FDA Enforcement Class II Terminated

Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient hemodynamic data in the Philips Interventional Hemodynamic System.

Recall: Z-1041-2022 · Reported May 18, 2022

Enforcement

Recall Number
Z-1041-2022
Event ID
89878
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips North America Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 18, 2022
Initiation Date
February 22, 2022
Classification Date
May 6, 2022
Termination Date
November 6, 2024
Address
222 Jacobs St, N/A, Cambridge, MA, 02141-2289, United States

Description

Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient hemodynamic data in the Philips Interventional Hemodynamic System.

Reason

When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.

Code Info

UDI: B-00884838084544; System Serial Numbers: 396 411 225 351 51 46 33 189 132 19 51 924 644 645 646 376 384 487 538 886 14 15 643 1194 220 1238 947 949 950 951 89 657 169 168 413 414 624 565 1060 393 512 278 818 362 123 802 395 576 932 291 834 539 605 161 970 686 761 229 230

Distribution

Devices were distributed to the following US states: FL, ID, NJ, OK, SC and TX. Devices were distributed to the following foreign countries: France, Germany, India, Mexico, Netherlands, New Zealand, Norway, Poland, South Africa, and Switzerland.

Quantity

77