FDA Enforcement Class I Ongoing

VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

Recall: Z-1070-2024 · Reported March 6, 2024

Enforcement

Recall Number
Z-1070-2024
Event ID
93806
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Ventec Life Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 6, 2024
Initiation Date
January 9, 2024
Classification Date
February 23, 2024
Address
22002 26th Ave Se, Bothell, WA, 98021-4903, United States

Description

VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

Reason

The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.

Code Info

Box Label - Part Numbers: PRT-01013-000 Pouch Label - Part Number: PRT-00802-001 UDI-DI Code: 00855573007594 Lot Number: 220922 Pack Slip Lot Number: 257337

Distribution

U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan

Quantity

250 breathing circuits