FDA Enforcement Class II Ongoing

Allura Xper FD20 OR Table; Model Number: 722023;

Recall: Z-1072-2026 · Reported January 21, 2026

Enforcement

Recall Number
Z-1072-2026
Event ID
98109
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 21, 2026
Initiation Date
December 15, 2025
Classification Date
January 13, 2026
Address
Veenpluis 4-6, Best, N/A, Netherlands

Description

Allura Xper FD20 OR Table; Model Number: 722023;

Reason

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Code Info

Model Number: 722023; System Product Name: Allura Xper FD20 OR Table; UDI-DI: 00884838059085; Serial Numbers: 103, 10, 85, 125;

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.

Quantity

4 units