FDA Enforcement
Class II
Ongoing
11274BCU1 11274BCUK1 Flexible Cystoscope Z18449US-BD (08-2018)
Recall: Z-1066-2022
·
Reported May 18, 2022
Enforcement
- Recall Number
- Z-1066-2022
- Event ID
- 89880
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Karl Storz Endoscopy
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- May 18, 2022
- Initiation Date
- April 1, 2022
- Classification Date
- May 12, 2022
- Address
- 2151 E Grand Ave, N/A, El Segundo, CA, 90245-5017, United States
Description
11274BCU1 11274BCUK1 Flexible Cystoscope Z18449US-BD (08-2018)
Reason
Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.
Code Info
11274BCU1 11274BCUK1 Flexible Cystoscope Z18449US-BD (08-2018) UDI Code: 4048551408155 All Serial Numbers manufactured/distributed since January 2018.
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA,WI, WV, and WY. O.U.S.: Mexico, Lebanon, and South Korea
Quantity
All Serial Numbers manufactured/distributed since January 2018.