FDA Enforcement
Class II
Ongoing
Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T
Recall: Z-1068-2024
·
Reported February 21, 2024
Enforcement
- Recall Number
- Z-1068-2024
- Event ID
- 93752
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- OriGen Biomedical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 21, 2024
- Initiation Date
- December 19, 2023
- Classification Date
- February 9, 2024
- Address
- 7000 Burleson Rd Bldg D, N/A, Austin, TX, 78744-3202, United States
Description
Accessory Sets Syringes: 6ml syringe with a female luer with 15cm sealed SCD tubing, REF: RF-T15, and 60ml syringe on a female luer with 15cm sealed SCD tubing, REF: 15-RF60-T
Reason
Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
Code Info
REF/UDI-DI/Lots: RF-T15/10816203020205/ V23277, V23278, V23639; 15-RF60-T/10816203020571/V23281
Distribution
US: FL, WA, NJ, MA, OH, CA, MD, TX, NE, ND OUS: GB, CA,NL,CH, PL, DE
Quantity
3613 Cases