FDA Enforcement Class II Ongoing

MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008

Recall: Z-1045-2023 · Reported February 8, 2023

Enforcement

Recall Number
Z-1045-2023
Event ID
91491
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
MEDLINE INDUSTRIES, LP - Northfield
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 8, 2023
Initiation Date
December 22, 2022
Classification Date
February 1, 2023
Address
3 Lakes Dr, N/A, Northfield, IL, 60093-2753, United States

Description

MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008

Reason

The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.

Code Info

GTIN 10193489120387, Lot Number 810304

Distribution

US Nationwide distribution.

Quantity

420 units