FDA Enforcement
Class II
Terminated
Siemens Artis zee Ceiling Systems. Angiographic x-ray system.
Recall: Z-1080-2014
·
Reported March 5, 2014
Enforcement
- Recall Number
- Z-1080-2014
- Event ID
- 67175
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 5, 2014
- Initiation Date
- December 16, 2013
- Classification Date
- February 24, 2014
- Termination Date
- March 9, 2015
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355, United States
Description
Siemens Artis zee Ceiling Systems. Angiographic x-ray system.
Reason
During the lifetime of Artis zee Ceiling systems, there is an increased wear due to potential sub-optimal cable guidance that can lead to cable damage.
Code Info
Artis zee Ceiling systems serial numbers from 146352 to 147617
Distribution
Nationwide Distribution.
Quantity
284