FDA Enforcement Class II Terminated

Fresenius 2008K@home Series : Hemodialysis System

Recall: Z-1029-2017 · Reported January 25, 2017

Enforcement

Recall Number
Z-1029-2017
Event ID
76037
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Fresenius Medical Care Renal Therapies Group, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
January 25, 2017
Initiation Date
December 21, 2016
Classification Date
January 13, 2017
Termination Date
March 17, 2023
Address
920 Winter St, Waltham, MA, 02451-1521, United States

Description

Fresenius 2008K@home Series : Hemodialysis System

Reason

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

Code Info

All serial numbers

Distribution

Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam

Quantity

1400 units (US) 556 units (OUS)