FDA Enforcement Class II Ongoing

Turbett Surgical Container, TS1200: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider

Recall: Z-1037-2023 · Reported February 8, 2023

Enforcement

Recall Number
Z-1037-2023
Event ID
91314
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Turbett Surgical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
February 8, 2023
Initiation Date
December 5, 2022
Classification Date
January 30, 2023
Address
668b Phillips Rd Ofc 1, N/A, Victor, NY, 14564-9741, United States

Description

Turbett Surgical Container, TS1200: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider

Reason

Pod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries.

Code Info

GTIN: 00859464007281 Lot Number: All Lot Numbers Serial Number: All Serial Numbers

Distribution

US Nationwide distribution in the states of CA, IA, IN, NC, NH, NY, OH, LA, SD, TX, VA, WI.

Quantity

61 units