FDA Enforcement
Class II
Terminated
Juno DFR X-ray system
Recall: Z-1079-2015
·
Reported March 18, 2015
Enforcement
- Recall Number
- Z-1079-2015
- Event ID
- 70241
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Villa Radiology Systems LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 18, 2015
- Initiation Date
- December 14, 2014
- Classification Date
- March 6, 2015
- Termination Date
- March 16, 2017
- Address
- 91 Willenbrock Rd Ste B1, N/A, Oxford, CT, 06478-1036, United States
Description
Juno DFR X-ray system
Reason
The system does not provide the appropriate audible signal, permanent activation, and manual override, although the system is in high-level control functionality
Code Info
P/N 709020
Distribution
US Distribution to the states of: AZ, AL, CA, CO, CT, DC, FL, IA, IL, IN, MI, MD, MS, MA, MO, NY, OH, PA, UT, TN and TX.
Quantity
30 units distributed in the US