FDA Enforcement Class II Terminated

Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.

Recall: Z-1043-2017 · Reported January 25, 2017

Enforcement

Recall Number
Z-1043-2017
Event ID
76048
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
The Binding Site Group, Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 25, 2017
Initiation Date
July 5, 2016
Classification Date
January 18, 2017
Termination Date
July 3, 2017
Address
8 Calthorpe Road, Birmingham, N/A, United Kingdom

Description

Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.

Reason

Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.

Code Info

Lot #405566, 405567

Distribution

U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA. Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium.

Quantity

809 kits