FDA Enforcement Class II Ongoing

AlluraXper FD20 System Code: (1) 722006 (2) 722012 (3) 722028

Recall: Z-1080-2025 · Reported February 12, 2025

Enforcement

Recall Number
Z-1080-2025
Event ID
96120
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 12, 2025
Initiation Date
January 13, 2025
Classification Date
February 4, 2025
Address
Veenpluis 4-6, Best, N/A, Netherlands

Description

AlluraXper FD20 System Code: (1) 722006 (2) 722012 (3) 722028

Reason

Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

Code Info

UDI: (1) 722006 00884838055513(21) 00884838059115(21) 00884838059191(10) (2) 722012 00884838059054(21) (3) 722028 00884838054202(21)

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, JAMAICA, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen.

Quantity

1528 units US; 3482 units OUS