FDA Enforcement
Class II
Ongoing
MEDLINE UNITE FOOT & ANKLE DRILL BIT, 1.3MM Solid Core, REF MPN10013
Recall: Z-1047-2023
·
Reported February 8, 2023
Enforcement
- Recall Number
- Z-1047-2023
- Event ID
- 91491
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- MEDLINE INDUSTRIES, LP - Northfield
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 8, 2023
- Initiation Date
- December 22, 2022
- Classification Date
- February 1, 2023
- Address
- 3 Lakes Dr, N/A, Northfield, IL, 60093-2753, United States
Description
MEDLINE UNITE FOOT & ANKLE DRILL BIT, 1.3MM Solid Core, REF MPN10013
Reason
The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.
Code Info
GTIN 10888277699113, Lot Number ?
Distribution
US Nationwide distribution.
Quantity
190 units