FDA Enforcement
Class II
Terminated
Fresenius 2008K2 Series : Hemodialysis System
Recall: Z-1028-2017
·
Reported January 25, 2017
Enforcement
- Recall Number
- Z-1028-2017
- Event ID
- 76037
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 25, 2017
- Initiation Date
- December 21, 2016
- Classification Date
- January 13, 2017
- Termination Date
- March 17, 2023
- Address
- 920 Winter St, Waltham, MA, 02451-1521, United States
Description
Fresenius 2008K2 Series : Hemodialysis System
Reason
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
Code Info
All serial numbers
Distribution
Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
Quantity
60,151 units (US) 1773 (OUS)