FDA Enforcement Class II Terminated

Single and Dual use Emission Computer Tomography gamma cameras (BrightView XCT, BrightView X & BrightView SPECT) Philips Healthcare Systems, Cleveland, OH. Gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies.

Recall: Z-1074-2014 · Reported March 5, 2014

Enforcement

Recall Number
Z-1074-2014
Event ID
67307
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 5, 2014
Initiation Date
December 19, 2013
Classification Date
February 24, 2014
Termination Date
January 12, 2015
Address
595 Miner Rd, Cleveland, OH, 44143-2131, United States

Description

Single and Dual use Emission Computer Tomography gamma cameras (BrightView XCT, BrightView X & BrightView SPECT) Philips Healthcare Systems, Cleveland, OH. Gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies.

Reason

Three problems were reported to the firm: When positioning for a relative 180 degree non circular acquisition the second (bottom) detector was not positioned correctly causing an inappropriate rotation orbit calculation. When the acquisition was started and the detectors began to move into position, the incorrect calculation resulted in the detectors coming in contact with the patient. As

Code Info

BrightView SPECT, Product code: 882480; Serial #'s: 400054, 4000002, 4000003, 4000004, 4000006, 4000007, 4000008, 4000009, 4000010, 4000011, 4000013, 4000029, 4000030, 4000031, 4000032, 4000033, 4000034, 4000035, 4000036, 4000037, 4000038, 4000039, 4000040, 4000041, 4000043, 4000044, 4000045, 4000046, 4000047, 4000048, 4000049, 4000050, 4000051, 4000052, 4000053, 4000054, 4000055, 4000056, 4000057, 4000058, 4000059, 4000060, 4000061, 4000062, 4000064, 4000065, 4000066, 4000067, 4000068, 4000070, 4000071, 4000072, 4000073, 4000074, 4000075, 4000076, 4000077, 4000078, 4000079, 4000080, 4000081, 4000083, 4000086, 4000087, 4000088, 4000089, 4000090, 4000091, 4000093, 4000094, 4000095, 4000099, 4000100, 4000102, 4000103, 4000104, 4000105, 4000106, 4000107, 4000108, 4000109, 4000110, 4000111, 4000112, 4000113, 4000114, 4000115, 4000116, 4000117, 4000118, 4000119, 4000120, 4000121, 4000122, 4000123, 4000124, 4000125, 4000126, 4000127, 4000128, 4000129, 4000130, 4000132, 4000133, 4000135, 4000137, 4000138, 4000139, 4000140, 4000141, 4000142, 4000143, 4000144, 4000145, 4000146, 4000148, 4000149, 4000151, 4000152, 4000153, 4000154, 4000155, 4000156, 4000157, 4000158, 4000159, 4000160, 4000161, 4000162, 4000164, 4000165, 4000166, 4000167, 4000168, 4000169, 4000170, 4000171, 4000173, 4000174, 4000175, 4000176, 4000177, 4000178, 4000179, 4000180, 4000181, 4000182, 4000183, 4000184, 4000185, 4000186, 4000187, 4000188, 4000189, 4000190, 4000191, 4000192, 4000193, 4000194, 4000195, 4000196, 4000197, 4000198, 4000199, 4000200, 4000202, 4000203, 4000204, 4000205, 4000206, 4000207, 4000208, 4000209, 4000212, 4000214, 4000215, 4000216, 4000218, 4000219, 4000220, 4000221, 4000222, 4000223, 4000224, 4000225, 4000226, 4000227, 4000228, 4000229, 4000230, 4000231, 4000232, 4000233, 4000234, 4000235, 4000236, 4000237, 4000238, 4000239, 4000240, 4000241, 4000242, 4000243, 4000244, 4000245, 4000246, 4000247, 4000248, 4000249, 4000250, 4000251, 4000252, 4000253, 4000254, 4000255, 4000256, 4000257, 4000258, 4000259, 4000260, 4000261, 4000262, 4000263, 4000264, 4000265, 4000266, 4000267, 4000268, 4000269, 4000270, 4000271, 4000272, 4000273, 4000274, 4000275, 4000276, 4000277, 4000278, 4000279, 4000280, 4000281, 4000282, 4000283, 4000284, 4000285, 4000287, 4000288, 4000289, 4000290, 4000291, 4000292, 4000293, 4000295, 4000296, 4000297, 4000298, 4000299, 4000300, 4000301, 4000302, 4000303, 4000304, 4000305, 4000306, 4000307, 4000308, 4000309, 4000310, 4000311, 4000312, 4000313, 4000314, 4000315, 4000316, 4000317, 4000318, 4000319, 4000320, 4000321, 4000322, 4000323, 4000324, 4000325, 4000326, 4000327, 4000328, 4000329, 4000330, 4000331, 4000332, 4000333, 4000334, 4000335, 4000336, 4000337, 4000338, 4000339, 4000340, 4000341, 4000342, 4000343, 4000344, 4000345, 4000346, 4000347, 4000348, 4000349, 4000351, 4000352, 4000353, 4000354, 4000355, 4000356, 4000357, 4000358, 4000359, 4000360, 4000361, 4000362, 4000363, 4000364, 4000365, 4000366, 4000367, 4000368, 4000369, 4000370, 4000371, 4000372, 4000374, 4000375, 4000376, 4000377, 4000378, 4000379, 4000380, 4000381, 4000382, 4000383, 4000384, 4000385, 4000386, 4000387, 4000388, 4000389, 4000390, 4000391, 4000392, 4000394, 4000396, 4000397, 4000398, 4000399, 4000400, 4000401, 4000402, 4000403, 4000404, 4000405, 4000406, 4000407, 4000408, 4000410, 4000411, 4000412, 4000414, 4000415, 4000416, 4000417, 4000418, 4000419, 4000420, 4000421, 4000422, 4000423, 4000424, 4000425, 4000426, 4000428, 4000429, 4000430, 4000431, 4000432, 4000433, 4000434, 4000435, 4000436, 4000437, 4000438, 4000439, 4000440, 4000441, 4000442, 4000443, 4000444, 4000445, 4000446, 4000447, 4000448, 4000449, 4000450, 4000451, 4000452, 4000453, 4000454, 4000455, 4000456, 4000459, 4000460, 4000461, 4000462, 4000463, 4000464, 4000466, 4000467, 4000468, 4000469, 4000470, 4000473, 4000474, 4000475, 4000476, 4000477, 4000478, 4000479, 4000480, 4000481, 4000482, 4000483, 4000485, 4000486, 4000487, 4000488, 4000489, 4000490, 4000491, 4000492, 4000493, 4000494, 4000495, 4000496, 4000497, 4000498, 4000499, 4000500, 4000501, 4000502, 4000503, 4000504, 4000505, 4000506, 4000507, 4000508, 4000509, 4000510, 4000511, 4000513, 4000516, 4000517, 4000518, 4000519, 4000520, 4000521, 4000522, 4000523, 4000524, 4000525, 4000526, 4000527, 4000528, 4000529, 4000530, 4000531, 4000532, 4000533, 4000534, 4000535, 4000536, 4000537, 4000538, 4000539, 4000540, 4000541, 4000542, 4000543, 4000544, 4000546, 4000547, 4000549, 4000550, 4000551, 4000552, 4000553, 4000554, 4000555, 4000556, 4000557, 4000558, 4000559, 4000560, 4000561, 4000562, 4000563, 4000564, 4000565, 4000566, 4000567, 4000568, 4000569, 4000571, 4000572, 4000573, 4000574, 4000575, 4000576, 4000577, 4000578, 4000580, 4000581, 4000582, 4000583, 4000584, 4000585, 4000586, 4000587, 4000588, 4000589, 4000590, 4000591, 4000592, 4000593, 4000594, 4000595, 4000596, 4000597, 4000598, 4000599, 4000600, 4000601, 4000602, 4000603, 4000604, 4000605, 4000606, 4000607, 4000608, 4000609, 4000610, 4000611, 4000612, 4000614, 4000615, 4000616, 4000617, 4000619, 4000620, 4000621, 4000622, 4000623, 4000624, 4000625, 4000626, 4000627, 4000628, 4000629, 4000630, 4000631, 4000632, 4000633, 4000634, 4000635, 4000636, 4000638, 4000639, 4000640, 4000642, 4000643, 4000644, 4000645, 4000646, 4000647, 4000648, 4000649, 4000650, 4000653, 4000654, 4000655, 4000656, 4000657, 4000658, 4000659, 4000660, 4000661, 4000662, 4000663, 4000664, 4000666, 4000667, 4000668, 4000669, 4000670, 4000671, 4000672, 4000673, 4000674, 4000675, 4000676, 4000677, 4000678, 4000679, 4000680, 4000681, 4000682, 4000683, 4000684, 4000685, 4000686, 4000689, 4000690, 4000692, 4000693, 4000694, 4340004, 40000101, 4000703001, 40007050005, 40007060007, 40007060008, 4000098A, 4000246A, 4000472A, 4000484A & TM1. BrightView X, Product code: 882478; Serial #'s: 11000003, 11000004, 11000005, 11000006, 11000008, 11000009, 11000011, 11000012, 11000013, 11000014, 11000015, 11000016, 11000017, 11000018, 11000019, 11000020, 11000022, 11000023, 11000024, 11000025, 11000026, 11000028, 11000029, 11000030, 11000031, 11000033, 11000034, 11000036, 11000037, 11000038, 11000039, 11000040, 11000041, 11000042, 11000043, 11000045, 11000046, 11000047, 11000048, 11000049, 11000050, 11000051, 11000052, 11000054, 11000055, 11000056, 11000057, 11000058, 11000059, 11000060, 11000061, 11000062, 11000063, 11000064, 11000065, 11000066, 11000067, 11000068, 11000069, 11000070, 11000071, 11000072, 11000073, 11000074, 11000075, 11000076, 11000077, 11000078, 11000079, 11000080, 11000083, 11000084, 11000085, 11000086, 11000087, 11000088, 11000089, 11000090, 11000091, 11000092, 11000093, 11000094, 11000095, 11000096, 11000097, 11000098, 11000099, 11000100, 11000101, 11000102, 11000103, 11000104, 11000105, 11000106, 11000107, 11000108, 11000109, 11000110, 11000111, 11000112, 11000113, 11000115, 11000116, 11000117, 11000118, 11340002, 11340004 & 11340005. BrightView XCT, Product code: 882482; Serial #'s: 6000002, 6000003, 6000004, 6000005, 6000006, 6000010, 6000011, 6000012, 6000013, 6000015, 6000016, 6000017, 6000018, 6000019, 6000020, 6000022, 6000023, 6000024, 6000025, 6000026, 6000027, 6000028, 6000029, 6000030, 6000031, 6000032, 6000033, 6000036, 6000037, 6000038, 6000040, 6000041, 6000042, 6000043, 6000044, 6000045, 6000047, 6000048, 6000049, 6000050, 6000051, 6000052, 6000053, 6000054, 6000055, 6000056, 6000058, 6000059, 6000060, 6000061, 6000062, 6000063, 6000064, 6000065, 6000066, 6000067, 6000069, 6000070, 6000071, 6000072, 6000073, 6000074, 6000076, 6000077, 6000078, 6000079, 6000080, 6000081, 6000082, 6000083, 6000084, 6000085, 6000086, 6000087, 6000088, 6000089, 6000090, 6000091, 6000092, 6000093, 6000094, 6000095, 6000096, 6000097, 6000098, 6000099, 6000100, 6000101, 6000102, 6000103, 6000104, 6000105, 6000106, 6000107, 6000108, 6000109, 6000110, 6000111, 6000112, 6000113, 6000114, 6000115, 6000116, 6000117, 6000118, 6000119, 6000120, 6000121, 6000122, 6000123, 6000125, 6000126, 6000127, 6000128, 6000129, 6000130, 6000131, 6000132, 6000133, 6000134, 6000136, 6000138, 6000139, 6000140, 6000141, 6000142, 6000143, 6000144, 6000145, 6000146, 6000147, 6000148, 6000149, 6000150, 6000151, 6000152, 6000153, 6000154, 6000155, 6000156, 6000157, 6000158, 6000159, 6000160, 6000161, 6000162, 6000163, 6000164, 6000165, 6000166, 6000167, 6000168, 6000169, 6000170, 6000171, 6000172, 6000173, 6000174, 6000175, 6000176, 6000177, 6000178, 6000179, 6000180, 6000181, 6000182, 6000183, 6000184, 6000185, 6000186, 6000187, 6000188, 6000189, 6000190, 6000191, 6000192, 6000193, 6000194, 6000195, 6000196, 6000197, 6000198, 6000199, 6000200, 6000201, 6000202, 6000203, 6000204, 6000205, 6000206, 6000208, 6000209, 6000210, 6000211, 6000212, 6000213, 6000214, 6000215, 6000216, 6000217, 6000218, 6000219, 6000220, 6000221, 6000222, 6000223, 6000224, 6000225, 6000226, 6000227, 6000228, 6000229, 6000230, 6000231, 6000232, 6000233, 6000234, 6000235, 6000236, 6000237, 6000238, 6000239, 6000240, 6000241, 6000242, 6000243, 6000244, 6000245, 6000246, 6000247, 6000249, 6000250, 6000251, 6000252, 6000253, 6000254, 6000255, 6000256, 6000257, 6000258, 6000259, 6000260, 6000262, 6000263, 6000264, 6000265, 6000266, 6000267, 6000268, 6000269, 6000270, 6000271, 6000272, 6000273, 6000274, 6000275, 6000276, 6000277, 6000278, 6000279, 6000280, 6000281, 6000282, 6000283, 6000284, 6000285, 6000286, 6000287, 6000288, 6000290, 6000291, 6000292, 6000293, 6000294, 6000295, 6000296, 6000297, 6000298, 6000299, 6000300, 6000301, 6000302, 6000303, 6000305, 6000306, 6000307, 6000308, 6000309, 6000310, 6000311, 6000312, 6000313, 6000314, 6000315, 6000316, 6000317, 6000318, 6000319, 6000320, 6000321, 6000322, 6000323, 6000324, 6000325, 6000326, 6000327, 6000328, 6000329, 6000330, 6000331, 6000332, 6000333, 6000334, 6000335, 6000336, 6000337, 6000338, 6000339, 6000340, 6000341, 6000342, 6000343, 6000344, 6000345, 6000346, 6000348, 6000349, 6000350, 6000351, 6000352, 6000353, 6000354, 6000355, 6000356, 6000357, 6000358, 6000359, 6000360, 6000361, 6000362, 6000364, 6000365, 6000366, 6000367, 6000368, 6000369, 6000370, 6000371, 6000372, 6000373, 6000374, 6000375, 6000376, 6000377, 6000378, 6000379, 6000380, 6000381, 6000382, 6000383, 6000384, 6000385, 6000386, 6000387, 6000388, 6000389, 6000390, 6000391, 6000392, 6000393, 6000394, 6000395, 6000396, 6000398, 6000399, 6000400, 6000401, 6000402, 6000403, 6000404, 6000405, 6000407, 6000408, 6000409, 6000410, 6000411, 6000412, 6000413, 6000414, 6000415, 6000416, 6000417, 6000418, 6340008, 6340009, 6340010, 6340011, 11000021, 11000082, 6000004A & IBA_RSC, P01_8128.

Distribution

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, VT, WA, WI, WV & WY, and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Gabon, Germany, Greece, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Palestine, Panama, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom & Viet Nam.

Quantity

1,143 units