FDA Enforcement Class II Terminated

ZUK PRC FEM SZB RMED/LLAT¿ ZUK PRC FEM SZC LMED/RLAT¿ ZUK PRC FEM SZC RMED/LLAT¿ NGU PRC FEM SZD LMED/RLAT¿ ZUK PRC FEM SZD RMED/LLAT¿ ZUK PRC FEM SZE LMED/RLAT ZUK PRC FEM SZF RMED/LLAT " These devices are indicated for patients with: - Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis. - Previous tibial condyle or plateau fractures with loss of anatomy or function. - Varus or valgus deformities. - Revision of previous arthroplasty procedures. " These devices are indicated for cemented use only. " The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees.

Recall: Z-1054-2018 · Reported March 21, 2018

Enforcement

Recall Number
Z-1054-2018
Event ID
78706
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer Biomet, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 21, 2018
Initiation Date
November 29, 2017
Classification Date
March 14, 2018
Termination Date
April 2, 2020
Address
1800 W Center St, Warsaw, IN, 46580-2304, United States

Description

ZUK PRC FEM SZB RMED/LLAT¿ ZUK PRC FEM SZC LMED/RLAT¿ ZUK PRC FEM SZC RMED/LLAT¿ NGU PRC FEM SZD LMED/RLAT¿ ZUK PRC FEM SZD RMED/LLAT¿ ZUK PRC FEM SZE LMED/RLAT ZUK PRC FEM SZF RMED/LLAT " These devices are indicated for patients with: - Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis. - Previous tibial condyle or plateau fractures with loss of anatomy or function. - Varus or valgus deformities. - Revision of previous arthroplasty procedures. " These devices are indicated for cemented use only. " The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees.

Reason

The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.

Code Info

00-5842-012-02 61777606 00-5842-012-02 61774093 00-5842-013-01 61774095 00-5842-013-02 61777611 00-5842-014-01 11002921 00-5842-014-01 61777613 00-5842-014-02 61777615 00-5842-014-02 61777616 00-5842-015-01 61777617 00-5842-016-02 61756843

Distribution

Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.

Quantity

13,227 in total