CPT DEGREE TAPER (various sizes) and FEMORAL HD (various sizes)
Enforcement
- Recall Number
- Z-1036-2014
- Event ID
- 66500
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- February 26, 2014
- Initiation Date
- October 11, 2013
- Classification Date
- February 20, 2014
- Termination Date
- April 14, 2017
- Address
- 345 E Main St, Warsaw, IN, 46580-2746, United States
Description
CPT DEGREE TAPER (various sizes) and FEMORAL HD (various sizes)
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
60275282 60089260 60145841 60147718 60156456 60089259 60156459 60200116 62135079 62079190 60732916 60638837 62094973 61887194 60457227
Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.
7,044,680 total