Power PORT-A-CATH II, TRAY POWER PAC II. Labeled as: a. VEN, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); b. VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); c. VEN LP, PU, 1.9MM, UL, 6 FR/INTRO, VALVED (1/EA); d. DL, LP, PU, 2.2MM, WING-LOCK, 7FR, INTRO (1/EA); e. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO (1/EA); f. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO, VALVED (1/EA); g. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO, VALVED (1/EA)
Enforcement
- Recall Number
- Z-1074-2018
- Event ID
- 79429
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Smiths Medical ASD Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 21, 2018
- Initiation Date
- December 4, 2017
- Classification Date
- March 15, 2018
- Termination Date
- June 12, 2020
- Address
- 6000 Nathan Ln N, Minneapolis, MN, 55442-1690, United States
Description
Power PORT-A-CATH II, TRAY POWER PAC II. Labeled as: a. VEN, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); b. VEN LP, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); c. VEN LP, PU, 1.9MM, UL, 6 FR/INTRO, VALVED (1/EA); d. DL, LP, PU, 2.2MM, WING-LOCK, 7FR, INTRO (1/EA); e. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO (1/EA); f. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO, VALVED (1/EA); g. DL, LP, PU, 3.2MM, WING-LOCK, 10 FR/INTRO, VALVED (1/EA)
Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.
a. Item No. 21-4474-24 (Lot No. 87X040, 87X117); b. Item No. 21-4475-24 (Lot No. 87X044); c. Item No. 21-4477-24 (Lot No. 87X097); d. Item No. 21-8466-24 (Lot No. 87X171); e. Item No. 21-8468-24 (Lot No. 87X172, 87X244, 87X180); f. Item No. 21-8469-24 (Lot No. 87X151); g. Item No. 21-8470-24 (Lot No. 87X173, 87X254, 87X194, 87X198, 87X237)
Worldwide Distribution. US nationwide including Puerto Rico, Austria, Bahamas, Belgium, Dominican Republic, Netherlands, Spain, and Turks and Caicos.
1,077 units total