FDA Enforcement
Class II
Terminated
Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.
Recall: Z-1043-2017
·
Reported January 25, 2017
Enforcement
- Recall Number
- Z-1043-2017
- Event ID
- 76048
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- The Binding Site Group, Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 25, 2017
- Initiation Date
- July 5, 2016
- Classification Date
- January 18, 2017
- Termination Date
- July 3, 2017
- Address
- 8 Calthorpe Road, Birmingham, N/A, United Kingdom
Description
Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.
Reason
Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.
Code Info
Lot #405566, 405567
Distribution
U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA. Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium.
Quantity
809 kits