NovaGuide Intelligent Ultrasound System, Model: NA-RBTSYS1. Comprised of NovaBot Automated Headset (AKA: NeuralBot, Model: NA-RBT1) and Lucid TCD 2.0 Transcranial Doppler Ultrasound (Model: NA-LUCM1)

Recall: Z-1050-2023 · Reported February 8, 2023

Enforcement

Recall Number
Z-1050-2023
Event ID
91467
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
NovaSignal Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 8, 2023
Initiation Date
December 16, 2022
Classification Date
February 1, 2023
Address
2440 S Sepulveda Blvd, N/A, Los Angeles, CA, 90064-1784, United States

Description

NovaGuide Intelligent Ultrasound System, Model: NA-RBTSYS1. Comprised of NovaBot Automated Headset (AKA: NeuralBot, Model: NA-RBT1) and Lucid TCD 2.0 Transcranial Doppler Ultrasound (Model: NA-LUCM1)

Reason

Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.

Code Info

System (Tablet)UDI-DI/ Serial Numbers: 00850681007443/ 22040001, 00850681007443/ 22040002, 00850681007443/21021223 (00850681007221/ 21090003), 00850681007443/ 22064227 (00850681007221/ 21090004)

Distribution

US Nationwide and Worldwide Distribution: AZ, TX, SC, PA, IL. OUS: CA, SE, NL

Quantity