FDA Enforcement Class II Terminated

Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.

Recall: Z-1034-2015 · Reported February 11, 2015

Enforcement

Recall Number
Z-1034-2015
Event ID
69299
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Animas Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
February 11, 2015
Initiation Date
September 6, 2011
Classification Date
January 30, 2015
Termination Date
August 10, 2015
Address
200 Lawrence Dr, N/A, West Chester, PA, 19380-3428, United States

Description

Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.

Reason

Calibration factors in the pump overwritten during a programming step. The force sensor could send a lower signal value to the pump processor, with loss of prime warnings, occlusion alarms and the pump unable to detect a cartridge during the prime sequence. Field action initiated 8/29/2011.

Code Info

Model Number(s): 100515-63 100510-63 100514-63 100512-63 100511-63 101200-03 100201-03 101202-03 101204-03 101205-03 101200-53 101202-53 101205-53 101200-02 101200-57 101201-57 101202-57 101204-57 101205-57 101200-63 101201-63 101202-63 101204-63 101205-63 101206-63

Distribution

No US distribution, Distributors are located in France, Germany, Sweden and United Kingdom.

Quantity

1235