Giraffe OmniBed¿ and Giraffe Incubator p/n 6600-0723-700 (rev A and B) labeled in part: ***GE Healthcare 8880 Gorman Road Laurel, MD 20723 U.S.A. www.gehealthcare.com***. GE Healthcare Giraffe Omnibed is a combination of an infant incubator and an infant warmer.
Enforcement
- Recall Number
- Z-1051-2013
- Event ID
- 64189
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- GE Healthcare, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 17, 2013
- Initiation Date
- January 8, 2013
- Classification Date
- April 5, 2013
- Termination Date
- July 10, 2013
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615, United States
Description
Giraffe OmniBed¿ and Giraffe Incubator p/n 6600-0723-700 (rev A and B) labeled in part: ***GE Healthcare 8880 Gorman Road Laurel, MD 20723 U.S.A. www.gehealthcare.com***. GE Healthcare Giraffe Omnibed is a combination of an infant incubator and an infant warmer.
Neonatal beds and incubator display boards may randomly overwrite stored information and radiant heater and oxygen set points upon power up or recovery from a power failure.
OmniBed Serial Numbers: HDGR52546, HDGR52736, HDGR52802, HDGR52861, HDGR52910, HDGR52958, HDGR53007, HDGR53058, HDGR53110, HDGR53185, HDGR52547, HDGR52737, HDGR52803, HDGR52862, HDGR52911, HDGR52959, HDGR53008, HDGR53059, HDGR53112, HDGR53186, HDGR52558, HDGR52739, HDGR52804, HDGR52863, HDGR52912, HDGR52960, HDGR53009, HDGR53060, HDGR53113, HDGR53187, HDGR52609, HDGR52740, HDGR52805, HDGR52864, HDGR52913, HDGR52961, HDGR53010, HDGR53061, HDGR53114, HDGR53188, HDGR52610, HDGR52741, HDGR52806, HDGR52865, HDGR52914, HDGR52962, HDGR53011, HDGR53062, HDGR53115, HDGR53189, HDGR52613, HDGR52742, HDGR52813, HDGR52866, HDGR52915, HDGR52963, HDGR53012, HDGR53063, HDGR53116, HDGR53190, HDGR52620, HDGR52743, HDGR52814, HDGR52867, HDGR52916, HDGR52964, HDGR53013, HDGR53064, HDGR53117, HDGR53191, HDGR52621, HDGR52744, HDGR52815, HDGR52868, HDGR52917, HDGR52965, HDGR53014, HDGR53065, HDGR53119, HDGR53192, HDGR52622, HDGR52745, HDGR52816, HDGR52869, HDGR52918, HDGR52966, HDGR53015, HDGR53066, HDGR53120, HDGR53193, HDGR52623, HDGR52746, HDGR52817, HDGR52870, HDGR52919, HDGR52967, HDGR53016, HDGR53067, HDGR53125, HDGR53194, HDGR52624, HDGR52747, HDGR52818, HDGR52871, HDGR52920, HDGR52968, HDGR53018, HDGR53068, HDGR53126, HDGR53195, HDGR52626, HDGR52748, HDGR52820, HDGR52872, HDGR52921, HDGR52969, HDGR53019, HDGR53069, HDGR53127, HDGR53196, HDGR52657, HDGR52749, HDGR52821, HDGR52873, HDGR52922, HDGR52970, HDGR53021, HDGR53070, HDGR53128, HDGR53197, HDGR52658, HDGR52750, HDGR52822, HDGR52874, HDGR52923, HDGR52971, HDGR53022, HDGR53071, HDGR53129, HDGR53213, HDGR52664, HDGR52751, HDGR52823, HDGR52875, HDGR52924, HDGR52972, HDGR53023, HDGR53072, HDGR53134, HDGR53214, HDGR52665, HDGR52752, HDGR52824, HDGR52876, HDGR52925, HDGR52973, HDGR53024, HDGR53073, HDGR53135, HDGR53215, HDGR52668, HDGR52753, HDGR52825, HDGR52877, HDGR52926, HDGR52974, HDGR53025, HDGR53074, HDGR53136, HDHR53216, HDGR52669, HDGR52754, HDGR52827, HDGR52878, HDGR52927, HDGR52975, HDGR53026, HDGR53075, HDGR53137, HDGR53217, HDGR52670, HDGR52755, HDGR52828, HDGR52879, HDGR52928, HDGR52976, HDGR53027, HDGR53076, HDGR53138, HDGR53219, HDGR52671, HDGR52756, HDGR52829, HDGR52880, HDGR52929, HDGR52977, HDGR53028, HDGR53077, HDGR53139, HDGR53220, HDGR52672, HDGR52757, HDGR52830, HDGR52881, HDGR52930, HDGR52978, HDGR53029, HDGR53078, HDGR53140, HDGR53221, HDGR52673, HDGR52758, HDGR52831, HDGR52882, HDGR52931, HDGR52979, HDGR53030, HDGR53079, HDGR53141, HDGR53222, HDGR52674, HDGR52759, HDGR52832, HDGR52883, HDGR52932, HDGR52980, HDGR53031, HDGR53080, HDGR53142, HDGR53223, HDGR52675, HDGR52760, HDGR52833, HDGR52884, HDGR52933, HDGR52981, HDGR53032, HDGR53081, HDGR53145, HDGR53224, HDGR52678, HDGR52761, HDGR52835, HDGR52885, HDGR52934, HDGR52982, HDGR53033, HDGR53085, HDGR53146, HDGR53225, HDGR52682, HDGR52762, HDGR52836, HDGR52886, HDGR52935, HDGR52983, HDGR53034, HDGR53086, HDGR53147, HDGR53226, HDGR52683, HDGR52763, HDGR52839, HDGR52887, HDGR52936, HDGR52984, HDGR53035, HDGR53087, HDGR53152, HDGR53231, HDGR52684, HDGR52764, HDGR52840, HDGR52888, HDGR52937, HDGR52985, HDGR53036, HDGR53088, HDGR53153, HDGR53232, HDGR52685, HDGR52765, HDGR52841, HDGR52890, HDGR52938, HDGR52986, HDGR53037, HDGR53089, HDGR53154, HDGR53235, HDGR52686, HDGR52768, HDGR52842, HDGR52891, HDGR52939, HDGR52987, HDGR53038, HDGR53090, HDGR53155, HDGR53241, HDGR52687, HDGR52769, HDGR52843, HDGR52892, HDGR52940, HDGR52988, HDGR53039, HDGR53091, HDGR53156, HDGR53242, HDGR52700, HDGR52770, HDGR52844, HDGR52893, HDGR52941, HDGR52990, HDGR53040, HDGR53092, HDGR53157, HDGR53251, HDGR52709, HDGR52771, HDGR52845, HDGR52894, HDGR52942, HDGR52991, HDGR53041, HDGR53093, HDGR53158, HDGR52714, HDGR52775, HDGR52846, HDGR52895, HDGR52943, HDGR52992, HDGR53042, HDGR53094, HDGR53159, HDGR52718, HDGR52776, HDGR52847, HDGR52896, HDGR52944, HDGR52993, HDGR53043, HDGR53095, HDGR53160, HDGR52720, HDGR52777, HDGR52848, HDGR52897, HDGR52945, HDGR52994, HDGR53044, HDGR53096, HDGR53161, HDGR52721, HDGR52778, HDGR52849, HDGR52898, HDGR52946, HDGR52995, HDGR53045, HDGR53097, HDGR53162, HDGR52723, HDGR52779, HDGR52850, HDGR52899, HDGR52947, HDGR52996, HDGR53047, HDGR53098, HDGR53163, HDGR52726, HDGR52780, HDGR52851, HDGR52900, HDGR52948, HDGR52997, HDGR53048, HDGR53099, HDGR53164, HDGR52727, HDGR52781, HDGR52852, HDGR52901, HDGR52949, HDGR52998, HDHR53049, HDGR53100, HDGR53165, HDGR52728, HDGR52782, HDGR52853, HDGR52902, HDHR52950, HDGR52999, HDGR53050, HDGR53102, HDGR53166, HDGR52729, HDGR52789, HDGR52854, HDGR52903, HDGR52951, HDGR53000, HDGR53051, HDGR53103, HDGR53167, HDGR52730, HDGR52790, HDGR52855, HDGR52904, HDGR52952, HDGR53001, HDGR53052, HDGR53104, HDGR53168, HDGR52731, HDGR52791, HDGR52856, HDGR52905, HDGR52953, HDGR53002, HDGR53053, HDGR53105, HDGR53169, HDGR52732, HDGR52792, HDGR52857, HDGR52906, HDGR52954, HDGR53003, HDGR53054, HDGR53106, HDGR53171, HDGR52733, HDGR52793, HDGR52858, HDGR52907, HDGR52955, HDGR53004, HDGR53055, HDGR53107, HDGR53175, HDGR52734, HDGR52794, HDGR52859, HDGR52908, HDGR52956, HDGR53005, HDGR53056, HDGR53108, HDGR53176, HDGR52735, HDGR52799, HDGR52860, HDGR52909, HDGR52957, HDGR53006, HDGR53057, HDGR53109, and HDGR53181. ***Incubator Serial Numbers: HDHR52549, HDHR52708, HDHR52774, HDHR52881, HDHR53029, HDHR52551, HDHR52709, HDHR52830, HDHR52882, HDHR53030, HDHR52552, HDHR52710, HDHR52831, HDHR52883, HDHR53031, HDHR52553, HDHR52711, HDHR52832, HDHR52884, HDHR53032, HDHR52554, HDHR52720, HDHR52833, HDHR52885, HDHR53039, HDHR52555, HDHR52721, HDHR52834, HDHR52894, HDHR53040, HDHR52556, HDHR52722, HDHR52835, HDHR52900, HDHR53041, HDHR52557, HDHR52723, HDHR52836, HDHR52901, HDHR53042, HDHR52559, HDHR52724, HDHR52837, HDHR52902, HDHR53043, HDHR52561, HDHR52725, HDHR52838, HDHR52903, HDHR53044, HDHR52562, HDHR52726, HDHR52839, HDHR52904, HDHR53045, HDHR52563, HDHR52727, HDHR52840, HDHR52905, HDHR53046, HDHR52564, HDHR52728, HDHR52841, HDHR52906, HDHR53047, HDHR52565, HDHR52729, HDHR52842, HDHR52907, HDHR53048, HDHR52567, HDHR52730, HDHR52843, HDHR52908, HDHR53049, HDHR52568, HDHR52731, HDHR52844, HDHR52909, HDHR53030, HDHR52569, HDHR52732, HDHR52845, HDHR52951, HDHR52910, HDHR52572, HDHR52733, HDHR52847, HDHR52911, HDHR53052, HDHR52573, HDHR52734, HDHR52848, HDHR52912, HDHR53053, HDHR52575, HDHR52735, HDHR52849, HDHR52913, HDHR53054, HDHR52576, HDHR52736, HDHR52850, HDHR52914, HDHR53055, HDHR52577, HDHR52737, HDHR52851, HDHR52915, HDHR53059, HDHR52578, HDHR52738, HDHR52852, HDHR52926, HDHR53062, HDHR52579, HDHR52739, HDHR52853, HDHR52927, HDHR52580, HDHR52740, HDHR52854, HDHR52928, HDHR52581, HDHR52741, HDHR52855, HDHR52929, HDHR52588, HDHR52742, HDHR52857, HDHR52971, HDHR52589, HDHR52743, HDHR52858, HDHR52972, HDHR52650, HDHR52744, HDHR52859, HDHR52974, HDHR52653, HDHR52745, HDHR52860, HDHR52975, HDHR52654, HDHR52746, HDHR52861, HDHR52976, HDHR52689, HDHR52747, HDHR52862, HDHR52977, HDHR52690, HDHR52748, HDHR52863, HDHR52978, HDHR52691, HDHR52749, HDHR52864, HDHR52979, HDHR52692, HDHR52750, HDHR52865, HDHR52981, HDHR52693, HDHR52751, HDHR52866, HDHR52982, HDHR52694, HDHR52752, HDHR52867, HDHR52983, HDHR52695, HDHR52753, HDHR52868, HDHR52985, HDHR52696, HDHR52754, HDHR52869, HDHR52986, HDHR52697, HDHR52759, HDHR52870, HDHR52987, HDHR52698, HDHR52760, HDHR52871, HDHR52988, HDHR52699, HDHR52767, HDHR52872, HDHR52989, HDHR52700, HDHR52768, HDHR52873, HDHR52990, HDHR52701, HDHR52769, HDHR52874, HDHR52991, HDHR52702, HDHR52770, HDHR52875, HDHR52992, HDHR52703, HDHR52771, HDHR52876, HDHR52993, HDHR52704, HDHR52772, HDHR52877, HDHR52994, HDHR52707, HDHR52773, HDHR52880, and HDHR52995.
Worldwide Distribution -- USA in the states of AR ,CA, CO, DC, FL, GA, HI, ID, IL, IN, KS, MA, MD, MI, MO, MS, NC, NE, NJ, NY, NV, OH, PA, TN, TX, VA, WA, WI, and WV, and the countries of Arab Emirates, Albania, Australia, Belgium, Brazil, Canada, Chile, China, Croatia, Czech., Denmark, Ecuador, Estonia, Finland, France, Germany, Hungary, Italy ,Korea, Kurdistan, Lebanon, Macedonia, Malaysia, Mexico, Norway, Oman, Poland, Russia, San Salvador, Saudi Arabia, Singapore, South Africa, Sweden, Turkey, Uruguay, Venezuela, and United Kingdom.
679 units