3.5 COMP PLT, 4 HOLE 3.5 COMP PLT, 5 HOLE 3.5 COMP PLT, 6 HOLE 3.5 COMP PLT, 8 HOLE 3.5 COMP PLT, 9 HOLE 3.5 COMP PLT, 10 HOLE
Enforcement
- Recall Number
- Z-1043-2014
- Event ID
- 66500
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 26, 2014
- Initiation Date
- October 11, 2013
- Classification Date
- February 20, 2014
- Termination Date
- April 14, 2017
- Address
- 345 E Main St, Warsaw, IN, 46580-2746, United States
Description
3.5 COMP PLT, 4 HOLE 3.5 COMP PLT, 5 HOLE 3.5 COMP PLT, 6 HOLE 3.5 COMP PLT, 8 HOLE 3.5 COMP PLT, 9 HOLE 3.5 COMP PLT, 10 HOLE
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
61270590 61270591 61450581 61270594 61270595 61270596
Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
7,044,680 total