FDA Enforcement
Class II
Terminated
EPWorks software used in the Protektor 32 Product Usage: Uses electroencephalography (EEG), evoked potentials (EP), electromyography (EMG) and transcranial motor evoked potentials (TcMEP) stimulation techniques to provide healthcare professionals with information to help assess patients neurological status during surgery.
Recall: Z-1067-2015
·
Reported February 18, 2015
Enforcement
- Recall Number
- Z-1067-2015
- Event ID
- 70157
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Natus Medical Incorporated
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- February 18, 2015
- Initiation Date
- January 2, 2015
- Classification Date
- February 9, 2015
- Termination Date
- March 22, 2017
- Address
- 2568 Bristol Circle, N/A, Oakville, Ontario, N/A, Canada
Description
EPWorks software used in the Protektor 32 Product Usage: Uses electroencephalography (EEG), evoked potentials (EP), electromyography (EMG) and transcranial motor evoked potentials (TcMEP) stimulation techniques to provide healthcare professionals with information to help assess patients neurological status during surgery.
Reason
Software error occurs when using remote monitoring; if the remote user tries to stop the free run waveform group, the system will display a message informing the user that they do not have sufficient privilege.
Code Info
EPWorks Software Versions prior to 6.0
Distribution
US Nationwide Distribution
Quantity
248 units