Allura Xper FD20; Model Number: 722012;
Enforcement
- Recall Number
- Z-1070-2026
- Event ID
- 98109
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 21, 2026
- Initiation Date
- December 15, 2025
- Classification Date
- January 13, 2026
- Address
- Veenpluis 4-6, Best, N/A, Netherlands
Description
Allura Xper FD20; Model Number: 722012;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Model Number: 722012; System Product Name: Allura Xper FD20; UDI-DI: 00884838059054; Serial Numbers: 2942, 1515, 1189, 1594, 535, 1964, 2156, 2742, 2092, 1673, 1803, 1655, 2020, 1973, 2037, 1853, 2098, 1248, 1082, 1576, 1057, 1129, 646, 930, 1421, 462, 203, 627, 2005, 1428, 353, 288, 191, 1994, 1802, 156, 853, 1283, 1956, 2361, 1980, 872, 931, 1295, 2513, 1090, 1374, 468, 1822, 812, 1379, 1817, 1580, 1050, 309, 144, 1841, 1845, 1468, 558, 1968, 1959, 2168, 1168, 1562, 967, 2114, 47, 167, 2091, 2053, 604, 603, 217, 2918, 373, 850, 1820, 1821, 1619;
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.
79 units