FDA Enforcement
Class I
Ongoing
regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.
Recall: Z-1047-2025
·
Reported February 12, 2025
Enforcement
- Recall Number
- Z-1047-2025
- Event ID
- 96094
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- ROi CPS LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 12, 2025
- Initiation Date
- December 26, 2024
- Classification Date
- February 4, 2025
- Address
- 3000 E Sawyer Rd, Republic, MO, 65738-2758, United States
Description
regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.
Reason
There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..
Code Info
Lot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, UDI (01)10194717119074(17)260831(10)103486.
Distribution
Distribution was made to MO. There was no government/military/foreign distribution.
Quantity
414 kits