FDA Enforcement
Class III
Terminated
Siemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). Intended to measure a variety of analytes in human body fluids.
Recall: Z-1047-2013
·
Reported April 10, 2013
Enforcement
- Recall Number
- Z-1047-2013
- Event ID
- 64570
- Classification
- Class III
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 10, 2013
- Initiation Date
- February 27, 2013
- Classification Date
- April 1, 2013
- Termination Date
- January 14, 2014
- Address
- 500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States
Description
Siemens Dimension(R) EXL(TM) LM STM (Siemens Material Number 106393377). Intended to measure a variety of analytes in human body fluids.
Reason
Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h
Code Info
Siemens Material Number 106393377
Distribution
Worldwide Distribution.
Quantity
2207 total