PORT-A-CATH II PS/Titan, Tray PAC II. Labeled as: a. VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO (1/EA); b. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO (1/EA); c. DL, LP, PU, 3.2MM, WING-LOCK, 10FR/INTRO (1/EA)
Enforcement
- Recall Number
- Z-1073-2018
- Event ID
- 79429
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Smiths Medical ASD Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 21, 2018
- Initiation Date
- December 4, 2017
- Classification Date
- March 15, 2018
- Termination Date
- June 12, 2020
- Address
- 6000 Nathan Ln N, Minneapolis, MN, 55442-1690, United States
Description
PORT-A-CATH II PS/Titan, Tray PAC II. Labeled as: a. VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO (1/EA); b. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO (1/EA); c. DL, LP, PU, 3.2MM, WING-LOCK, 10FR/INTRO (1/EA)
Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.
a. Item No. 21-4071-24 (Lot No. 86X046, 87X050); b. Item No. 21-8066-24 (Lot No. 87X178); c. Item No. 21-8068-24 (Lot No. 87X179)
Worldwide Distribution. US nationwide including Puerto Rico, Austria, Bahamas, Belgium, Dominican Republic, Netherlands, Spain, and Turks and Caicos.
1,077 units total