MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Enforcement
- Recall Number
- Z-1078-2026
- Event ID
- 98116
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Elekta, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Report Date
- January 21, 2026
- Initiation Date
- December 10, 2025
- Classification Date
- January 13, 2026
- Address
- 400 Perimeter Center Ter NE Ste 50, Atlanta, GA, 30346-1227, United States
Description
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
MOSAIQ v 3.10.200, and 3.1.1.0 up to and including v 3.2.3.1 with Particle Base. Software Versions: 3.10.200, 3.1.1.0, 3.1.2.0, 3.1.3.0, 3.1.3.1, 3.1.3.2, 3.1.3.3, 3.1.3.4, 3.2.0.0, 3.2.1.0, 3.2.1.1, 3.2.1.2, 3.2.1.3, 3.2.1.4, 3.2.2.0, 3.2.2.1, 3.2.3.0, 3.2.3.1 UDI-DI: 07340201500026, 07340201500071
US Nationwide distribution in the state of TX.