Azurion 5 M12 System Code: (1) 722227 (2) 722231
Enforcement
- Recall Number
- Z-1089-2025
- Event ID
- 96120
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 12, 2025
- Initiation Date
- January 13, 2025
- Classification Date
- February 4, 2025
- Address
- Veenpluis 4-6, Best, N/A, Netherlands
Description
Azurion 5 M12 System Code: (1) 722227 (2) 722231
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
UDI: (1) 722227 00884838099227(21); (2) 722231 00884838116740(21)
Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, JAMAICA, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen.
6 units US; 246 units OUS