FDA Enforcement Class II Terminated

BVI Visitec Soft Tip Cannula, 0.8MM .50 x 33mm (25G x 1.3 in). The BVI Soft Tip Cannula is intended to irrigate and/or aspirate ophthalmic solutions and liquids in or out of the eye, and remove materials where appropriate (e.g. retinal, subretinal procedures).

Recall: Z-1066-2019 · Reported April 3, 2019

Enforcement

Recall Number
Z-1066-2019
Event ID
81757
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beaver Visitec
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 3, 2019
Initiation Date
November 30, 2018
Classification Date
March 27, 2019
Termination Date
May 1, 2020
Address
500 Totten Pond Rd, 10 City Point, Waltham, MA, 02451-1916, United States

Description

BVI Visitec Soft Tip Cannula, 0.8MM .50 x 33mm (25G x 1.3 in). The BVI Soft Tip Cannula is intended to irrigate and/or aspirate ophthalmic solutions and liquids in or out of the eye, and remove materials where appropriate (e.g. retinal, subretinal procedures).

Reason

Certain lots are missing a protective sheath component over the silicone tip, which may result in the silicone tip breaking or detaching at the metal tube component. If the tip becomes detached during a procedure while in the eye, the fragment will need to be retrieved to mitigate the possibility of injury.

Code Info

Lots 60047S4, 60060S9, 600623S

Distribution

Worldwide distribution: US (nationwide) to states of: AR, AZ, CA, CO, FL, GA, LA, MD, MO, NC, NE, NJ, NY, OH, OK, TX, UT, VA, WA; and countries of: Mexico and Peru.

Quantity

3430 units