FDA Enforcement Class III Terminated

Siemens Dimension(R) EXL(TM) 200 (Siemens Material Number 10636929). Intended to measure a variety of analytes in human body fluids.

Recall: Z-1048-2013 · Reported April 10, 2013

Enforcement

Recall Number
Z-1048-2013
Event ID
64570
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 10, 2013
Initiation Date
February 27, 2013
Classification Date
April 1, 2013
Termination Date
January 14, 2014
Address
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States

Description

Siemens Dimension(R) EXL(TM) 200 (Siemens Material Number 10636929). Intended to measure a variety of analytes in human body fluids.

Reason

Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h

Code Info

Siemens Material Number 10636929

Distribution

Worldwide Distribution.

Quantity

2207 total