FDA Enforcement Class II Terminated

SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, and 4.4.0.0

Recall: Z-1065-2015 · Reported February 18, 2015

Enforcement

Recall Number
Z-1065-2015
Event ID
69837
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
SCC Soft Computer
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 18, 2015
Initiation Date
November 26, 2014
Classification Date
February 9, 2015
Termination Date
May 16, 2016
Address
5400 Tech Data Drive, N/A, Clearwater, FL, 33760, United States

Description

SoftPath Laboratory Information System. Versions 4.3.0.8, 4.3.0.9, 4.3.0.10, 4.3.0.11, 4.3.0.12, 4.3.0.14, 4.3.0.15, and 4.4.0.0

Reason

Modifications to diagnostic text may be: 1) Saved to the database but not appear on the report sent to the physician; or 2) Documented on the report, but not saved to the database.

Code Info

Version 4.3.0.8 Release date: 07/31/2009 - Present; Version 4.3.0.9 Release date: 11/23/2009 - Present; Version 4.3.0.10 Release date: 09/20/2010- Present; Version 4.3.0.11 Release date: 11/11/2010- Present; Version 4.3.0.12 Release date: 02/07/2011- Present; Version 4.3.0.14 Release date: 11/14/2012- Present; Version 4.3.0.15 Release date: 09/17/2013- Present; Version 4.4.0.1 Release date: 12/13/2013 - Present;

Distribution

Distribution US Nationwide and CANADA.

Quantity

108 units