Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBP Module, REF: 01-2017. Used with User/Operator Manual
Enforcement
- Recall Number
- Z-1027-2026
- Event ID
- 97889
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Remote Diagnostic Technologies Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- January 14, 2026
- Initiation Date
- November 26, 2025
- Classification Date
- January 8, 2026
- Address
- Farnborough Aerospace Centre, Aerospace Boulevard, Farnborough, N/A, United Kingdom
Description
Philips Tempus Pro Patient Monitor, REF: 00-1004, 00-1007, 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R. USB 2-Channel IBP Module, REF: 01-2017. Used with User/Operator Manual
Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated measurements, this could potentially lead to incorrect clinical decisions.
REF/UDI-DI: 00-1004/05060472440624; 00-1007/05060472440655; 00-1004-R/05060472440020, 05060472442901; 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932; User/Operator Manual Version 41-2001EN Serial Numbers: 30264 616174 616689 616695 616701 616703 616706 615658 615668 615551 615552 615555 615556 615557 615559 602803 602941 603335 603342 603406 603407 603409 603410 603412 602157 602159 602160 602163 602164 602168 602169 602170 602171 602172 615637 616688 614764 615435 615526 615604 615606 615607 615623 603631 604118 604119 604120 604127 604130 604138 616466 602877 602895 616584 616581 603518 603529 603555 614730 614734 614736 614737 614958 614970 614971 614972 614973 616719 616720 616725 616727 616978 616432 30100 30201 30202 30203 616491 616492 616493 616494 616496 616500 616503 616456 600613 615937 616338 616380 616678 616679 616683 616685 616687 616702 616344 616351 616384 616604 616608 616614 616682 616690 616691 616825 603857 603858 603859 603860 603862 603863 603865 603868 603869 603870 603871 603872 603876 603877 603881 603883 603885 603892 603897 603901 603906 603907 603908 603909 603914 603915 603916 603917 603918 603919 603920 603921 603922 603923 603924 603925 603926 603927 603928 603929 615717 615719 615728 615729 615742 615860 615886 630322 630714 630679 631728 632096 632135 616480 616488 614532 614605 614752 615707 616991 616999 602639 614844 614759 614767 614769 614841 614856 614891 614893 614894 614896 615047 615048 615049 615050 615051 615052 615053 615055 614780 614276 614546 614548 614549 614550 614552 614553 614626 630653 630632 630643 630647 630358 630621 630340 630611 630617 630622 632137 602214 602217 601897 602423 603646 630150 630782 630803 630827 630835 615301 604063 604060 604093 604097 616465 614772 614774 614777 614815 614820 614821 630344 630625 630634 616375 616410 615481 600692 630033 630039 630111 630191 630218 630233 615290 615383 615393 615401 615411 615413 615433 615434 615436 602165 602177 602194 602205 616455 604150 614281 615087 604047 615057 615058 615070 615074 615077 615078 615079 615080 615081 615446 615114 615116 615221 615222 615223 615224 615225 615226 615227 616937 30093 30094 30096 30098 30099 600062 600063 600696 601984 602448 616490 616508 616510 600698 600731 602521 615999 616002 616007 616026 616031 40105 603823 603660 603831 603835 603836 603850 603849 615549 614888 615730 615132 615231 615235 602049 602444 615615 616295 616306 616308 616309 616311 616315 616317 616318 616321 616346 616348 616352 616363 616364 616385 616398 616404 616408 616411 616413 616415 616429 616436 616437 616446 616448 616449 616450 616452 616454 616457 616458 616459 616462 616464 616467 616471 616473 616474 616475 616477 616478 616479 616481 616483 616485 616487 616616 616617 616618 616619 616623 616627 616628 616630 616633 616634 616635 616636 616637 616638 616640 616642 616645 616647 616656 616658 616659 616660 616662 616666 616668 602317 601950 602431 602051 614802 614370 614392 603700 603701 603703 603705 603707 603712 603719 603723 603724 614718 615562 615590 615698 615402 615412 615418 615420 614940 614941 614944 615399 615405 615408 615409 616712 616714 616738 616743 616745 616747 616749 602046 602047 602048 616586 616587 616676 616696 616717 616718 616723 616726 616728 616730 616740 616742 40141 40140 603847 603852 615565 616037 616036 602079 602464 602545 30220 30221 30222 30223 30224 30225 30226 30228 30229 30232 30233 30234 30235 30236 30237 30238 30244 30227 617009 614854 614927 615228 615229 615230 615232 615233 615234 616580 602409 601915 602041 614750 614790 615558 630792 615864 615865 601977 604094 604107 614573 602083 602103 604106 601969 600053 600054 600368 600827 600831 600832 602062 602428 603837 603841 40127 40142 602218 602506 602515 40199 40200 600611 630148 630149 630161 630162 630178 630190 630204 630739 630743 630752 630781 630799 630806 630810 630829 630830 630832 630833 631040 631041 631044 631051 631052 631061 631088 630003 630054 630194 630206 630740 630749 630779 630790 630793 630796 630809 630813 630816 630819 630826 630828 630837 630838 631002 631012 631026 631031 631043 631063 631081 630022 630082 630095 630121 630200 630287 630787 630789 630801 630836 630996 631013 631035 631319 631324 631326 631331 631333 631346 631347 631349 631350 631358 631365 631402 630004 630009 630012 630015 630019 630026 630055 630058 630060 630160 630208 630695 630760 630762 631005 631008 631019 631033 631062 631066 631071 631075 631079 631080 631101 630010 630053 630131 630140 630158 630164 630175 630176 630180 630183 630195 630203 630207 630713 630716 630717 630764 630769 631092 631105 614905 615876 600008 602434 603638 603639 603644 616377 616522 600960 601839 601920 601928 602394 602395 602474 602477 602020 603610 600901 601849 603478 603479 602747 602490 601583 601854 601941 601946 602009 602018 602533 602730 602771 602777 602891 602892 602922 602972 615797 602013 602557 602576 601909 603833 600003 616272 616271 614853 614859 614861 614862 614863 614864 614865 614866 614867 614868 614869 614870 614871 614872 614873 614874 614875 614876 614877 614878 40115 40116 40125 40126 616185 616188 614371 615421 615448 615482 615631 614883 614884 614885 614886 614887 614889 614890 615829 615397 614610 615258 615262 615265 615268 615269 615276 616192 616215 615577 615579 615585 615599 615620 614911 614912 614913 614914 614915 615278 615567 615574 615586 615589 615591 615594 615598 616958 616989 615769 615778 615779 616795 616798 616800 616801 616830 616851 630081 615880 615881 615882 601916 614901 614895 614897 615751 614500 615168 615346 615353 615360 615361 615364 615381 615386 615388 615389 616207 40261 40263 40264 631218 631229 614809 614814 602085 615384 616218 614789 615450 614618 614640 615471 615701 615255 615256 615257 614907 603642 603640 601532 601588 603326 615075 615851 615900 616499 640126 630373 630455 631177 631201 631203 631756 631784 631785 631786 631799 640383 640390 640417 640427 640423 640480 640421 640463 640055 640056 640065 640066 640067 640069 640075 640085 640089 640112 601004 30110 601070 600763 600191 604099 604100 616127 630021 630049 616849 616893 616896 614917 614918 600137 600138 600139 600909 614222 615753 615765 615767 615770 616498 602076 615456 616838 604066 614831 614835 614315 616502 616504 616513 616519 616524 616531 616539 616541 616550 616520 616546 615825 616316 616378 616505 616509 616521 616523 616527 616530 616533 616535 616548 600373 600375 600374 602199 630167 630215 630219 630226 630227 630230 630232 630235 630236 630239 630241 630245 630247 630249 630250 630251 630266 630268 630270 630273 630292 630294 630295 630297 630299 602057 614796 614251 615277 604115 615522 615700 615613 615877 615879 616463 616470 603878 603880 603882 603884 603886 603887 603888 603889 603890 603891 603893 603895 603896 603898 603899 603900 603902 603903 603904 603905 616388 616497 616529 603657 603658 603976 604003 601447 601460 601462 601473 601498 615943 614210 615382 616412 616445 616476 616482 601456 601449 615907 630028 630030 630038 30081 30082 30083 30064 30063 602329 602324 603633 630221 601488 615919 30051 616177 616178 616184 616200 616212 616175 616216 616094 615487 615488 615507 615510 615512 615516 615521 615524 604078 615059 615065 616833 616945 603308 614328 630035 630776 630802 630821 615365 615367 615387 615406 615248 615249 615252 615253 603845 602398 615609 615611 615630 615633 615635 616392 616414 616416 616441 603632 604051 614236 614247 616820 630034 630110 616722 616750 616752 616764 616765 616769 616772 616782 616784 616786 616791 616793 616797 616799 616804 616805 616807 616815 616816 615419 614414 600416 616374 615933 615948 616334 616362 616376 616495 616661 616680 616681 616686 616693 603867 603873 603641 603643 601954 604072 615462 615469 616453 616299 616422 616440 602251 604073 604062 630608 615314 615334 615337 615339 615608 615624 615629 615632 615997 616001 616003 616024 616033 616034 616004 615592 615662 616577 614823 614839 615872 615561 615581 615583 615690 615254 604085 600391 600396 602443 614829 614833 614834 614858 614860 614347 614348 614349 614350 604074 616281 616312 616319 616320 616332 616333 616359 616365 616366 616369 616372 616386 616390 616420 616424 616428 616431 616434 614355 614356 603629 600033 615032 615034 615036 615037 615038 615039 615040 615060 615061 615088 615089 615090 615091 615092 615093 615094 615095 615096 615097 615098 615099 615100 615101 615102 615103 615104 615105 615106 615107 615108 615109 615110 615111 615112 615113 615115 615117 615118 615119 615120 615121 615123 615124 615125 615126 615127 615128 615131 615578 615575 616159 614964 614966 614967 631544 631601 631629 615062 615063 601952 30077 40124 40143 600166 600167 614817 615441 604113 614238 603825 615064 615073 630636 616439 616010 616015 614576 615760 615236 614335 615784 615478 615494 615506 602525 600027 614566 615281 615286 615288 615289 615291 615294 615296 615299 615307 615447 40149 40150 40151 600185 600186 600187 600369 600521 600523 600524 600680 600681 600683 600684 600685 600689 600693 600972 601624 601919 601935 601943 601947 601979 601980 601982 601985 601999 602014 602021 602027 602029 602123 603348 603385 603420 603436 600700 602696 602704 602705 602706 600208 602038 600207 600209 600210 603311 615452 616472 630024 630031 614784 614786 614830 614850 616133 616143 616179 600107 30085 40144 602156 601397 600324 601394 600367 614898 616443 616489 603910 616077 616078 601428 30052 616812 615890 614816 615042 615673 616109 616118 616170 615806 615812 603866 614625 615870 616779 616787 616792 616821 616829 616834 616835 616836 616837 616839 616840 616842 600701 600709 600710 602265 603411 615083 616468 615675 615676 615678 615679 615680 615681 601197 601235 601303 600809 600434 600356 600522 600508 600858 600575 601146 601048 600853 601152 601114 600997 601199 603481 600235 600237 600239 600227 600270 600280 600283 600319 600320 600330 600331 600333 600347 600350 600354 600355 600357 600362 600381 600417 600427 600435 600438 600439 600440 600442 600443 600444 600445 600448 600450 600451 600452 600453 600456 600457 600478 600500 600509 600512 600089 600231 600244 600245 600246 600248 600249 600250 600251 600253 600254 600256 600257 600258 600259 600260 600261 600262 600263 600264 600265 600268 600269 600271 600272 600274 600275 600278 600281 600282 600284 600300 600308 600318 600322 600326 600327 600328 600329 600341 600343 600344 600346 600351 600352 600359 600360 600364 600365 600370 600377 600379 600382 600398 600401 600403 600413 600414 600415 600418 600419 600421 600423 600425 600428 600429 600430 600431 600432 600436 600454 600469 600472 600474 600475 600476 600479 600480 600482 600483 600485 600486 600487 600488 600489 600490 600492 600493 600494 600496 600498 600499 600501 600503 600504 600505 600507 600511 600514 600515 600516 600517 600518 600519 600526 600310 600345 600348 600404 600409 600420 600422 600437 600441 600447 600449 600455 600463 600467 600470 600473 600477 600481 600484 600491 600495 600502 600510 600513 600520 601953 602065 602097 602161 602176 602296 602297 602299 602300 602302 602303 602305 602306 602309 602310 602311 602312 602313 602315 602322 602327 602328 602331 602520 602523 601191 601348 600363 600342 600446 600433 600252 600247 600397 600977 614694 30036 600141 600157 600158 600159 600190 600198 600199 600202 600203 614674 614678 614679 614680 614681 614682 614683 614686 614687 614688 614689 614690 614691 614692 614693 602733 602741 603337 603424 603538 603547 603558 603564 603573 603574 603575 603576 603577 603578 603579 603580 603582 603584 603585 603589 603603 603607 603609 603666 603667 603726 603728 603729 603730 603731 603733 603734 603736 603737 603943 603944 603946 603948 603961 603963 614659 614980 614985 614989 614990 615002 615005 615007 615184 615188 615196 615197 615199 615207 615219 603323 603328 603341 603367 603463 603970 603978 603999 604001 604005 604006 604009 604027 614684 614685 614738 614942 614945 614969 615011 615012 615029 615165 615190 615192 615200 615204 615206 615211 615212 615215 615218 602710 603090 603947 603969 603974 603980 603981 603983 603984 603985 603986 603991 603993 603994 603997 604002 604004 604010 604011 604013 604015 604017 604021 604023 604026 604028 604033 604036 614654 614656 614664 614665 614667 614672 614675 614677 614731 614732 614733 614740 614741 614943 614960 614968 614978 614987 614991 614992 614993 614994 614996 615003 615004 615008 615009 615015 615018 615022 615024 615027 615028 615030 615161 615166 615186 615187 615191 615193 615194 615201 615208 615213 615214 614669 614671 614676 614729 614735 614981 614982 614984 614986 614995 614998 614999 615000 615001 615209 614655 614696 614697 614700 614702 614721 614722 614727 614728 614988 615172 615179 615180 615181 615216 630368 630370 630407 630408 630409 630416 630421 630423 630425 630426 630428 630429 630430 630437 630444 630447 630451 630461 630465 630466 630472 630476 630361 630362 630376 630417 630419 630420 630424 630431 630433 630440 630441 630442 630445 630446 630448 630449 630450 630453 630458 630463 630464 630467 630471 630473 630475 630389 630391 630432 630456 630457 630470 630474 630477 630479 630481 630482 630490 630491 630492 630495 630497 630499 630507 630508 630515 630525 630531 630532 630535 630536 630537 630540 630384 630387 630398 630410 630412 630414 630415 630452 630460 630493 630500 630505 630545 630553 630557 630562 630567 630568 630571 630576 630577 630578 630579 630584 630587 630363 630364 630378 630388 630390 630392 630394 630395 630400 630406 630411 630413 630422 630427 630580 630582 630583 630585 630586 630588 630589 630590 630593 630594 630596 630598 630600 631160 631166 631168 631171 631173 631175 631181 631183 631188 631189 631190 631194 631195 631196 631198 631199 631200 631202 631204 631205 631206 631210 631212 631213 631216 631176 631217 631227 631228 631233 631220 631231 631234 631236 631239 631240 631242 631244 631245 631247 630501 630502 630529 630572 630581 631180 631182 631185 631187 631251 631252 631733 631734 631735 631736 631737 631738 631739 631743 631745 631746 631747 631749 631751 631753 630386 630418 630509 630563 630591 631740 631741 631742 631744 631748 631750 631752 631754 631755 631757 631758 631760 631761 631763 631766 631767 631768 631774 631775 631778 631779 631780 631781 631782 631787 631790 631792 631797 631798 631803 631804 631805 631806 631808 631809 631811 631817 631821 631822 631825 631832 631837 631839 630595 631762 631765 631769 631770 631771 631772 631773 631776 631777 631788 631789 631791 631793 631794 631795 631796 631801 631802 631813 631814 631816 631819 631824 631826 630552 631246 631248 631249 631764 631800 631807 631810 631812 631815 631818 631820 631823 631827 631828 631829 631833 631834 631835 631836 640004 640005 640006 640008 640009 640013 640015 640016 640017 640019 640020 640021 640022 640024 640025 640026 640027 640028 640029 640030 640031 640032 640033 640034 640035 640036 640041 640044 640045 640050 640063 640064 640072 640073 640076 640078 640093 640105 640106 640123 631159 631162 631163 631164 631167 631172 631178 631179 631186 631191 631192 631193 631197 631208 631209 631241 631243 630380 630385 630393 630396 630397 630399 630401 630402 630405 630434 630435 630436 630438 630439 630443 630365 630369 630371 630372 630375 630377 630379 630381 630382 631207 631211 631214 631215 631219 631223 631161 631165 631169 631170 631222 631224 631232 631235 615185 614902 614909 614910 604156 614214 614260 614287 614289 614292 614303 614804 615076 615861 615775 615732 615531 615532 615533 615534 615535 615536 615537 615546 615603 615622 615660 615691 615692 615695 615696 615697 615712 615560 615576 615628 615311 615315 615320 615326 615329 615343 616022 616028 601901 601956 601972 602093 602111 614228 614229 614230 614231 614232 614233 614290 614293 614294 615454 615480 615499 602063 602066 602078 602314 602316 602518 615519 615503 615523 616152 616154 616155 616164 616149 631374 632109 631250 631830 631831 631838 631840 616157 616166 616168 616172 603843 603844 603851 603853 603855 615550 614791 616582 615250 615251 615133 615241 615242 615243 615244 615572 604092 616106 616112 616113 616114 616119 616182 616186 616190 616193 616194 616198 616202 616203 616204 616205 616214 616219 616223 616224 616227 616231 616232 616234 616236 616237 616241 616700 616715 614635 616117 616196 616208 616210 603477 615563 602253 616307 616325 616329 616342 616350 616356 616358 616360 616381 616382 616383 616394 616396 616402 616417 616418 616430 603955 603956 603959 602960 602206 615495 615502 615511 615529 615530 615616 603614 603679 616824 601528 630312 630320 630321 630325 30080 40102 40103 40104 40106 40107 40109 40110 600010 600059 40119 600884 600910 600913 600915 600916 600919 600920 600921 600922 600924 600925 600926 600927 600930 600931 600932 600934 600935 600936 600937 600938 600939 600942 600943 600944 600945 600946 600950 600953 600954 600955 600956 600957 600958 600959 600961 600962 600963 600964 600965 600966 600967 600968 600971 600973 601540 601990 602008 601268 601276 601361 601559 601560 601563 601564 601565 601566 601569 601570 601579 601581 601584 601585 601587 601590 601591 601592 601628 601629 601827 601828 601829 601832 601833 601834 601835 601836 601837 601838 601840 601841 601842 601843 601844 601845 601850 601851 601853 601855 601857 601858 601921 601922 601924 601925 601926 601927 601929 601930 601931 601932 601933 601934 601937 601938 601939 601940 601942 601944 601945 601948 601981 601983 601986 601987 601988 601989 601992 601993 601996 601997 601998 602000 602001 602002 602003 602005 602007 602010 602011 602015 602016 602017 602019 602022 602023 602024 602025 602028 602030 602033 602035 602036 602037 601492 601666 601673 601674 601682 601676 601677 601679 601659 601672 602154 602173 602178 602210 602219 602220 602526 602530 602532 602543 602377 602385 602388 602389 602391 602392 602393 602399 602401 602402 602403 602455 602457 602459 602462 602473 602480 602483 602485 602491 602454 602456 602461 602465 602467 602468 602470 602472 602475 602476 602478 602484 602488 602489 602527 602528 602529 602531 602534 602537 602538 602539 602540 602542 602544 602546 602547 602548 602549 602550 602551 602552 602553 602554 602555 602556 602558 602559 602561 602562 602563 602564 602568 602569 602571 602572 602573 602574 602575 602578 602579 602580 602581 600940 601558 601571 601573 601595 601609 601618 601995 602374 602375 602376 602378 602379 602381 602382 602383 602356 602380 602463 602588 602604 602606 602608 602616 602620 602623 602626 602633 602640 602641 602642 602644 602647 602648 602649 602652 602653 602654 602458 602469 602524 602560 602702 602703 602711 602715 602716 602717 602718 602719 602720 602721 602723 602724 602725 602727 602728 602731 602732 602734 602735 602737 602738 602740 602742 602743 602744 602746 602748 602749 602751 602752 602753 602754 602755 602757 602758 602760 602761 602763 602764 602766 602769 602783 602785 602790 602793 602804 602729 602750 602756 602775 602784 602787 602791 602800 602801 602838 602840 602842 602843 602844 602847 602850 602854 602861 602865 602866 602868 602869 602874 602879 602889 602893 602896 602897 602898 602899 602901 602913 602915 602916 602917 602918 602924 602927 602966 602967 602968 602969 602970 602745 602765 602770 602806 602809 602815 602872 602875 602883 602894 602900 602907 602912 602914 602919 602921 602923 602925 602926 602932 602971 602977 602981 602983 603005 600923 601847 603361 603461 603462 603464 603465 603466 603467 603468 603470 603471 603472 603473 603475 603476 603480 603483 603487 603488 603490 603491 603492 603493 603495 603498 603506 603507 603510 603512 603514 603515 603517 603519 603521 603522 603525 603526 603527 603528 603530 603531 603533 603535 603536 603537 603541 603544 30076 40101 603310 603313 603331 603351 603357 603358 603360 603388 603394 603397 603405 603415 603429 603565 603568 603583 603595 603597 603604 603608 603611 603615 603617 603674 603677 603717 603718 603722 603314 603332 603352 603379 603427 603566 603567 603569 603570 603572 603594 603599 603600 603601 603602 603612 603613 603678 603720 603721 614670 602031 601455 602053 602307 602326 614318 614319 614320 614334 614336 614438 614619 614623 614627 614628 614631 615682 615684 615685 615686 615722 615723 615724 615789 615790 615791 615793 615794 615795 615796 615798 615838 640007 640010 640012 640014 640018 640037 640038 640039 640042 640046 640051 640052 640054 640057 640059 640077 640084 640097 640098 640100 603630 600205 600206 602207 600241 600255 600366 600376 600279 600424 615888 600321 600222 600405 600616 601115 601175 601353 601439 601534 601538 630374 600773 602186 602187 602026 602280 602541 602565 602691 603431 602499 603606 603321 614699 603725 600779 603598 603962 614997 615203 40145 40215 40216 40217 40218 40219 40220 40224 40225 40226 40228 40229 40232 40233 40234 40235 40236 40238 40239 40240 40241 40242 40245 40246 40247 40248 40249 40250 40251 40252 40253 40254 40256 40257 40258 600013 600014 600018 600021 600023 600024 600026 600028 600029 600030 600032 600036 600037 600038 600039 600041 600044 600045 600046 600047 600048 600051 600052 600060 600643 600647 600649 600652 600653 600654 600656 600657 600658 600659 600660 600661 600662 600663 600723 600724 600725 600726 600727 600728 600729 600730 600732 600733 600734 600735 600736 600738 600740 600742 600744 600745 600746 600747 600748 40259 600049 40230 600040 600043 40227 40223 40244 600017 602577 600050 600031 40221 40222 40255 602888 602929 602931 602937 602939 602942 602944 602950 602951 602985 602989 602993 602995 603001 603004 603334 603336 603339 603340 603343 603344 603346 603353 603354 603355 603404 603413 603416 603417 603418 603419 603421 603422 603425 603426 603428 603430 603434 603435 603437 602805 602902 603338 603440 603442 603446 603449 603450 603451 603452 603453 603456 603469 603484 603486 603497 603500 603501 603502 603503 603586 603590 603591 603592 603593 603596 603605 603664 603665 603668 603669 603670 603671 603672 603673 603675 603680 603681 603682 603683 603686 603688 603693 603697 603698 603732 603958 603988 603989 603992 603996 603998 604014 604016 604030 604039 604042 604043 604044 40231 40237 40243 600022 600025 600042 600642 614638 603684 603685 603687 603690 603691 603692 603694 603695 603696 604018 604031 615509 617094 617106 617131 617142 617167 617171 617182 617196 617207 617215 617218 617221 617233 617234 617235 617244 617246 617261 617267 617271 617272 617274 617283 617300 617304 617098 617100 617101 617103 617119 617120 617125 617128 617136 617141 617145 617151 617157 617161 617162 617169 617190 617213 617226 617228 617230 617238 617248 617276 617299 617121 617126 617130 617134 617137 617147 617152 617154 617172 617175 617185 617192 617199 617201 617208 617209 617216 617241 617243 617259 617273 617289 617297 617298 617307 617023 617066 617069 617072 617075 617083 617093 617095 617123 617129 617132 617144 617146 617149 617178 617179 617194 617210 617220 617117 617150 617156 617183 617200 617223 617227 617270 617291 617306 617122 617135 617148 617159 617168 617174 617180 617184 617186 617195 617203 617212 617229 617232 617254 617256 617266 617269 617278 617284 617286 617288 617303 617305 617309 630468 630469 630478 630480 630484 630485 630486 630487 630488 630489 630503 630504 630510 630511 630512 630513 630517 630518 630522 630542 630496 630498 630516 630519 630520 630521 630524 630526 630527 630528 630530 630533 630539 630546 630549 630551 630558 630561 630564 630565 630494 630506 630514 630523 630534 630538 630541 630543 630544 630548 630550 630554 630555 630556 630559 630566 630570 630573 630574 630357 630603 615489 615500 615508 604061 616512 616514 616516 616518 602105 602112 602113 614799 615758 614792 614812 614819 614826 614832 614843 614847 614848 616965 616969 616992 617018 617020 614248 614252 614253 614254 614256 614261 614265 614266 614268 614270 614271 614272 614273 614274 614275 614277 614286 614327 614331 614415 614416 614417 614418 614419 614420 614422 614423 614424 614426 614427 614428 614429 614430 614431 614432 614433 614476 614479 614480 614481 614482 614483 614484 614487 614492 614499 614368 614373 614374 614375 614378 614379 614401 614403 614404 614405 614406 614407 614409 614410 614411 614434 614435 614437 614439 614440 614441 614442 614443 614444 614445 614446 614449 614473 614485 614486 614488 614489 614490 614491 614493 614495 614496 614497 614498 614502 614504 614505 614506 614510 614511 614512 614513 614514 614515 614516 614517 614518 614519 614520 614521 614522 614580 601663 603661 603861 604086 604095 604096 604098 604101 604103 604105 604108 604110 604111 604112 604114 604116 604117 604140 604160 604161 604164 604165 614200 614203 614205 614206 614207 614208 614211 614212 614213 614216 614219 614220 614221 614284 614296 614297 614298 614301 614302 614304 614305 614306 614307 614308 614309 614310 614311 614312 614313 614314 614316 614317 614337 614338 614339 614340 614341 614342 614343 614344 614345 614346 614394 614395 614399 614400 614402 614436 614447 614450 614453 614454 614457 614460 614461 614462 614465 614466 614467 614468 614469 614470 614471 614472 614474 614478 614503 614507 614509 614526 614527 614529 614530 614533 614534 614535 614537 614538 614539 614540 614541 614542 614543 614544 614545 614554 614555 614556 614559 614560 614561 614562 614563 614565 614567 614568 614569 614571 614574 614575 614577 614578 614579 614581 614582 614583 614584 614585 614586 614587 614589 614590 614591 614592 614593 614594 614595 614596 614597 614598 614599 614600 614602 614603 614606 614607 614608 614611 614613 614614 614616 614617 614459 615694 615688 615699 615708 615710 615711 630657 630664 630667 630671 630675 630683 630686 630687 630705 630711 630720 630722 630724 630729 630730 630732 630737 630738 630742 630748 630754 630755 630758 630765 630768 630770 630771 630772 630773 630778 630329 630334 630336 630354 630359 630605 630646 630648 630652 630678 630681 630684 630685 630703 630706 630709 630718 630725 630726 630727 630731 630734 630736 630744 630745 630746 630751 630753 630756 630281 630293 630307 630309 630310 630313 630314 630315 630316 630317 630318 630319 630326 630330 630331 630332 630333 630335 630338 630339 630341 630342 630343 630345 630347 630348 630349 630350 630351 630353 630355 630356 630360 630601 630604 630606 630607 630609 630610 630613 630614 630615 630616 630618 630619 630620 630624 630627 630628 630629 630633 630635 630637 630640 630650 630654 630655 630656 630658 630659 630660 630663 630668 630669 630670 630673 630677 630690 630692 630721 630007 630018 630228 630276 630300 630328 630642 630662 630665 630666 630672 630674 630680 630682 630689 630691 630693 630694 630696 630697 630698 630699 630700 630702 630704 630707 630715 630719 630728 630747 630750 630757 630763 630766 630775 631032 631310 631369 631406 631449 631529 631530 631532 631541 631551 631697 631720 631861 632004 632027 632056 632070 632093 632103 632105 630786 630823 631009 631073 631074 631076 631089 631253 631273 631282 631284 631287 631315 631345 631362 631381 631385 631397 631453 631490 631499 631533 631537 631538 631539 631548 631550 631554 631556 631575 631609 631701 631717 631863 631896 631910 631914 631934 631935 631936 631937 631939 631941 631946 631950 631952 631953 631956 631962 631968 631969 631973 631976 631978 631979 631981 631983 631987 632006 632025 632028 632029 632030 632032 632033 632034 632035 632037 632039 632054 632065 632066 632069 632081 632085 632090 632097 632108 632110 632124 632125 632131 632140 632141 632159 632166 601951 601959 602092 614448 614463 614755 615285 615437 616255 616256 616526 616566 614620 616087 616088 616090 616095 616098 616104 616107 616111 616123 616124 604149 614217 614245 614259 614329 615400 615805 615813 616122 616125 616131 616150 616151 616165 616167 616235 616238 616240 616242 616243 616244 616245 616246 616247 616248 616249 616251 616257 616273 616274 616276 616277 616279 616280 630222 630231 630237 630238 630252 630253 630267 630274 630275 630277 630278 630285 630301 630302 630214 630223 630225 630240 630242 630243 630248 630255 630257 630260 630264 630269 630271 630272 630280 630282 630283 630284 630286 630290 630291 630042 630043 630105 630113 630115 630116 630205 630211 630213 630220 630224 630229 630244 630246 630254 630256 630258 630265 630298 630305 630306 630815 630839 630840 614330 615525 615858 615859 602396 614624 614641 603474 603494 614879 614880 615347 615369 615845 601550 601552 616071 616072 614393 615031 615035 615041 615518 615496 615504 615514 615527 615528 616844 616845 616847 616871 616895 616606 616657 616669 616674 616677 616050 615054 615271 615297 615321 615332 615338 615438 601112 614757 614758 614760 614763 614765 615748 615759 600914 600349 601285 30084 30215 30217 30219 40181 40193 600094 600095 600109 603444 603445 603447 603457 603496 603499 603505 600825 600870 600106 600193 600195 600196 600226 600267 600273 600361 600378 600380 600388 600389 600390 600393 600399 600407 600408 600410 600411 600412 600458 600459 600460 600462 600465 600466 600468 600471 600525 600527 600528 600530 600532 600533 600534 600535 600540 600542 600544 600555 600563 600564 600565 600566 600567 600568 600569 600571 600574 600576 600577 600578 600579 600580 600581 600582 600583 600584 600586 600589 600590 600591 600592 600593 600594 600595 600596 600597 600598 600599 600600 600601 600602 600603 600604 600605 600607 600608 600609 600666 600667 600668 600669 600670 600671 600672 600673 600674 600675 600739 600741 600743 600749 600750 600752 600755 600756 600757 600758 600759 600761 600762 600765 600766 600767 600768 600769 600770 600772 600775 600776 600777 600781 600784 600786 600787 600788 600789 600790 600791 600793 600798 600799 600800 600801 600803 600804 600805 600806 600807 600808 600810 600811 600812 600813 600814 600815 600816 600817 600818 600819 600820 600826 600828 600829 600830 600266 600289 600400 600464 600531 600543 600556 600570 600572 600585 600587 600588 600606 600737 600751 600753 600782 600783 600785 600794 600796 600802 600821 600822 600833 600834 600837 600838 600839 600841 600842 600843 600844 600848 600849 600851 600854 600862 600863 600864 600865 600866 600867 600869 600875 600878 600879 600880 600835 600845 600846 600847 600850 600852 600855 600856 600857 600859 600860 600861 600868 600871 600872 600874 600876 600877 600881 600882 600883 600885 600886 600887 600888 600889 600891 600893 600895 600896 600897 600898 600900 600902 600904 600905 600907 600908 600911 600912 600276 600386 600392 600780 600918 600970 600975 600976 600978 600979 600980 600981 600982 600983 600984 600985 600986 600987 600988 600989 600990 600991 600993 600994 600995 600998 600999 601018 601019 601020 601041 601042 601043 601044 601045 601047 601049 601050 601052 601053 601054 601055 601056 601058 601059 601060 601061 601062 601063 601064 601065 601066 601067 601068 601069 601071 601072 601073 601074 601075 601076 601077 601078 601079 601080 601081 601082 601083 601084 601085 601086 601087 601088 601089 601090 601091 601092 601093 601094 601095 601096 601097 601098 601099 601100 601101 601102 601103 601104 601105 601106 601107 601108 601109 601110 601111 601113 601117 601118 601119 601120 601121 601122 601123 601124 601125 601126 601127 601129 601130 601132 601133 601134 601135 601136 601137 601138 601139 601140 601141 601142 601143 601144 601145 601147 601148 601149 601150 601151 601153 601154 601155 601156 601157 601158 601159 601160 601161 601162 601163 601164 601165 601166 601167 601168 601169 601170 601171 601172 601173 601174 601176 601178 601179 601180 601181 601182 601183 601184 601185 601186 601187 601188 601189 601190 601192 601193 601194 601195 601196 601198 601201 601202 601203 601204 601205 601206 601207 601209 601210 601211 601212 601213 601214 601215 601216 601217 601218 601219 601220 601221 601222 601223 601226 601227 601228 601229 601230 601231 601232 601233 601234 601236 601237 601238 601239 601240 601241 601242 601243 601244 601245 601246 601247 601248 601249 601250 601251 601253 601254 601255 601256 601257 601258 601259 601260 601261 601262 601263 601264 601265 601266 601267 601269 601270 601271 601272 601273 601274 601275 601277 601278 601279 601280 601281 601282 601283 601284 601286 601287 601288 601289 601290 601291 601292 601294 601295 601296 601297 601299 601302 601304 601305 601306 601307 601308 601309 601310 601311 601312 601313 601314 601315 601316 601317
601318 601319 601320 601321 601322 601324 601325 601327 601328 601329 601331 601128 601323 601326 601330 601332 601334 601335 601336 601337 601338 601339 601340 601341 601343 601345 601346 601347 601349 601352 602133 602135 602136 602143 602144 602146 602147 602150 602152 602179 602184 602185 602736 602762 602772 602773 602774 602776 602778 602779 602782 602789 602796 602798 602802 602808 602810 602811 602812 602813 602814 602816 602817 602820 602821 602822 602823 602824 602825 602826 602827 602828 602829 602830 602832 602834 602835 602836 602837 602841 602848 602851 602857 602858 602859 602860 602864 602867 602871 602873 602876 602878 602880 602882 602885 602886 602887 602890 602903 602904 601116 601298 601301 601300 601051 616665 616667 615862 615863 631610 615457 615465 615473 615543 615568 615569 615573 615672 615738 614920 614923 602295 616757 616796 616810 615356 615416 614201 614215 614218 614239 614881 615595 614768 614770 30207 30214 30216 600690 600694 601342 600704 40133 614551 614564 615066 615067 615068 615069 615071 614778 614779 614781 615263 615284 615295 615323 615351 615352 614570 615335 614421 615266 615272 615283 615292 615308 615309 615312 615316 615318 615327 615342 615344 615348 615349 615355 615359 614359 614360 614363 602234 615553 614808 615548 614846 615713 616101 614806 614805 614807 614810 615474 615571 615582 615741 615771 616100 616105 615056 615331 615363 615370 615385 615403 615362 615905 615909 600200 602964 30104 30166 30500 30116 30048 30170 614824 614837 615275 615279 615280 615282 615293 615298 615302 615306 615477 615043 615044 615045 600873 601668 602261 601899 601683 601903 601900 603824 601904 604052 604056 615492 615539 615587 615588 615619 604067 615086 616461 616540 616553 616556 616571 616547 600242 614572 615677 615702 615718 615735 603485 603710 603711 603713 603714 603715 603716 614743 614744 614747 614749 614924 614925 614926 615792 616438 616447 602441 604102 614785 614793 614798 632133 631348 631488 631526 631534 631706 631993 632126 632130 632134 632138 631288 631372 631393 631463 631486 631710 631732 632021 632049 632076 632082 631996 632047 603561 603571 616817 616819 616843 616848 616853 616858 616861 616862 616864 616865 616866 616867 616869 616873 616874 616875 616879 616880 616881 616883 616884 616886 616892 616899 616900 616902 616907 616910 616914 616915 616921 616925 616928 616941 616942 616944 616954 616970 616988 614788 615706 615317 615336 602268 602273 602278 602288 602345 602346 602347 602349 602350 602351 602353 602355 602357 602359 602360 602362 602363 602364 602365 602367 602368 602369 602370 602372 602373 616156 616161 616163 616160 616162 615593 615597 602155 602175 602189 602190 602191 602195 602196 602200 602201 601911 603949 603951 614390 614391 603654 604084 604088 601889 601892 614782 615513 602241 602258 603832 614801 601898 601914 602505 601974 602511 603662 601489 601958 601973 614524 614525 601893 602082 602091 602332 602427 603627 603628 603930 603931 603934 603932 603935 603938 603940 602420 602449 602519 603937 603941 604046 604050 603827 603828 603834 603839 616115 616116 616121 602416 602193 631039 631091 601957 604070 616191 630824 630834 615731 615850 615247 600162 600163 602256 615733 615747 615756 615763 615773 615777 614882 616826 630630 630639 601890 615721 600823 602032 616039 616044 616046 616038 616048 601480 601470 601448 601450 601453 601457 601461 601463 601464 601465 601467 601471 601475 601487 601490 601491 601500 601501 601507 601664 601669 601670 601654 601902 601479 602212 602430 602514 601494 602059 602099 603875 603879 603894 614761 615084 615313 601503 602245 602509 600323 601469 601472 601655 603822 603842 601468 601452 602338 601451 601502 601454 601483 601905 601895 601476 630626 630631 630644 630163 631442 616515 616517 615600 615602 615612 615625 603933 603936 615006 616484 616486 614928 614929 614976 614977 601435 601436 601438 601441 601443 600334 600335 600336 600337 600338 600339 30055 601955 602166 614637 614822 616694 616708 30065 30066 30067 30068 30069 30070 30041 40113 40114 40135 40136 40138 40139 40146 40147 40148 40197 30454 30455 40207 40260 600056 600055 600071 600065 600066 600072 600090 600091 600092 600093 600104 600105 600113 600096 602833 640171 614536 615704 615727 615749 615750 615757 615761 615762 616206 615443 615444 615451 615459 615460 615467 615468 615472 615476 615636 615640 615642 615643 615644 615646 615647 615648 615649 615657 615554 615584 615601 615605 615610 615617 615627 615634 632091 615274 615544 600824 631472 631547 632129 614919 615358 615410 615415 616147 616148 632050 632077 632145 632155 601496 603826 603874 614369 614388 614396 614451 614452 614455 614456 614477 614494 614508 614523 614531 614588 614601 614604 614609 614630 615085 615245 615638 615653 615656 615670 615674 615720 615725 615726 615743 615745 616305 616310 616355 616406 616407 616419 616421 616423 616425 616427 616809 616850 616854 616855 616870 616876 616877 616878 616901 616923 630044 630057 630085 630088 630090 630096 630097 630098 630099 630100 630102 630103 630106 630108 630651 630733 630783 630808 603856 603864 603635 630056 630132 630151 630987 630999 631001 631027 631085 631096 631375 631380 631387 631390 631399 631405 631424 631465 631467 631468 631469 631471 631473 604079 604147 604282 614204 614234 614237 614282 615350 615368 616085 616086 615846 615929 615303 615304 615305 615774 615856 615857 616963 40111 40112 40128 40129 40206 600061 600114 600144 600188 600189 600213 600214 600384 600385 600340 602262 602433 602502 602447 603829 602446 615458 615639 600917 600928 600929 600933 600941 600947 600948 600949 600951 600952 600969 601005 601562 601572 601586 601830 601936 602004 602366 602726 602839 602852 602911 603534 603384 616542 616567 616583 616596 616601 615340 614795 616025 630181 615645 614842 615564 615570 615683 615687 615689 614762 614908 614916 614921 614922 615330 615345 615270 615449 615461 615463 615498 616108 616110 603647 615239 602088 602098 616221 616918 616922 616924 616926 616935 614278 614458 615333 615614 615666 616195 616222 616225 616228 616230 614771 616138 616451 615650 615651 616460 616340 616368 616426 615391 615392 30078 603645 602419 603626 603636 631514 631730 631856 631859 631872 631873 615844 615847 615395 615396 616846 616856 616857 616859 616872 616885 616887 616890 616891 616894 616897 616908 616909 616931 602216 602231 602244 602321 602330 602451 602422 602425 602421 602249 602333 602335 602336 602339 602417 602452 602453 602100 614223 614225 614226 614227 614857 615240 615394 615407 615414 615497 615517 616814 616831 616832 616593 616594 616598 616602 616612 616620 616622 616624 616625 616626 616697 616698 616704 616705 616707 616709 616711 615836 40137 615752 615754 630075 630825 616040 616041 604076 604077 616079 616080 616081 616082 616083 616084 616089 616092 616093 616096 616097 616099 616103 616130 616132 616135 616144 616153 616169 616171 616173 616176 616180 616181 616183 616187 616189 616197 616209 616211 616213 604159 614264 614269 615371 615372 615373 615374 615375 615376 615377 615378 615379 615380 616252 616253 616254 616258 616259 616260 616261 616262 616263 616264 616265 616266 616267 616268 616269 616282 616283 616284 616286 616291 616296 616297 616298 616300 616301 616314 616330 616331 616335 616336 616337 616339 616341 616343 616347 616349 616354 616357 616361 616367 616371 616373 616387 616389 616391 616393 616395 616397 616399 616400 616401 616409 616091 614787 615328 615354 615390 615541 615542 615545 615547 614903 614904 615432 615442 600996 602286 602287 602293 602664 602936 602289 602107 602294 602301 602304 602325 602504 614751 630623 630661 614547 615848 604121 604122 604123 604124 604125 604126 604128 604129 604131 604132 604133 604134 604135 604137 604351 614351 614352 614353 614354 614364 614365 614367 614372 614376 614377 604136 615439 630259 630261 630263 614773 614775 614776 614825 614845 616947 615259 615260 615261 615816 615823 615824 615827 615831 615849 615852 615853 615854 615855 615826 631395 600181 600182 602845 600180 630079 630723 630735 630820 30005 30003 30004 30006 30008 30010 30012 30014 30018 30054 30056 30057 30058 30009 30013 30016 30020 30021 30022 30023 30024 30025 30026 30027 30028 30029 30073 30074 30030 30031 30032 30033 30034 30035 30038 30088 30091 30092 30095 30097 30204 30205 30206 30210 30212 30213 30239 30240 30241 30242 30045 30046 30047 30049 30101 30102 30103 30105 30106 30107 30108 30109 30111 30112 30113 30114 30115 30117 40134 30050 30118 30119 30120 30121 30122 30123 30124 30125 30126 30127 30128 30130 30131 30132 30133 30134 30135 30053 30136 30137 30138 30230 30231 30245 30246 30247 30248 30250 30251 30252 30253 30254 30255 30256 30257 30258 30259 30260 30139 30140 30141 30142 30143 30144 30145 30146 30147 30148 30149 30150 30151 30152 30154 30155 30156 30157 30158 30159 30160 30161 30162 30163 30165 30167 30168 30169 30171 30172 30173 30174 30175 30176 30177 30178 30179 30180 30181 30182 30183 30184 30185 30186 30187 30188 30189 30190 30191 30192 30193 30195 30196 30197 30198 30199 30200 30451 30452 30453 40198 30456 30457 30458 30459 30460 30461 30462 30463 30464 30465 30466 30467 30468 30469 30470 30471 30472 30473 30474 30475 30476 30477 30478 30479 30481 30482 30483 30484 30485 30486 30487 30488 30489 30490 30491 30492 30493 30494 30495 30496 30497 600067 600068 600069 600070 600074 600075 600076 600077 600078 600079 600080 600081 600083 600084 30498 30499 30501 30502 30503 30504 30505 30506 30507 30508 30509 30510 30511 30512 30513 30514 30515 30516 30517 600098 600102 600103 600119 600120 600121 600122 600125 600129 600135 600140 600145 600146 600147 600148 600082 600087 600097 600099 600100 600118 600123 600124 600126 600168 600169 600171 600172 600173 600174 600175 600176 600178 600179 600228 600539 600170 600461 600538 600541 600545 600546 600547 600548 600549 600550 600551 600552 600553 600554 600558 600559 600560 600561 600562 600612 600618 600619 600712 600713 600718 600626 600636 600676 600615 600650 600664 600754 600771 600992 600974 601000 601001 601002 601003 601006 601007 601008 601009 601010 601011 601012 601013 601015 601016 601017 600406 600637 601014 601224 601225 601363 601344 601351 601364 601366 601367 601368 601369 601370 601371 601372 601373 601375 601377 601378 601379 601380 601381 601382 601510 601519 601521 601523 601525 601446 601509 601512 601514 601516 601517 601527 601529 601530 601535 601536 601537 601543 601508 601619 601365 601542 601544 601548 601437 601440 601442 601444 601445 601511 601513 601515 601518 601522 601531 601533 601545 601551 601553 601555 601557 601621 601582 601600 601602 601614 601417 601556 601561 601613 601615 601616 601617 601620 601625 601630 601631 602114 602116 602117 602118 602119 602120 602121 602122 602124 602127 602128 602129 602131 602132 602134 602138 602139 602140 602141 602145 602149 602153 602264 602267 602269 602270 602271 602272 602274 602275 602276 602277 602282 602283 602284 602285 602279 602281 602266 602348 602352 602354 602358 602361 602371 602386 602390 602400 602486 602487 602492 602493 602535 601568 601593 601597 601599 601852 601991 602006 602115 602180 602182 602183 602188 602344 602384 602460 602479 602583 602587 602589 601623 602034 602137 602582 602584 602586 602590 602591 602593 602594 602595 602596 602597 602598 602599 602600 602601 602603 602607 602613 602614 602628 602629 602630 602631 602634 602635 602637 602645 602663 602665 602666 602667 602290 602291 602585 602612 602619 602621 602624 602625 602627 602632 602636 602650 602651 602655 602657 602659 602660 602662 602669 602670 602671 602672 602674 602675 601923 602570 602592 602602 602605 602609 602610 602611 602615 602617 602618 602622 602638 602643 602656 602658 602661 602668 602673 602676 602677 602678 602679 602680 602681 602682 602683 602684 602685 602686 602687 602688 602689 602690 602693 602694 602695 602697 602698 602699 602700 602701 602707 602708 602709 602712 602713 602714 602722 602780 602781 602788 602794 602799 602807 602818 602819 602855 602862 602863 602856 602870 602910 602935 602958 603359 603380 603386 603387 603398 603401 603402 602976 603324 603327 603433 603508 603509 603511 603513 603516 603520 603523 603524 603543 603553 603554 603556 603557 603560 603562 602646 602881 602974 602982 602997 603002 602940 602943 602954 602979 603727 603945 602692 603333 603563 614643 614644 614645 614646 614647 614648 614649 603581 604020 614651 614652 614653 614703 614704 614705 614706 614707 614974 614975 614931 614932 614933 614934 614935 614936 614937 614938 614939 614946 614947 614948 614950 614952 614953 614954 614959 614961 614962 614963 603973 603977 615010 615013 615014 615016 615017 615019 615020 615021 615023 615025 615026 615159 615160 615162 615163 615164 615167 615170 615171 615173 615174 615175 603950 603952 603954 603957 603971 614955 603967 604019 604022 604024 604025 604029 604032 604034 604035 604037 604040 615182 615183 615189 615195 615202 615205 615217 615220 615198 600177 601354 601355 601356 601358 601359 601362 600697 600714 600715 600716 600717 600719 600720 600890 601391 601395 601402 601408 616554 616610 616641 616675 615693 615705 615709 616469 600402 616029 616030 614412 614413 631569 631992 40152 40153 40154 40155 40157 40158 600223 600224 600225 600229 600230 600892 600894 600899 600903 601526 601541 601546 601547 601554 600906 601484 602050 602104 602110 603911 603912 603913 616000 616017 600506 615580 615739 616250 616433 616435 614803 603559 600035 30480 601520 601524 601622 30129 615491 615538 615566 616615 616621 616629 616631 616639 616644 616646 616648 616650 616651 616653 616663 616670 616671 616672 616673 616684 600537 30164 30194 30039 30261 30262 30263 30265 30266 30267 30268 30269 30270 30271 30044 30153 615839 615840 615841 615842 616789 616802 614906 604082 630029 630032 630050 630070 630107 615801 615804 615810 615830 615832 615833 615834 615875 616128 601910 614408 615310 614754 615322 615325 615404 615422 615423 615424 615425 615426 615427 615428 615430 615431 615445 615455 615464 615475 615484 615485 615486 615621 615641 615734 615736 615740 614358 30087 40194 40195 40196 40160 40161 40163 40164 40165 40166 40167 40168 40169 40170 40171 40172 40175 40176 40177 40178 40179 40182 40183 40184 40185 40186 40187 40188 40189 40190 40191 40192 600160 600161 600155 600156 600536 600764 600703 600679 600686 600688 600691 600699 600702 600705 600708 600711 600797 600617 600620 600622 600623 600624 600627 600628 600629 600631 600632 600633 600634 600635 600638 600651 600677 600682 600695 600721 600722 600760 600774 600778 600625 600706 30218 40162 601376 601357 601360 601374 601383 601384 601385 601386 601387 601389 601390 601392 601393 601396 601398 601399 601401 601403 601404 601405 601406 601407 601409 601410 601411 601412 601413 601414 601415 601416 601418 601419 601420 601421 601422 601423 601424 601425 601426 601427 601429 601430 601433 601434 600614 601431 601567 601574 601575 601576 601577 601578 601580 601589 601594 601596 601598 601601 601603 601604 601605 601606 601607 601608 601610 601611 601612 601627 601831 601846 601848 601856 40159 600621 40173 602759 602792 602846 602906 602908 602928 602930 602933 602938 602946 602953 602956 602959 602961 602963 602965 602973 602978 602980 602984 602986 602987 602990 602991 602992 602994 603000 603003 603309 603362 603364 603365 603366 602795 602849 602853 602884 602909 602934 602945 602948 602949 602955 602962 602988 602998 602999 603306 603307 603315 603316 603317 603318 603319 603320 603363 603368 603370 603371 603373 603374 603375 603376 603377 603378 603381 603382 603383 601432 601388 40174 603634 603325 603330 603345 603350 603389 603390 603391 603395 603396 603399 603403 603408 603423 603432 603438 603439 603441 603443 603448 603454 603455 603458 603459 603460 603489 603539 603540 603542 603545 603546 603548 603549 603550 603551 603552 603616 603618 603619 603620 603621 603622 603624 603625 603676 603689 603699 603704 603706 603708 603709 600687 40180 614380 614382 614383 614387 614389 614708 614709 614710 614712 614714 614716 614657 614666 614668 614720 614742 614951 614956 614957 614756 615885 603979 603982 603987 604000 604008 614711 615911 615925 615927 615931 615935 615936 615940 615941 615944 615946 616005 616060 616061 616062 616063 616064 616065 616066 616067 616068 602786 602797 602920 602947 603968 603972 614719 614949 614983 615169 630337 630602 630645 630638 602418 601975 602068 602495 602500 614425 614475 614612 614622 614633 615912 615918 615800 615803 604104 615072 615811 615828 615866 615867 615868 615871 615873 615891 615893 615894 615895 615896 615897 615898 615903 615904 615906 615908 615910 615914 615920 616069 616070 615815 600110 600111 600639 600644 600655 615515 600004 600019 600165 601661 630145 630210 632015 600101 615238 616042 616043 603854 602070 603840 604054 604058 604068 604071 604153 604158 604162 604163 614745 602058 615470 616442 615652 615654 615655 615659 615661 615843 615237 616270 616275 616278 616009 616014 616032 616035 614849 614851 614852 614855 614766 615746 615755 614892 615398 602039 602040
Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, GI, ZA, AF, SE, FI.
7129