FDA Enforcement
Class I
Ongoing
AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8005
Recall: Z-1061-2024
·
Reported March 6, 2024
Enforcement
- Recall Number
- Z-1061-2024
- Event ID
- 93821
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Vyaire Medical
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 6, 2024
- Initiation Date
- January 10, 2024
- Classification Date
- February 24, 2024
- Address
- 26125 N Riverwoods Blvd, Mettawa, IL, 60045-3420, United States
Description
AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8005
Reason
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
Code Info
UDI/DI Case: 50190752114157, Each: 10190752114159; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.
Distribution
Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
Quantity
3,683,407 units