FDA Enforcement Class II Terminated

CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.

Recall: Z-1062-2015 · Reported February 18, 2015

Enforcement

Recall Number
Z-1062-2015
Event ID
69863
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Maquet Datascope Corp - Cardiac Assist Division
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 18, 2015
Initiation Date
December 16, 2014
Classification Date
February 6, 2015
Termination Date
June 9, 2017
Address
1300 MacArthur Blvd., N/A, Mahwah, NJ, 07430-2052, United States

Description

CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.

Reason

Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit and may become dislodged and remain in IABPs fiber-optic receptacle.

Code Info

Sensation 7 fr.34 cc IAB Demonstration Unit 0684-DC-0433 0684-DO-0433 0684-DC-0469 0684-DO-0469 Sensation 7 Fr. 40 cc IAB Demonstration Unit 0684-DC-0434 0684-DO-0434 0684-DC-0470 0684-DO-0470 Sesnsation Plus 7.5 Fr. 40cc IAB Demonstration Units 0684-DC-0567 0684-DO-0567 0684-DC-0568 0684-DO-0568 Sensation Plus 7.5 Fr. 50cc IAB Demonstration Units 0684-DC-0575 0684-DO-0575 0684-DC-0576 0684-DO-0576 Sensation Plus 8 Fr. 50 cc IAB Demonstration Unit 0684-DC-0272 0684-DO-0272

Distribution

Worldwide Distribution

Quantity

838 units-US; 364 units - OUS