FDA Enforcement Class II Ongoing

Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.

Recall: Z-1075-2024 · Reported February 21, 2024

Enforcement

Recall Number
Z-1075-2024
Event ID
93801
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Microbiologics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 21, 2024
Initiation Date
January 8, 2024
Classification Date
February 12, 2024
Address
200 Cooper Ave N, N/A, Saint Cloud, MN, 56303-4440, United States

Description

Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.

Reason

The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.

Code Info

Lot #8184-68, Exp. 5/31/2025, UDI-DI 70845357041448. UPDATE: The recall was expanded on 1/23/2024 to include the following additional lot numbers: 8184-65, exp. 3/31/2025; 8184-66, exp. 4/30/2025; 8184-67, exp. 5/31/2025; 8184-69, exp. 6/30/2025; 8184-70, exp. 8/31/2025; and 8184-71, exp. 8/31/2025.

Distribution

US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NE, NH, NJ, NY, OH, PA, RI, TN, TX, VA, WA, WI, WV, and WY. There was government distribution and no military distribution. Foreign distribution was made to France.

Quantity

580 packages