FDA Enforcement Class III Terminated

Flyte Togas are components of the Stryker Flyte System and are intended to be worn over any Stryker Flyte Helmet. This toga provides Level 1 and 4 protection according to the Association for the Advancement of Medical Instrumentation (AAMI) Liquid Barrier Performance and Classification System (PB70:2012).

Recall: Z-1077-2015 · Reported February 18, 2015

Enforcement

Recall Number
Z-1077-2015
Event ID
70332
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Stryker Instruments Div. of Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 18, 2015
Initiation Date
December 22, 2014
Classification Date
February 10, 2015
Termination Date
May 5, 2015
Address
4100 E Milham Ave, N/A, Portage, MI, 49002-9704, United States

Description

Flyte Togas are components of the Stryker Flyte System and are intended to be worn over any Stryker Flyte Helmet. This toga provides Level 1 and 4 protection according to the Association for the Advancement of Medical Instrumentation (AAMI) Liquid Barrier Performance and Classification System (PB70:2012).

Reason

A single lot (11121171) of Flyte Toga Peel-away X-Large (catalog number 0408-830-100) was labeled with an incorrect expiration date of 2412-12-01. The correct expiration date should have read 2014-12-01.

Code Info

Part/Catalog number: 0408-830-100 Lot number: 11121171 Expiration date (as printed): 2412-12-01

Distribution

Nationwide Distribution including AL, AR, AZ, IA, IL, IN, LA, MA, NC, NE, NM, SC, TN, TX, WI, and WV.

Quantity

1530 units (170 cases)