FDA Enforcement Class II Ongoing

MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007

Recall: Z-1083-2024 · Reported February 21, 2024

Enforcement

Recall Number
Z-1083-2024
Event ID
93651
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
SEDECAL SA
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 21, 2024
Initiation Date
October 14, 2021
Classification Date
February 14, 2024
Address
PELAYA, 9 POL., POL. IND. RIO DE JANEIRO, Algete, N/A, N/A, Spain

Description

MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007

Reason

While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.

Code Info

Model 712001 - UDI-DI 00884838028586, all serial numbers; Model 712002 - UDI-DI 00884838082373, all serial numbers; Model 712004 - UDI-DI 08436046002173, all serial numbers; Model 712006 - UDI-DI 00884838091504, serial numbers: SN20410556 and SN21410184; Model 712007 - UDI-DI)08436046002357, serial numbers: 26, 27, 19410038, 19410044, 19410046, 21410261, 21410266, 21410280, 21410288, 21410290, 21410291, 21410293, 21410294, 21410339, 21410340, 21410341, 21410342, 21410345, 22410022, 22410023, SN0410157, SN1410049, SN19410034, SN19410035, SN19410036, SN19410037, SN19410071, SN19410092, SN19410093, SN19410094, SN19410097, SN19410098, SN19410099, SN19410100, SN19410101, SN19410102, SN19410115, SN19410116, SN19410123, SN19410127, SN19410129, SN19410135, SN19410136, SN19410137, SN20410029, SN20410030, SN20410032, SN20410033, SN20410034, SN20410044, SN20410045, SN20410049, SN20410060, SN20410063, SN20410064, SN20410065, SN20410066, SN20410071, SN20410072, SN20410073, SN20410074, SN20410081, SN20410082, SN20410083, SN20410084, SN20410093, SN20410094, SN20410095, SN20410096, SN20410097, SN20410115, SN20410116, SN20410117, SN20410118, SN20410119, SN20410120, SN20410154, SN20410171, SN20410172, SN20410202, SN20410218, SN20410219, SN20410242, SN20410260, SN20410261, SN20410279, SN20410280, SN20410288, SN20410289, SN20410290, SN20410291, SN20410298, SN20410299, SN20410300, SN20410313, SN20410321, SN20410334, SN20410360, SN20410367, SN20410373, SN20410395, SN20410396, SN20410428, SN20410472, SN20410473, SN20410482, SN20410555, SN20410556, SN20410557, SN20410566, SN20410567, SN20410570, SN21410017, SN21410034, SN21410035, SN21410048, SN21410060, SN21410061, SN21410062, SN21410081, SN21410093, SN21410094, SN21410094, SN21410099, SN21410103, SN21410104, SN21410105, SN21410117, SN21410118, SN21410119, SN21410130, SN21410131, SN21410132, SN21410133, SN21410149, SN21410150, SN21410153, SN21410154, SN21410155, SN21410169, SN21410170, SN21410171, SN21410172, SN21410187, SN21410188, SN21410189, SN21410190, SN21410200, SN21410201, SN21410203, SN21410204, SN21410205, SN21410206, SN21410207, SN21410208, SN21410212, SN21410213, SN21410214, SN21410232, SNZY20410009, SNZY20410011, SNZY20410014, and SNZY204110015.

Distribution

US Nationwide distribution. There was government distribution but no military distribution.

Quantity

851 devices