iChem VELOCITY Automated Urine Chemistry, Catalog No. 800-3049, 800-3050, 800-3079, 800-3080, 800-3530, 800-7167, 800-7190, 800-7713, 800-7714, 800-7719, 800-7720 iChem VELOCITY Urine Chemistry Strips, Catalog No. 800-7212 & 800-7212-001
Enforcement
- Recall Number
- Z-1067-2016
- Event ID
- 73100
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 16, 2016
- Initiation Date
- January 13, 2016
- Classification Date
- March 7, 2016
- Termination Date
- August 29, 2016
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
iChem VELOCITY Automated Urine Chemistry, Catalog No. 800-3049, 800-3050, 800-3079, 800-3080, 800-3530, 800-7167, 800-7190, 800-7713, 800-7714, 800-7719, 800-7720 iChem VELOCITY Urine Chemistry Strips, Catalog No. 800-7212 & 800-7212-001
Iris International noted loose or missing analyte pads found in several locations (strip vial, strip provider module, strip conveyor system or waste container) that are undetectable by the iChemVELOCITY System.
all serial numbers
Affected geographies: Argentina, Australia, Austria, Bangladesh, Bermuda, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Canada, Chile, China, Columbia, Croatia, Czech Republic, Ecuador, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Lebanon, Lithuania, Macao, Malaysia, Mexico, Montenegro, Netherlands, New Zealand, Nigeria, Oman, Panama Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico , Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Republic, United States (including Puerto Rico), and Venezuela
2,661 units total (1,424 units in US)