn. C-SECTION PACK, Item Number: DYNJ40458B, Case UDI/GTIN: 40889942109334, Unit UDI/GTIN: 10889942109333, Lot Number: 22IBS430,22HBH047; ooo. C-SECTION PACK, Item Number: DYNJ40503B, Case UDI/GTIN: 40193489614931, Unit UDI/GTIN: 10193489614930, Lot Number: 22IDB276; ppp. C SECTION, Item Number: DYNJ41268B, Case UDI/GTIN: 40193489487627, Unit UDI/GTIN: 10193489487626, Lot Number: 22IMF995,22IMF998; qqq. WR C SECTION PACK-LF, Item Number: DYNJ41570K, Case UDI/GTIN: 40195327089772, Unit UDI/GTIN: 10195327089771, Lot Number: 22IMF970,22IMG359; rrr. C-SECTION PACK, Item Number: DYNJ42226D, Case UDI/GTIN: 40195327017935, Unit UDI/GTIN: 10195327017934, Lot Number: 22IMG051,22IMG068; sss. PK, L&D-C SECTION, Item Number: DYNJ42476I, Case UDI/GTIN: 40193489266017, Unit UDI/GTIN: 10193489266016, Lot Number: 22JBU242; ttt. C-SECTION PACK HIGHLAND-LF, Item Number: DYNJ42634B, Case UDI/GTIN: 40888277582805, Unit UDI/GTIN: 10888277582804, Lot Number: 22IBD370; uuu. C- SECTION, Item Number: DYNJ42651, Case UDI/GTIN: 40888277003584, Unit UDI/GTIN: 10888277003583, Lot Number: 22IME346,22IME347,22HMC131; vvv. C SECTION PACK-LF, Item Number: DYNJ42850A, Case UDI/GTIN: 40193489880909, Unit UDI/GTIN: 10193489880908, Lot Number: 22JBN852; www. C-SECTION PACK, Item Number: DYNJ42946A, Case UDI/GTIN: 40195327196838, Unit UDI/GTIN: 10195327196837, Lot Number: 22IBV689,22HBH787; xxx. C-SECTION PACK, Item Number: DYNJ43284D, Case UDI/GTIN: 40888277984128, Unit UDI/GTIN: 10888277984127, Lot Number: 22IMG361; yyy. C SECTION PACK, Item Number: DYNJ43732F, Case UDI/GTIN: 40193489549370, Unit UDI/GTIN: 10193489549379, Lot Number: 22IBI952; zzz. PK, L&D-C SECTION, Item Number: DYNJ43746C, Case UDI/GTIN: 40193489310635, Unit UDI/GTIN: 10193489310634, Lot Number: 22JBO122; aaaa. C SECTION PACK, Item Number: DYNJ43810A, Case UDI/GTIN: 40193489298056, Unit UDI/GTIN: 10193489298055, Lot Number: 22IBK239; bbbb. C SECTION PACK, Item Number: DYNJ46508F, Case UDI/GTIN: 40195327111787, Unit UDI/GTIN: 10195327111786, Lot Number: 22JME929; cccc. C-SECTION PACK, Item Number: DYNJ47220C, Case UDI/GTIN: 40889942131359, Unit UDI/GTIN: 10889942131358, Lot Number: 22JMD043; dddd. SECTION PACK (CSSJE)642-LF, Item Number: DYNJ47485C, Case UDI/GTIN: 40193489586597, Unit UDI/GTIN: 10193489586596, Lot Number: 22IMI230; eeee. C-SECTION PACK, Item Number: DYNJ47606B, Case UDI/GTIN: 40193489730693, Unit UDI/GTIN: 10193489730692, Lot Number: 22IBV197; ffff. C-SECTION PACK-RFD, Item Number: DYNJ47761K, Case UDI/GTIN: 40193489896986, Unit UDI/GTIN: 10193489896985, Lot Number: 22IME390; gggg. C-SECTION DELIVERY PACK-LF, Item Number: DYNJ47808F, Case UDI/GTIN: 40195327232604, Unit UDI/GTIN: 10195327232603, Lot Number: 22IMB729; hhhh. C-SECTION PACK, Item Number: DYNJ48272, Case UDI/GTIN: 40888277575371, Unit UDI/GTIN: 10888277575370, Lot Number: 22IBO504,22HMD672; iiii. C SECTION PACK, Item Number: DYNJ48483D, Case UDI/GTIN: 40195327030811, Unit UDI/GTIN: 10195327030810, Lot Number: 22JMI251; jjjj. C-SECTION PACK, Item Number: DYNJ48746A, Case UDI/GTIN: 40889942385691, Unit UDI/GTIN: 10889942385690, Lot Number: 22JMA582; kkkk. C SECTION, Item Number: DYNJ49172G, Case UDI/GTIN: 40193489308243, Unit UDI/GTIN: 10193489308242, Lot Number: 22IMC797,22HME909; llll. C SECTION PLUS PACK, Item Number: DYNJ49370D, Case UDI/GTIN: 40193489213318, Unit UDI/GTIN: 10193489213317, Lot Number: 22JBO103,22HBR997,22HBG119; mmmm. C SECTION PACK-6220, Item Number: DYNJ49519A, Case UDI/GTIN: 40889942024835, Unit UDI/GTIN: 10889942024834, Lot Number: 22IBN420; nnnn. C-SECTION PACK, Item Number: DYNJ49668A, Case UDI/GTIN: 40889942390176, Unit UDI/GTIN: 10889942390175, Lot Number: 22JBS277,22IBR285; oooo. C-SECTION PACK, Item Number: DYNJ49771, Case UDI/GTIN: 40888277881144, Unit UDI/GTIN: 10888277881143, Lot Number: 22IBQ282; pppp. C-SECTION PACK - WCH, Item Number: DYNJ49829F, Case UDI/GTIN: 40193489631570, Unit UDI/GTIN: 10193489631579, Lot Number: 22JBX669; qqqq. C SECTION WITH BASIN PACK, Item Number: DYNJ51378G, Case UDI/GTIN: 40193489996389, Unit UDI/GTIN: 10193489996388, Lot Number: 22IBN196; rrrr. PHP C-SECTION PACK-LF, Item Number: DYNJ51603D, Case UDI/GTIN: 40195327167050, Unit UDI/GTIN: 10195327167059, Lot Number: 22JMG359,22IMA243; ssss. C-SECTION KIT, Item Number: DYNJ51655C, Case UDI/GTIN: 40195327187850, Unit UDI/GTIN: 10195327187859, Lot Number: 22IME130; tttt. C SECTION PACK, Item Number: DYNJ51872D, Case UDI/GTIN: 40195327043439, Unit UDI/GTIN: 10195327043438, Lot Number: 22JBI664; uuuu. C-SECTION, Item Number: DYNJ51997, Case UDI/GTIN: 40889942233107, Unit UDI/GTIN: 10889942233106, Lot Number: 22JBP756,22IBR427; vvvv. C-SECTION I PACK-LF, Item Number: DYNJ52040F, Case UDI/GTIN: 40195327096145, Unit UDI/GTIN: 10195327096144, Lot Number: 22JBO202,22JBN048,22HBG758; wwww. C-SECTION PACK 2, Item Number: DYNJ52168A, Case UDI/GTIN: 40193489539739, Unit UDI/GTIN: 10193489539738, Lot Number: 22JBD226,22IBP554,22IBM605; xxxx. C-SECTION PACK-LF, Item Number: DYNJ52908A, Case UDI/GTIN: 40889942851981, Unit UDI/GTIN: 10889942851980, Lot Number: 22IMC339,22HMC091; yyyy. C-SECTION PACK, Item Number: DYNJ52931D, Case UDI/GTIN: 40195327200221, Unit UDI/GTIN: 10195327200220, Lot Number: 22IBR773; zzzz. C-SECTION PACK, Item Number: DYNJ53386C, Case UDI/GTIN: 40195327169696, Unit UDI/GTIN: 10195327169695, Lot Number: 22JMF746; aaaaa. C SECTION PACK, Item Number: DYNJ53399I, Case UDI/GTIN: 40193489968478, Unit UDI/GTIN: 10193489968477, Lot Number: 22JMF705; bbbbb. C-SECTION PACK, Item Number: DYNJ53598F, Case UDI/GTIN: 40195327055821, Unit UDI/GTIN: 10195327055820, Lot Number: 22IMB353,22IMB057; ccccc. C SECTION, Item Number: DYNJ53654F, Case UDI/GTIN: 40193489454674, Unit UDI/GTIN: 10193489454673, Lot Number: 22IBI089; ddddd. C SECTION PACK-LF, Item Number: DYNJ54186A, Case UDI/GTIN: 40195327062607, Unit UDI/GTIN: 10195327062606, Lot Number: 22JMF820; eeeee. C SECTION PACK-LF, Item Number: DYNJ54374J, Case UDI/GTIN: 40193489404075, Unit UDI/GTIN: 10193489404074, Lot Number: 22IBM226; fffff. BSMI CSECTION BIRTHING PACK, Item Number: DYNJ54946G, Case UDI/GTIN: 40195327135455, Unit UDI/GTIN: 10195327135454, Lot Number: 22IBJ472,22HBA609; ggggg. C-SECTION PACK, Item Number: DYNJ55336A, Case UDI/GTIN: 40889942781097, Unit UDI/GTIN: 10889942781096, Lot Number: 22KBB430; hhhhh. C SECTION PACK-LF, Item Number: DYNJ55466C, Case UDI/GTIN: 40193489408158, Unit UDI/GTIN: 10193489408157, Lot Number: 22IBN820,22HBW539; iiiii. C-SECTION PACK, Item Number: DYNJ55719, Case UDI/GTIN: 40889942574590, Unit UDI/GTIN: 10889942574599, Lot Number: 22IBR318; jjjjj. C SECTION PACK, Item Number: DYNJ57193A, Case UDI/GTIN: 40193489647656, Unit UDI/GTIN: 10193489647655, Lot Number: 22IBF463; kkkkk. C SECTION, Item Number: DYNJ57622, Case UDI/GTIN: 40889942757245, Unit UDI/GTIN: 10889942757244, Lot Number: 22IBF901; lllll. PK, L&D-C SECTION, Item Number: DYNJ58239B, Case UDI/GTIN: 40195327248766, Unit UDI/GTIN: 10195327248765, Lot Number: 22IMH256,22IMH497; mmmmm. C-SECTION PACK, Item Number: DYNJ58493A, Case UDI/GTIN: 40193489517423, Unit UDI/GTIN: 10193489517422, Lot Number: 22JBU325; nnnnn. C SECTION, Item Number: DYNJ58551, Case UDI/GTIN: 40889942870982, Unit UDI/GTIN: 10889942870981, Lot Number: 22JBO158; ooooo. C-SECTION PACK, Item Number: DYNJ59414A, Case UDI/GTIN: 40195327066452, Unit UDI/GTIN: 10195327066451, Lot Number: 22JMF097; ppppp. C-SECTION SMMC, Item Number: DYNJ59457A, Case UDI/GTIN: 40195327243266, Unit UDI/GTIN: 10195327243265, Lot Number: 22JMD987; qqqqq. C SECTION PACK, Item Number: DYNJ61573, Case UDI/GTIN: 40193489601078, Unit UDI/GTIN: 10193489601077, Lot Number: 22JMA868,22IMF177; rrrrr. C SECTION PACK, Item Number: DYNJ64493A, Case UDI/GTIN: 40193489323260, Unit UDI/GTIN: 10193489323269, Lot Number: 22IBI098; sssss. C-SECTION PACK, Item Number: DYNJ64678A, Case UDI/GTIN: 40195327225996, Unit UDI/GTIN: 10195327225995, Lot Number: 22JDB087; ttttt. C SECTION #2 SHARED, Item Number: DYNJ64773C, Case UDI/GTIN: 40195327221066, Unit UDI/GTIN: 10195327221065, Lot Number: 22IDA431; uuuuu. C SECTION #3 SHARED, Item Number: DYNJ64774D, Case UDI/GTIN: 40195327233885, Unit UDI/GTIN: 10195327233884, Lot Number: 22IDC063,22IDA788; vvvvv. C SECTION, Item Number: DYNJ65646B, Case UDI/GTIN: 40195327146789, Unit UDI/GTIN: 10195327146788, Lot Number: 22IBF374; wwwww. C-SECTION PACK, Item Number: DYNJ65961B, Case UDI/GTIN: 40193489984300, Unit UDI/GTIN: 10193489984309, Lot Number: 22IME243; xxxxx. CESAREAN BIRTH PACK, Item Number: DYNJ67620, Case UDI/GTIN: 40193489418164, Unit UDI/GTIN: 10193489418163, Lot Number: 22JBX125,22HBR751,22HBS988; yyyyy. C SECTION PACK-LF, Item Number: DYNJ67957, Case UDI/GTIN: 40193489458979, Unit UDI/GTIN: 10193489458978, Lot Number: 22JMH427,22JMD099,22IMH525; zzzzz. C-SECTION PACK, Item Number: DYNJ68954, Case UDI/GTIN: 40193489863469, Unit UDI/GTIN: 10193489863468, Lot Number: 22JMF850; aaaaaa. C-SECTION PACK-LF, Item Number: DYNJ69445, Case UDI/GTIN: 40193489942515, Unit UDI/GTIN: 10193489942514, Lot Number: 22JBR185,22IBG076; bbbbbb. C SECTION PACK, Item Number: DYNJ82554, Case UDI/GTIN: 40195327198245, Unit UDI/GTIN: 10195327198244, Lot Number: 22JBQ815; cccccc. C-SECTION PACK, Item Number: DYNJ83021, Case UDI/GTIN: 40195327228911, Unit UDI/GTIN: 10195327228910, Lot Number: 22JBN586; dddddd. C-SECTION, Item Number: DYNJ900260A, Case UDI/GTIN: 40195327178537, Unit UDI/GTIN: 10195327178536, Lot Number: 22IBQ946; eeeeee. C-SECTION, Item Number: DYNJ900266C, Case UDI/GTIN: 40193489278430, Unit UDI/GTIN: 10193489278439, Lot Number: 22IBS357; ffffff. C SECTION, Item Number: DYNJ900794, Case UDI/GTIN: 40888277906601, Unit UDI/GTIN: 10888277906600, Lot Number: 22KBF556,22JBM372,22IBI225; gggggg. C-SECTION, Item Number: DYNJ902761D, Case UDI/GTIN: 40193489891059, Unit UDI/GTIN: 10193489891058, Lot Number: 22JMG886,22JMD346; hhhhhh. C SECTION, Item Number: DYNJ903771B, Case UDI/GTIN: 40193489935968, Unit UDI/GTIN: 10193489935967, Lot Number: 22JBN602,22JBN603,22HBX117; iiiiii. C-SECTION, Item Number: DYNJ904357D, Case UDI/GTIN: 40195327223091, Unit UDI/GTIN: 10195327223090, Lot Number: 22JBE816,22IBR279; jjjjjj. C-SECTION, Item Number: DYNJ904612F, Case UDI/GTIN: 40193489844048, Unit UDI/GTIN: 10193489844047, Lot Number: 22JBA083; kkkkkk. PROHEALTH C-SECTION, Item Number: DYNJ904879D, Case UDI/GTIN: 40193489400107, Unit UDI/GTIN: 10193489400106, Lot Number: 22KBC325; llllll. C SECTION, Item Number: DYNJ905314G, Case UDI/GTIN: 40193489930536, Unit UDI/GTIN: 10193489930535, Lot Number: 22JBR953; mmmmmm. C SECTION, Item Number: DYNJ905776C, Case UDI/GTIN: 40195327179862, Unit UDI/GTIN: 10195327179861, Lot Number: 22KBF490; nnnnnn. C-SECTION, Item Number: DYNJ905792C, Case UDI/GTIN: 40195327108428, Unit UDI/GTIN: 10195327108427, Lot Number: 22IMF389; oooooo. C-SECTION, Item Number: DYNJ906470D, Case UDI/GTIN: 40195327204243, Unit UDI/GTIN: 10195327204242, Lot Number: 22IDC048; pppppp. C-SECTION PACK, Item Number: DYNJ906557A, Case UDI/GTIN: 40193489436373, Unit UDI/GTIN: 10193489436372, Lot Number: 22IBT728; qqqqqq. C SECTION, Item Number: DYNJ906917A, Case UDI/GTIN: 40193489460378, Unit UDI/GTIN: 10193489460377, Lot Number: 22JBJ828; rrrrrr. C SECTION, Item Number: DYNJ907073, Case UDI/GTIN: 40193489337069, Unit UDI/GTIN: 10193489337068, Lot Number: 22IBE728; ssssss. C-SECTION, Item Number: DYNJ907228F, Case UDI/GTIN: 40195327247448, Unit UDI/GTIN: 10195327247447, Lot Number: 22JBC329; tttttt. C SECTION, Item Number: DYNJ907660A, Case UDI/GTIN: 40195327168668, Unit UDI/GTIN: 10195327168667, Lot Number: 22IMG417; uuuuuu. C-SECTION - SMH, Item Number: DYNJ907724A, Case UDI/GTIN: 40193489976381, Unit UDI/GTIN: 10193489976380, Lot Number: 22JBH655,22IBA564; vvvvvv. C-SECTION MRMC, Item Number: DYNJ907745B, Case UDI/GTIN: 40195327195411, Unit UDI/GTIN: 10195327195410, Lot Number: 22KBE700,22IBE731,22HBU387; wwwwww. SCHED C SECTION, Item Number: DYNJ907956, Case UDI/GTIN: 40193489949439, Unit UDI/GTIN: 10193489949438, Lot Number: 22IBQ951; xxxxxx. C-SECTION, Item Number: DYNJ908197, Case UDI/GTIN: 40195327006502, Unit UDI/GTIN: 10195327006501, Lot Number: 22JDB121; yyyyyy. C-SECTION -LF, Item Number: DYNJ9932207R, Case UDI/GTIN: 40193489624152, Unit UDI/GTIN: 10193489624151, Lot Number: 22JMD352; zzzzzz. ADVENTIST C-SECTION PACK, Item Number: DYNJADV101A, Case UDI/GTIN: 40888277576170, Unit UDI/GTIN: 10888277576179, Lot Number: 22IBS980,22HBW715; aaaaaaa. C-SECTION PK-LF, Item Number: DYNJC2363J, Case UDI/GTIN: 40195327131341, Unit UDI/GTIN: 10195327131340, Lot Number: 22HBI849,22HBI850; bbbbbbb. C-SECTION PROCEDURE, Item Number: DYNJQ9038I, Case UDI/GTIN: 40889942629948, Unit UDI/GTIN: 10889942629947, Lot Number: 22HBR429; ccccccc. C-SECTION II-LF, Item Number: DYNJS0612, Case UDI/GTIN: 40080196942504, Unit UDI/GTIN: 10080196942503, Lot Number: 22JBK525,22IBP836,22HBX278,22HBN836,22HBG183,22HBG185; ddddddd. C-SECTION I-LF, Item Number: DYNJS3038, Case UDI/GTIN: 40884389450216, Unit UDI/GTIN: 10884389450215, Lot Number: 22JDA185,22HDC283,22HDB455,22HDA724; eeeeeee. C-SECTION PACK, Item Number: DYNJT0473I, Case UDI/GTIN: 40889942954057, Unit UDI/GTIN: 10889942954056, Lot Number: 22IMH365; fffffff. C-BIRTH PACK-LF, Item Number: DYNJT1915K, Case UDI/GTIN: 40193489899376, Unit UDI/GTIN: 10193489899375, Lot Number: 22IBN792,22IBE982; ggggggg. C-SECTION PACK, Item Number: OBJ602U, Case UDI/GTIN: 40193489964470, Unit UDI/GTIN: 10193489964479, Lot Number: 22JBP919,22IBK343; hhhhhhh. THEDACARE C-SECTION, Item Number: OBP112X, Case UDI/GTIN: 40195327147649, Unit UDI/GTIN: 10195327147648, Lot Number: 22HBB733; iiiiiii. C-SECTION BASIN PACK, Item Number: OBP211C, Case UDI/GTIN: 40884389069746, Unit UDI/GTIN: 10884389069745, Lot Number: 22IBM827,22HBW292; jjjjjjj. C-SECTION PACK, Item Number: OBP323G, Case UDI/GTIN: 40193489886413, Unit UDI/GTIN: 10193489886412, Lot Number: 22IBA633,22HBI799,22HBI800; kkkkkkk. C-SECTION PACK-LF, Item Number: PHS164410001, Case UDI/GTIN: 40889942035749, Unit UDI/GTIN: 10889942035748, Lot Number: 22IMG290; lllllll. PPH C-SECTION PACK-LF, Item Number: PHS322253008B, Case UDI/GTIN: 40193489636889, Unit UDI/GTIN: 10193489636888, Lot Number: 22IBN335; mmmmmmm. C SECTION PACK-LF, Item Number: PHS410129014C, Case UDI/GTIN: 40889942936961, Unit UDI/GTIN: 10889942936960, Lot Number: 22JBC784; nnnnnnn. C-SECTION PACK, Item Number: PHS774420G, Case UDI/GTIN: 40889942942511, Unit UDI/GTIN: 10889942942510, Lot Number: 22IMC958; ooooooo. C-SECTION PACK-LF, Item Number: PHS988341C, Case UDI/GTIN: 40889942484820, Unit UDI/GTIN: 10889942484829, Lot Number: 22IMH544
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
11,886 cases (27,144 units)