FDA Enforcement Class II Ongoing

Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system

Recall: Z-1092-2022 · Reported May 25, 2022

Enforcement

Recall Number
Z-1092-2022
Event ID
90023
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Luminex Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
May 25, 2022
Initiation Date
April 5, 2022
Classification Date
May 13, 2022
Address
12212 Technology Blvd, N/A, Austin, TX, 78727-6101, United States

Description

Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system

Reason

Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.

Code Info

Serial Numbers for 10-0000-07: 17340003 19044007 20133004 18128002 18214004 12061048 18191010 13193254 13051068 21029902 16277059 19233005 11250096 14069051 18255003 *18039010 *20212006 *19317004 *19206009 *18057008 Serial Numbers added 6.23.22 for 10-0000-07R: 18191010 13051068 14069051 *12198198 *12278344 *14322267 *12271338 Serial Numbers added 7.14.22 for 10-0000-07R: 18226007 and 16242043 Serial Numbers added 7.14.22 for 10-0000-07: 17062030, 18235008 and 19192006

Distribution

US:OH, ME, PA, FL, IL, NE, IN, TX, GA, WA, MI, CO, AZ, OH, KY, CA, VT OUS: None

Quantity

24 units