NeoCoil 3.0T GEM Flex Coil. The 3.0T GEM Flex coil system includes three sizes of coil arrays and one interface box. 3.0T GEM Flex, Interface 16ch Fixed, P -Connector, NC032204, GEHC part number 5430008-5. 3.0T GEM Flex Coil16-L Array, NC040200, GEHC part number 5430008-2. 3.0T GEM Flex Coil 16-M Array, NC033000, GEHC part number 5430008-3. 3.0T GEM Flex Coil 16-S Array, NC034200, GEHC part number5430008-4. The 3.0T GEM Flex Coil (interface box and arrays) is used in conjunction with a 3.0T GE Magnetic Resonance Scanner
Enforcement
- Recall Number
- Z-1027-2016
- Event ID
- 72976
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- NeoCoil, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 9, 2016
- Initiation Date
- December 18, 2015
- Classification Date
- February 29, 2016
- Termination Date
- October 12, 2016
- Address
- N27W23910 Paul Rd Ste A, N/A, Pewaukee, WI, 53072-6204, United States
Description
NeoCoil 3.0T GEM Flex Coil. The 3.0T GEM Flex coil system includes three sizes of coil arrays and one interface box. 3.0T GEM Flex, Interface 16ch Fixed, P -Connector, NC032204, GEHC part number 5430008-5. 3.0T GEM Flex Coil16-L Array, NC040200, GEHC part number 5430008-2. 3.0T GEM Flex Coil 16-M Array, NC033000, GEHC part number 5430008-3. 3.0T GEM Flex Coil 16-S Array, NC034200, GEHC part number5430008-4. The 3.0T GEM Flex Coil (interface box and arrays) is used in conjunction with a 3.0T GE Magnetic Resonance Scanner
Potential for higher than specified surface temperatures. Use of 3.0T GEM Flex system may result in patient warming or burns.
Mfg Lot or Serial # System ID 0000034-4-0130 608263MR5W 0000040-4-0135 608263MR5W 000032E-4-0151 608263MR5W 033-4-0151 608263MR5W 0000040-4-0202 602863MR 0000033-4-0204 602863MR 034-4-0265 602863MR 040-4-0076 602863MR 0000034-4-0250 703478MR2DV 0000040-4-0270 703478MR2DV 033-4-022 703478MR2DV 0000033-4-0272 703478MR2DV 034-4-0255 506648MR4 040-4-0275 506648MR4 040-4-0020 506648MR4 033-4-0277 506648MR4 034-4-0071 608265MR750W 0000040-4-0381 608265MR750W 0000033-4-0384 608265MR750W 034-4-0345 608265MR750W 0000034-4-0093 708346MR3 0000040-4-0096 708346MR3 0000033-4-0088 708346MR3 034-4-0352 269DVMR 040-4-0392 269DVMR 033-4-0395 269DVMR 0000034-4-0144 808873MR750W 0000033-4-0151 808873MR750W 0000040-4-0150 808873MR750W 0000034-4-0112 270798MR2 0000040-4-0118 270798MR2 0000033-4-0118 270798MR2 034-4-0089 650723NCMR1 033-4-0258 650723NCMR1 033-4-0134 650723NCMR1 0000034-4-0086 281274MR2 0000040-4-0089 281274MR2 0000033-4-0090 281274MR2 0000034-4-0076 757827MR2 0000033-4-0081 757827MR2 0000SGH225PYJJ 757827MR2 0000034-4-0007 305461MR1 0000040-4-0070 305461MR1 0000033-4-0070 305461MR1 0000034-4-0110 847872ZMR2 0000040-4-0114 847872ZMR2 0000033-4-0114 847872ZMR2 0000034-4-0095 330344MR750W 0000040-4-0098 330344MR750W 0000033-4-0099 330344MR750W 0000034-4-0094 616774MR750W 0000040-4-0097 616774MR750W 0000033-4-0098 616774MR750W 0000040-4-0276 386FNSBMR3T 0000033-4-0279 386FNSBMR3T 0000034-4-0260 386FNSBMR3T 0000034-4-0140 217876DMH3T 0000040-4-0145 217876DMH3T 0000033-4-0146 217876DMH3T 0000034-4-0216 410328MR3T 0000040-0-0224 410328MR3T 0000033-4-0227 410328MR3T 034-4-0313 716740MR1 040-4-0329 716740MR1 033-4-0329 716740MR1 0000034-4-0177 415CBMRP 0000040-4-0182 415CBMRP 0000033-4-0183 415CBMRP 0000034-4-0043 808485MR2 0000040-4-0047 808485MR2 033-4-0323 808485MR2 034-4-0267 415UCBHMR1 040-4-0284 415UCBHMR1 033-4-0285 415UCBHMR1 033-4-0352 843815750W 040-4-0055 843815750W 0000034-4-0221 843815750W 034-4-0272 415UCBHMR2 040-4-0288 415UCBHMR2 033-4-0290 415UCBHMR2 0000034-4-0108 856581MR3T 4040109 856581MR3T 0000033-4-0112 856581MR3T 034-4-0380 423493MR750W 040-4-0432 423493MR750W 033-4-0436 423493MR750W 0000034-4-0099 RADNETMR1648 0000040-4-0102 RADNETMR1648 0000033-4-0102 RADNETMR1648 034-4-0369 480342MR3 044-4-0403 480342MR3 033-4-0405 480342MR3 3440372 617726MR1A2 4040420 617726MR1A2 3340423 617726MR1A2 034-4-0069 503251PMR3 040-4-0322 503251PMR3 033-4-0144 503251PMR3 3440200 702382MR750W 4040192 702382MR750W 3340199 702382MR750W 0000-34-4-0235 207MMC3TMR 0000040-4-0249 207MMC3TMR 0000033-4-0252 207MMC3TMR 0000040-4-0291 703573MR1DV 0000033-4-0292 703573MR1DV 0000034-4-0275 703573MR1DV 0000034-4-0189 513636MR4 0000040-4-0196 513636MR4 0000033-4-0198 513636MR4 0000034-4-0188 708346MR4 0000040-4-0195 708346MR4 0000033-4-0197 708346MR4 0000034-4-0055 559734VIMR1 0000040-4-0056 559734VIMR1 0000033-4-0057 559734VIMR1 034-4-0384 720848MR4 040-4-0425 720848MR4 033-4-0428 720848MR4 0000034-4-0143 214645UT3TMR1 0000034-4-0149 214645UT3TMR1 0000034-4-0150 214645UT3TMR1 0000034-4-0115 757983MR 0000033-4-0120 757983MR 0000034-4-0015 757983MR 0000034-4-0037 604875MR750 0000040-4-0033 604875MR750 0000033-4-0044 604875MR750 0000034-4-0033 808545MR 0000040-4-0068 808545MR 0000033-4-0012 808545MR 0000034-4-0282 608262PETMR1 0000040-4-0299 608262PETMR1 0000033-4-0300 608262PETMR1 0000034-4-0138 808YMMR 0000040-4-0148 808YMMR 033-4-0081 808YMMR 034-4-0365 215829MR750W 040-4-0416 215829MR750W 033-4-0420 215829MR750W 034-4-0361 847316MR3T 040-4-0412 847316MR3T 033-4-0414 847316MR3T 0000034-4-0245 914750W 0000040-4-0264 914750W 0000033-4-0267 914750W 0000034-4-0147 856247MR1 0000040-4-0155 856247MR1 0000033-4-0155 856247MR1 0000034-4-0199 937641MR2 0000040-4-0204 937641MR2 0000033-4-0207 937641MR2 034-4-0379 859334MR 040-4-0448 859334MR 033-4-0451 859334MR 0000034-4-0166 COR526967 0000040-4-0179 COR526967 0000033-4-0191 COR526967 0000034-4-0058 RADNETMR1631 0000040-4-0059 RADNETMR1631 0000033-4-0059 RADNETMR1631 034-4-0405 CHOPPETMR 040-4-0467 CHOPPETMR 033-4-4070 CHOPPETMR 034-4-0338 608890MR750 040-4-0367 608890MR750 033-4-0370 608890MR750 034-4-0296 E416249MR2 040-4-0315 E416249MR2 033-4-0314 E416249MR2 0000034-4-0180 608CHILDMR1 0000040-4-0184 608CHILDMR1 0000033-4-0185 608CHILDMR1 0000034-4-0009 RADNETMR1632 0000040-4-0016 RADNETMR1632 0000033-4-0018 RADNETMR1632 0000034-4-0264 6142343T 0000040-4-0281 6142343T 0000033-4-0282 6142343T 034-4-0284 614293MR9 040-4-0304 614293MR9 033-4-0302 614293MR9 034-4-0394 650PH3T 040-4-0451 650PH3T 033-4-0454 650PH3T 0000033-4-0253 843750WGEM 00000040-40250 843750WGEM 033-4-0410 212288MR4 040-4-0407 212288MR4 0000033-4-0040 608WIMRMR2 0000040-4-0038 608WIMRMR2 034-4-0269 415UCBHMR3 040-4-0286 415UCBHMR3 0000033-4-0124 414647MR1 0000040-4-0116 414647MR1 034-4-0150 301570MR 0000034-4-0150 301570MR 033-4-0417 608HERIMR 040-4-0413 608HERIMR 033-3-0484 303492MR 040-4-0482 303492MR 00000040-4-289 865984MR750 0000033-4-0289 865984MR750 0000033-4-0180 713798MR1 0000040-4-0180 713798MR1 033-4-0382 970624MR1 040-4-0380 970624MR1 0000033-4-0224 715847MR3T 0000040-4-0216 715847MR3T 033-4-0430 205592750 040-4-0427 205592750 0000033-4-0268 309655MR750W 0000040-4-0265 309655MR750W 0000033-4-0159 337289MR1 0000040-4-0160 337289MR1 033-4-0372 718204MR750W 040-4-0369 718204MR750W 0000040-4-0274 702731MR750W 0000033-4-0276 702731MR750W 0000034-4-0142 312355DVMR 0000033-4-0140 312355DVMR 033-4-0446 860446MR750W 040-4-0443 860446MR750W 033-4-0391 734845VA750MR 040-4-0390 734845VA750MR 033-4-0361 904483MR3T 040-4-0358 904483MR3T 0000033-4-0075 314577CGMR3T 0000040-4-0077 314577CGMR3T 0000033-4-0246 949760MR3 0000040-4-0243 949760MR3 040-4-0039 317865MR4 000032E-4-0302 317865MR4 0000034-4-0160 BRIGHAM750WMR1 3440160 BRIGHAM750WMR1 033-4-0444 773665SJMR2 040-4-0440 773665SJMR2 0000040-4-0172 212543MR750 0000033-4-0173 212543MR750 0000033-4-0288 317887MR3 0000040-4-0287 317887MR3 0000033-4-0121 671FHPMR750W 0000040-4-0122 671FHPMR750W 033-4-0360 318443MR2 040-4-0357 318443MR2 033-4-0398 608VAMR750W 040-4-0397 608VAMR750W 034-4-0294 MR229049 0000040-4-0066 206292SR750W 0000034-4-0080 RADNETMR1639 4040197 201445750W 0000033-4-0287 415UCBHMR3 0000034-4-0198 913322MISHMRW 0000034-4-0070 6033343TW 0000034-4-0049 405271OUCMR 0000034-4-0109 RADNETMR1652 034-4-0274 412914MR 0000034-4-0116 202966MR3TW 0000040-4-0104 703698MR4 0000034-4-0269 415UCBHMR3 0000034-4-0233 954355MR750W 0000034-4-0262 PRESBYMR1 0000040-4-0075 856247MR3 040-4-0169 864529MR 0000034-4-0100 BOSTONMEDMR1 040-4-0198 301295MR3T2 0000034-4-0124 VAWESTROX750W 3340162 RADNETMR613 0000034-4-0196 217528MR1 040-4-0349 817433HSWMR2 0000034-4-0156 858622HDMR1 0000040-4-0076 0920279031 0000033-4-0032 0920279031 0000034-4-0040 0920279031 0000040-4-0022 0920279031 0000040-4-0248 EM0253 0000033-4-0251 EM0253 0000034-4-0232 EM0253 0000034-4-0290 EM0253 0000040-4-0185 EM0271 0000033-4-0162 EM0271 0000033-4-0186 EM0271 0000034-4-0181 EM0271 0000040-4-0101 EM0223 0000033-4-0101 EM0223 0000034-4-0098 EM0223 0000040-4-0055 M2322246 0000033-4-0016 M2322246 0000034-4-0054 M2322246 0000040-4-0094 0850270411 0000033-4-0095 0850270411 0000034-4-0091 0850270411 0000040-4-0159 0850270430 0000033-4-0158 0850270430 0000034-4-0151 0850270430 0000040-4-0079 PC9595MR01 0000033-4-0079 PC9595MR01 0000034-4-0074 PC9595MR01 0000040-4-0194 0850270436 0000033-4-0196 0850270436 0000034-4-0187 0850270436 0000040-4-0158 EM0234 0000033-4-0176 EM0234 0000034-4-0170 EM0234 0000040-4-0272 0850270455 0000033-4-0274 0850270455 0000034-4-0256 0850270455 0000040-4-0019 HU1098MR02 0000033-4-0020 HU1098MR02 0000034-4-0020 HU1098MR02 0000040-4-0472 0850270478 0000033-4-0474 0850270478 0000034-4-0407 0850270478 0000040-4-0146 MRUA0184 0000033-4-0145 MRUA0184 0000034-4-0139 MRUA0184 0000040-4-0045 0910274040 0000033-4-0046 0910274040 0000034-4-0041 0910274040 0000040-4-0069 UDN0121 0000033-4-0069 UDN0121 0000034-4-0066 UDN0121 0000040-4-0254 0920279028 0000033-4-0257 0920279028 0000034-4-0238 0920279028 0000040-4-0111 EM0230 0000033-4-0111 EM0230 0000034-4-0107 EM0230 0000040-4-0127 0850270409 0000033-4-0127 0850270409 0000034-4-0120 0850270409 0000040-4-0210 EM0239 0000033-4-0214 EM0239 0000034-4-0201 EM0239 0000040-4-0228 082427040146 0000033-4-0232 082427040146 0000034-4-0219 082427040146 0000040-4-0295 EM0258 0000033-4-0297 EM0258 0000034-4-0276 EM0258 0000040-4-0273 082427060052 0000033-4-0275 082427060052 0000034-4-0254 082427060052 0000040-4-0375 KW1100MR02 0000033-4-0378 KW1100MR02 0000034-4-0341 KW1100MR02 0000040-4-0285 082427100219 0000033-4-0286 082427100219 0000034-4-0270 082427100219 0000040-4-0449 M4035023 0000033-4-0452 M4035023 0000034-4-0391 M4035023 0000040-4-0290 082427110051 0000033-4-0291 082427110051 0000034-4-0273 082427110051 0000040-4-0419 MRUA0459 0000033-4-0422 MRUA0459 0000034-4-0370 MRUA0459 0000040-4-0163 082427120110 0000033-4-0165 082427120110 0000034-4-0159 082427120110 0000040-4-0181 QA1042MR01 0000033-4-0182 QA1042MR01 0000034-4-0176 QA1042MR01 0000040-4-0246 082427120116 0000033-4-0249 082427120116 0000034-4-0231 082427120116 0000040-4-0460 ZC0002 0000033-4-0462 ZC0002 0000034-4-0399 ZC0002 0000040-4-0269 082427120124 0000033-4-0273 082427120124 0000034-4-0252 082427120124 0000040-4-0147 EM0227 0000033-4-0147 EM0227 0000034-4-0141 EM0227 0000040-4-0309 082427120131 0000033-4-0306 082427120131 0000034-4-0288 082427120131 0000040-4-0174 EM0232 0000033-4-0175 EM0232 0000034-4-0168 EM0232 0000040-0-0303 082427130076 0000033-4-0309 082427130076 0000034-4-0291 082427130076 0000040-4-0170 EM0236 0000033-4-0171 EM0236 0000034-4-0164 EM0236 0000040-4-0156 83092403928 0000033-4-0156 83092403928 0000034-4-0148 83092403928 0000040-4-0221 EM0242 0000033-4-0222 EM0242 0000034-4-0202 EM0242 0000040-4-0165 83096535911 0000033-4-0187 83096535911 0000034-4-0178 83096535911 0000040-4-0231 EM0256 0000033-4-0234 EM0256 0000034-4-0220 EM0256 0000040-4-0279 EM0265 0000033-4-0280 EM0265 0000034-4-0266 EM0265 0000040-4-0082 0850270414 0000033-4-0082 0850270414 0000034-4-0077 0850270414 0000040-4-0282 EM0275 0000033-4-0283 EM0275 0000034-4-0265 EM0275 0000040-4-0211 83096536248 0000033-4-0215 83096536248 0000034-4-0203 83096536248 0000040-4-0405 EM0281 0000033-4-0407 EM0281 0000034-4-0377 EM0281 0000040-4-0260 83104653897 0000033-4-0263 83104653897 0000034-4-0246 83104653897 0000040-4-0346 KA1023MR02 0000033-4-0345 KA1023MR02 0000034-4-0319 KA1023MR02 0000040-4-0241 83105089955 0000033-4-0244 83105089955 0000034-4-0326 83105089955 0000040-4-0027 LB1331MR01 0000033-4-0030 LB1331MR01 0000034-4-0030 LB1331MR01 0000040-4-0262 83108956002 0000033-4-0265 83108956002 0000034-4-0243 83108956002 0000040-4-0298 M4018549 0000033-4-0299 M4018549 0000034-4-0281 M4018549 0000040-4-0311 83112923777 0000033-4-0310 83112923777 0000034-4-0292 83112923777 0000040-4-0083 MPX10000 0000033-4-0083 MPX10000 0000034-4-0078 MPX10000 000040-04-0429 85110163303 0000033-4-0432 85110163303 0000034-4-0371 85110163303 0000040-4-0310 MRUA0343 0000033-4-0307 MRUA0343 0000034-4-0268 MRUA0343 0000040-4-0453 083027020204715 0000033-4-0427 083027020204715 0000033-4-0456 083027020204715 0000034-4-0378 083027020204715 0000040-4-0424 083027020204715 0000040-4-0136 O009MR03 0000033-4-0136 O009MR03 0000034-4-0128 O009MR03 0000040-4-0187 083027200034313 0000033-4-0189 083027200034313 0000034-4-0184 083027200034313 0000040-4-0297 PL3277MR01 0000033-4-0296 PL3277MR01 0000034-4-0289 PL3277MR01 0000040-4-0452 06008MRS03 0000033-4-0455 06008MRS03 0000034-4-0395 06008MRS03 0000040-4-0188 T002MR01 0000033-4-+190 T002MR01 0000034-4-0183 T002MR01 0000040-4-0026 120810MR02 0000033-4-0028 120810MR02 0000034-4-0028 120810MR02 0000040-4-0383 X41944010 0000033-4-0385 X41944010 0000034-4-0346 X41944010 0000040-4-0166 120810MR03 0000033-4-0166 120810MR03 0000034-4-0161 120810MR03 0000040-4-0133 EM0221 0000033-4-0132 EM0221 0000034-4-0126 EM0221 0000040-4-0065 140024MR01 0000033-4-0066 140024MR01 0000034-4-0065 140024MR01 0000040-4-0144 EM0225 0000033-4-0144 EM0225 0000034-4-0137 EM0225 0000040-4-0378 150250MR01 0000033-4-0399 150250MR01 0000034-4-0343 150250MR01 0000040-4-0164 EM0229 0000033-4-0168 EM0229 0000034-4-0162 EM0229 0000040-4-0498 157022MR01 0000033-4-0501 157022MR01 0000034-4-0425 157022MR01 0000040-4-0162 EM0231 0000033-4-0164 EM0231 0000034-4-0158 EM0231 0000040-4-0408 20312MRS01 0000033-4-0410 20312MRS01 0000034-4-0368 20312MRS01 0000040-4-0168 EM0233 0000033-4-0169 EM0233 0000034-4-0163 EM0233 0000040-4-0219 220078MR01 0000033-4-0221 220078MR01 0000034-4-0208 220078MR01 0000040-4-0171 EM0235 0000033-4-0172 EM0235 0000034-4-0165 EM0235 0000040-4-0339 220826MR02 0000033-4-0339 220826MR02 0000034-4-0308 220826MR02 0000040-4-0275 EM0237 0000033-4-0194 EM0237 0000034-4-0185 EM0237 0000040-4-0099 26001MRS01 0000033-4-0097 26001MRS01 0000034-4-0096 26001MRS01 0000040-4-0186 EM0240 0000033-4-0188 EM0240 0000034-4-0182 EM0240 0000040-4-0464 34026MRS05 0000033-4-0466 34026MRS05 0000034-4-0403 34026MRS05 0000040-4-0238 EM0243 0000033-4-0241 EM0243 0000034-4-0222 EM0243 0000040-4-0078 34348MRS01 0000033-4-0076 34348MRS01 0000034-4-0072 34348MRS01 0000040-4-0234 EM0255 0000033-4-0237 EM0255 0000034-4-0224 EM0255 0000040-4-0134 387040MR10 0000033-4-0134 387040MR10 0000034-4-0127 387040MR10 0000040-4-0256 EM0257 0000033-4-0259 EM0257 0000034-4-0239 EM0257 0000040-4-0153 600042MR02 0000033-4-0154 600042MR02 0000034-4-0146 600042MR02 0000040-4-0230 EM0260 0000033-4-0233 EM0260 0000034-4-0218 EM0260 0000040-4-0018 70431MRS04 0000033-4-0026 70431MRS04 0000034-4-0026 70431MRS04 0000040-4-0393 EM0267 0000033-4-0394 EM0267 0000034-4-0279 EM0267 0000040-4-0263 725100MR02 0000033-4-0266 725100MR02 0000034-4-0244 725100MR02 0000040-4-0445 EM0274 0000033-4-0448 EM0274 0000034-4-0390 EM0274 0000040-4-0257 77266MRS03 0000033-4-0260 77266MRS03 0000034-4-0241 77266MRS03 0000040-4-0428 EM0280 0000033-4-0431 EM0280 0000034-4-0381 EM0280 0000040-4-0014 850020MR02 0000033-4-0017 850020MR02 0000034-4-0017 850020MR02 0000040-4-0463 EX0130 0000033-4-0465 EX0130 0000034-4-0402 EX0130 0000040-4-0414 A4493006 0000033-4-0418 A4493006 0000034-4-0363 A4493006 0000040-4-0178 IE1010MR01 0000033-4-0179 IE1010MR01 0000034-4-0172 IE1010MR01 0000040-4-0307 AC001MR02 0000033-4-0308 AC001MR02 0000034-4-0286 AC001MR02 0000040-4-0244 KW1019MR03 0000033-4-0247 KW1019MR03 0000034-4-0229 KW1019MR03 0000040-4-0400 AP7645MR01 0000033-4-0402 AP7645MR01 0000034-4-0353 AP7645MR01 0000040-4-0139 LB1160MR02 0000033-4-0139 LB1160MR02 0000034-4-0132 LB1160MR02 0000040-4-0371 B5129925 0000033-4-0374 B5129925 0000034-4-0339 B5129925 0000040-4-0128 M004MR01 0000033-4-0128 M004MR01 0000034-4-0119 M004MR01 0000040-4-0266 B5174327 0000033-4-0269 B5174327 0000034-4-0247 B5174327 0000040-4-0388 M345477701 0000033-4-0390 M345477701 0000034-4-0350 M345477701 0000040-4-0215 BH1017MR01 0000033-4-0226 BH1017MR01 0000034-4-0215 BH1017MR01 0000040-4-0411 M40345211 0000033-4-0412 M40345211 0000034-4-0360 M40345211 0000040-4-0251 C001MR01 0000033-4-0254 C001MR01 0000034-4-0237 C001MR01 0000040-4-0305 M54606125 0000033-4-0303 M54606125 0000034-4-0285 M54606125 0000040-4-0074 C4200438 0000033-4-0073 C4200438 0000034-4-0069 C4200438 0000040-4-0161 MRUA0122 0000033-4-0161 MRUA0122 0000034-4-0152 MRUA0122 4040058 DPN26332 3340060 DPN26332 3440059 DPN26332 0000040-4-0258 MRUA0191 0000033-4-0261 MRUA0191 0000034-4-0242 MRUA0191 0000040-4-0423 DPN36500 0000033-4-0426 DPN36500 0000034-4-0366 DPN36500 0000040-4-0308 MRUA0353 0000033-4-0305 MRUA0353 0000034-4-0287 MRUA0353 0000040-4-0034 E001MR02 0000033-4-0062 E001MR02 0000034-4-0062 E001MR02 0000040-4-0293 N002MR03 0000033-4-0293 N002MR03 0000034-4-0277 N002MR03 0000040-4-0213 EM0025 0000033-4-0217 EM0025 0000034-4-0205 EM0025 0000040-4-0138 PC3003MR01 0000033-4-0138 PC3003MR01 0000034-4-0131 PC3003MR01 0000040-4-0061 EM0201 0000033-4-0011 EM0201 0000034-4-0061 EM0201 0000040-4-0176 PL3116MR01 0000033-4-0181 PL3116MR01 0000034-4-0175 PL3116MR01 0000040-4-0085 EM0207 0000033-4-0085 EM0207 0000034-4-0079 EM0207 0000040-4-0012 PNUE08MR01 0000033-4-0025 PNUE08MR01 0000034-4-0006 PNUE08MR01 0000040-4-0106 EM0209 0000033-4-0106 EM0209 0000034-4-0102 EM0209 0000040-4-0267 SA1255MR03 0000033-4-0270 SA1255MR03 0000034-4-0248 SA1255MR03 0000040-4-0100 EM0212 0000033-4-0100 EM0212 0000034-4-0097 EM0212 0000040-4-0455 TP0008MR07 0000033-4-0467 TP0008MR07 0000034-4-0404 TP0008MR07 0000040-4-0108 EM0214 0000033-4-0108 EM0214 0000034-4-0104 EM0214 0000040-4-0296 UE929526 0000033-4-0298 UE929526 0000034-4-0280 UE929526 0000040-4-0140 EM0218 0000033-4-0092 EM0218 0000034-4-0087 EM0218 0000040-4-0479 ZC0001 0000033-4-0481 ZC0001 0000034-4-0412 ZC0001 0000040-4-0120 EM0219 0000033-4-0119 EM0219 0000034-4-0114 EM0219 0000040-4-0115 EM0220 0000033-4-0116 EM0220 0000034-4-0111 EM0220 0000040-4-0151 EM0228 0000033-4-0152 EM0228 0000034-4-0145 EM0228 0000040-4-0355 RU8937MR01 0000033-4-0356 RU8937MR01 0000040-4-0050 N4192951 0000033-4-0056 N4192951 0000040-4-0199 MR308183MR3 0000033-4-0201 MR308183MR3 0000040-4-0236 83104203907 0000033-4-0239 83104203907 0000040-4-0456 BG4680MR01 0000033-4-0458 BG4680MR01 0000040-4-0401 EM0264 0000033-4-0403 EM0264 0000040-4-0253 EM0249 0000033-4-0256 EM0249 0000040-4-0377 EM0266 0000033-4-0380 EM0266 0000034-4-0136 082427040135 0000033-4-0143 082427040135 0000040-4-0409 EM0268 0000033-4-0411 EM0268 0000040-4-0442 082427160149 0000033-4-0441 082427160149 0000040-4-0084 0910273043 0000033-4-0084 0910273043 0000040-4-0232 RU8615MR01 0000033-4-0235 RU8615MR01 0000040-4-0441 EM0276 0000033-4-0445 EM0276 0000040-4-0214 EM0247 0000033-4-0218 EM0247 0000040-4-0422 EM0277 0000033-4-0425 EM0277 0000034-4-0023 X59007321 48 X59007321 0000040-4-0117 EM0213 0000033-4-0117 EM0213 0000040-4-0450 A1765918 0000033-4-0453 A1765918 0000040-4-0049 EM0215 0000033-4-0110 EM0215 0000040-4-0382 A5719869 0000033-4-0383 A5719869 0000040-4-0124 EM0217 0000033-4-0123 EM0217 0000040-4-0207 EM0238 0000033-4-0210 EM0238 0000040-4-0062 541010MR04 0000033-4-0063 541010MR04 0000040-4-0220 130083MR05 0000034-4-0209 130083MR05 0000040-4-0152 RU5599MR01 0000033-4-0153 RU5599MR01 0000040-4-0119 M001MR02 0000034-4-0113 M001MR02 0000040-4-0306 RU8792MR01 0000033-4-0304 RU8792MR01 0000040-4-0459 M2850013 0000033-4-0461 M2850013 0000040-4-0154 SA1223MR01 0000034-4-0154 SA1223MR01 0000040-4-0259 0850270446 0000033-4-0262 0850270446 0000040-4-0240 EM0248 0000033-4-0243 EM0248 0000040-4-0126 EM0226 0000033-4-0126 EM0226 0000040-4-0252 EM0254 0000033-4-0255 EM0254 0000040-4-0385 M4172341 0000033-4-0386 M4172341 0000040-4-0053 M45052106 0000034-4-0052 M45052106 0000040-4-0439 MRUA0394A 0000033-4-0443 MRUA0394A 00040-4-0008-2 082427040135AW2 0000033-4-0315 A5547422 0000034-4-0149 85092252519 0000034-4-0191 EM0005 0000034-4-0258 0853270141 0000034-4-0053 0850270386 0000040-4-0294 6073MR0003A 0000034-4-0213 EM0244 0000034-4-0212 EM0245 0000034-4-0206 EM0246 0000034-4-0106 0853270119 0000034-4-0122 EM0216 00000034-4-230 083027860052014 0000004-4-0242 MRUA0272 0000033-4-0037 83027840038612 0000033-4-0080 AE1016MR01 0000033-4-0141 0850270427 0000033-4-0160 SA1223MR01 0000033-4-0184 KW1005MR04 0000033-4-0192 RU8438MR01 0000033-4-0193 RU8438MR03 0000033-4-0213 130083MR05 0000033-4-0225 10692MRS01 0000033-4-0240 00230MRS01 0000033-4-0248 083027860052014 0000033-4-0278 A5111746 0000033-4-0379 083027806724415 0000033-4-0387 53008MRS01 0000033-4-0388 06162MRS01 0000033-4-0389 EG1526MR05 0000033-4-0396 083027046705715 0000033-4-0397 20312MRS01 0000033-4-0408 DZ1060MR04 0000033-4-0435 082427200086 0000033-4-0439 0850270459 0000033-4-0460 210002MR01 0000033-4-0475 082427100254 0000033-4-0478 MRWB0843 0000033-4-0488 082427010087 0000033-4-0491 00440MRS08 0000033-4-0497 10722MRS01 0000034-4-0398 M2850013 0000040-4-0041 083027840038612 00000032E40288 83096322711 0000032-E-0389 082427130076 000022E-4-0544 M007MR01 000032E-4-0015 PNUE08MR01 000032E-4-0021 LB1331MR01 000032E-4-0027 HU1098MR02 000032E-4-0031 X59007321 000032E-4-0037 70431MRS04 000032E-4-0039 120810MR02 000032E-4-0048 083027840038612 000032E-4-0050 0920279031 000032E-4-0051 0910274040 000032E-4-0056 M2322246 000032E-4-0058 N4192951 000032E-4-0060 0850270386 000032E-4-0068 EM0201 000032E-4-0069 850020MR01 000032E-4-0071 541010MR04 000032E-4-0073 140024MR01 000032E-4-0075 UDN0121 000032E-4-0079 C4200438 000032E-4-0081 0850270411 000032E-4-0085 EM0212 000032E-4-0086 34348MRS01 000032E-4-0088 0850270414 000032E-4-0090 PC9595MR01 000032E-4-0092 MPX10000 000032E-4-0093 M001MR02 000032E-4-0097 0910273043 000032E-4-0098 EM0207 000032E-4-0105 EM0218 000032E-4-0107 EM0223 000032E-4-0109 26001MRS01 000032E-4-0113 EM0209 000032E-4-0119 EM0214 000032E-4-0120 EM0215 000032E-4-0122 0853270119 000032E-4-0123 EM0230 000032E-4-0128 EM0220 000032E-4-0129 EM0213 000032E-4-0132 EM0219 000032E-4-0136 EM0226 000032E-4-0138 EM0217 000032E-4-0141 0850270409 000032E-4-0142 M004MR01 000032E-4-0144 EM0216 000032E-4-0148 EM0221 000032E-4-0149 387040MR10 000032E-4-0150 O009MR03 000032E-4-0152 PC3003MR01 000032E-4-0153 LB1160MR02 000032E-4-0158 082427040135 000032E-4-0159 EM0225 000032E-4-0160 MRUA0184 000032E-4-0162 EM0227 000032E-4-0168 EM0228 000032E-4-0169 RU5599MR01 000032E-4-0174 600042MR02 000032E-4-0175 83092403928 000032E-4-0176 85092252519 000032E-4-0179 0850270430 000032E-4-0181 MRUA0122 000032E-4-0183 SA1223MR01 000032E-4-0190 EM0231 000032E-4-0191 082427120110 000032E-4-0194 120810MR03 000032E-4-0196 EM0229 000032E-4-0197 EM0233 000032E-4-0198 EM0236 000032E-4-0199 EM0235 000032E-4-0202 EM0232 000032E-4-0203 EM0234 000032E-4-0206 IE1010MR01 000032E-4-0208 PL3116MR01 000032E-4-0209 QA1042MR01 000032E-4-0212 T002MR01 000032E-4-0214 EM0271 000032E-4-0215 83096535911 000032E-4-0216 EM0240 000032E-4-0217 083027200034313 000032E-4-0224 EM0237 000032E-4-0226 0850270436 000032E-4-0228 EM0005 000032E-4-0233 MR308183MR3 000032E-4-0244 EM0238 000032E-4-0247 EM0239 000032E-4-0249 83096536248 000032E-4-0251 EM0025 000032E-4-0253 EM0246 000032E-4-0256 220078MR01 000032E-4-0257 130083MR05 000032E-4-0258 EM0242 000032E-4-0261 BH1017MR01 000032E-4-0264 EM0245 000032E-4-0265 EM0244 000032E-4-0272 082427040146 000032E-4-0273 EM0260 000032E-4-0275 RU8615MR01 000032E-4-0277 EM0256 000032E-4-0281 83104203907 000032E-4-0285 EM0255 000032E-4-0286 EM0243 000032E-4-0287 EM0248 000032E-4-0289 EM0247 000032E-4-0293 83105089955 000032E-4-0294 KW1019MR03 000032E-4-0296 082427120116 000032E-4-0298 EM0253 000032E-4-0305 C001MR01 000032E-4-0306 EM0249 000032E-4-0307 EM0254 000032E-4-0311 EM0257 000032E-4-0313 MRUA0191 000032E-4-0314 0850270446 000032E-4-0315 83108956002 000032E-4-0318 725100MR02 000032E-4-0320 83104653897 000032E-4-0322 B5174327 000032E-4-0323 SA1255MR03 000032E-4-0328 082427120124 000032E-4-0332 082427060052 000032E-4-0334 0850270455 000032E-4-0337 0853270141 000032E-4-0341 E001MR02 000032E-4-0349 EM0265 000032E-4-0352 EM0275 000032E-4-0357 082427100219 000032E-4-0364 082427110051 000032E-4-0368 N002MR03 000032E-4-0371 EM0258 000032E-4-0373 UE929526 000032E-4-0376 M4018549 000032E-4-0378 6073MR0007 000032E-4-0381 M54606125 000032E-4-0382 RU8792MR01 000032E-4-0383 AC001MR02 000032E-4-0384 MRUA0353 000032E-4-0385 PL3277MR01 000032E-4-0386 082427120131 000032E-4-0387 MRUA0343 000032E-4-0390 83112923777 000032E-4-0403 EM0262 000032E-4-0406 M365865301 000032E-4-0407 082427180028 000032E-4-0418 EM0259 000032E-4-0421 EM0261 000032E-4-0422 EM0270 000032E-4-0425 220826MR02 000032E-4-0426 KA1023MR02 000032E-4-0427 082427040167 000032E-4-0428 83112923995 000032E-4-0430 082427100225 000032E-4-0440 EM0263 000032E-4-0441 083027244278914 000032E-4-0443 0920279028 000032E-4-0446 RU8937MR01 000032E-4-0455 082427120122 000032E-4-0457 082427120135 000032E-4-0460 082427010080 000032E-4-0467 B5129925 000032E-4-0471 KW1100MR02 000032E-4-0474 EM0266 000032E-4-0479 A5719869 000032E-4-0480 X41944010 000032E-4-0481 M4172341 000032E-4-0485 EG1526MR05 000032E-4-0486 M345477701 000032E-4-0492 EM0267 000032E-4-0494 259130MR12 000032E-4-0496 150250MR01 000032E-4-0499 AP7645MR01 000032E-4-0502 EM0264 000032E-4-0511 EM0268 000032E-4-0513 M40345211 000032E-4-0515 20312MRS01 000032E-4-0518 A4493006 000032E-4-0525 MRUA0459 000032E-4-0528 EM0003 000032E-4-0530 EM0277 000032E-4-0531 DPN36500 000032E-4-0532 EM0281 000032E-4-0539 85110163303 000032E-4-0545 EM0280 000032E-4-0554 MRUA0394A 000032E-4-0556 EM0276 000032E-4-0561 082427160149 000032E-4-0563 M4035023 000032E-4-0567 EM0274 000032E-4-0569 EM0273 000032E-4-0572 A1765918 000032E-4-0577 083027020204715 000032E-4-0579 BG4693MR01 000032E-4-0580 EM0269 000032E-4-0584 M2850013 000032E-4-0585 ZC0002 000032E-4-0588 EX0130 000032E-4-0589 34026MRS05 000032E-4-0592 TP0008MR07 000032E-4-0595 EX0131 000032E-4-0598 0850270478 000032E-4-0610 ZC0001 000032E-4-0634 34225MRS02 000032E-4-0637 157022MR01 000032E-4-0014 140024MR01 000032E-4-0010 EM0223 000032E-4-0220 EM0237 000032E-4-0271 EM0260 000032E-4-0274 EM0256 000032E-4-0280 EM0255 000032E-4-0283 EM0243 000032E-4-0252 EM0247 000032E-4-0290 83105089955 000032E-4-0309 EM0257 000032E-4-0316 83104653897 000032E-4-0329 0850270455 000032E-4-0344 EM0265 000032E-4-0348 EM0275 000032E-4-0366 N002MR03 000032E-4-0367 EM0258 000032E-4-0375 M4018549 000032E-4-0380 M54606125 000032E-4-0374 PL3277MR01 00000032E-0386 082427120131 000032E-4-0356 MRUA0343 000032E-4-0405 M365865301 000032E-4-0408 EM0259 000032E-4-0415 220826MR02 000032E-4-0462 EM0266 000032E-4-0478 X41944010 000032E-4-0450 M4172341 000032E-4-0372 EM0267 278 150250MR01 000032E-4-0495 AP7645MR01 000032E-4-0483 20312MRS01 000032E-4-0529 EM0277 000032E-4-0521 DPN36500 000032E-4-0505 EM0281 000032E-4-0526 85110163303 000032E-4-0538 EM0280 000032E-4-0560 EM0274 000032E-4-0533 083027020204715 000032E-4-0559 EM0269 000032E-4-0582 ZC0002 000032E-4-0413 EX0130 000032E-4-0590 TP0008MR07 000032E-4-0607 ZC0001 000032E-4-0633 157022MR01 000032E-4-0509 20312MRS01 032E-4-0330 506648MR4 032E-4-0397 412914MR 033-04-0319 920738MR750 032E-4-0412 716740MR1 032E-4-0379 614293MR9 032e-4-0189 RADNETMR621 032E-4-0442 248533750W 000032E-4-0438 336832MR750 032E-4-0433 206CH750MR 032E-4-0487 734845VA750MR 000032E-4-0112 BOSTONMEDMR1 000032E-4-0193 BRIGHAM750WMR1 000032E-4-0447 MSK53MR02 032E-4-0477 970624MR1 032E-4-0358 415UCBHMR3 032E-4-0354 415UCBHMR1 032E-4-0362 415UCBHMR2 032e-4-0411 503251PMR3 032E-4-0465 718204MR750W 032E-4-0392 MR229049 032E-4-0488 269DVMR 032E-4-0207 201445750W 032e-4-0359 516NS7113T 032E-4-0368 904483MR3T 000032E-4-0476 608265MR750W 032E-4-0514 847316MR3T 032E-4-0497 608VAMR750W 032E-4-0546 720848MR4 032E-4-0522 480342MR3 032E-4-0547 423493MR750W 032E-4-0452 318443MR2 032E-4-0448 RADNETMR613 032E-4-0139 510449MR750W 032E-4-0509 212288MR4 notfnd MAYOPETMR01 000032E-4-0463 608890MR750 032E-4-0461 614722MR3 032E40527 617726MR1A2 032E-4-0537 205592750 032E-4-0520 215829MR750W 032E-4-0562 860446MR750W 032E-4-0570 859334MR 032E-4-0574 650PH3T 032E-4-0417 608HERIMR 032E-4-0555 773665SJMR2 032E-4-0394 E416249MR2 032E-4-0594 CHOPPETMR 032E-4-0613 303492MR 901813 864529MR 032E-4-0593 910907WAMR1 032E-4-0644 336397P3T 032E-4-0548 310264MR2 032E-4-0617 815968MR 00000032E40229 702382MR750W 0000034-4-0234 317865MR4 000022E-4-0190 305854MR1 000032E-4-0033 815395MR2 000032E-4-0041 604875MR750 000032E-4-0043 608WIMRMR2 032e-4-0065 RADNETMR1631 000032E-4-0074 808545MR 000032E-4-0076 305461MR1 000032E-4-0084 757827MR2 000032E-4-0099 RADNETMR1639 000032E-4-0101 281274MR2 000032E-4-0108 708346MR3 032E-4-0558 616774MR750W 000032E-4-0111 330344MR750W 000032E-4-0115 RADNETMR1648 000032E-4-0117 703698MR4 000032E-4-0124 856581MR3T 000032E-4-0126 RADNETMR1652 000032E-4-0127 847872ZMR2 000032E-4-0130 270798MR2 000032E-4-0133 757983MR 000032E-4-0134 671FHPMR750W 032E-40051 414647MR1 000032E-4-0146 VAWESTROX750W 000032E-4-0161 217876DMH3T 000032E-4-0163 312355DVMR 000032E-4-0164 808YMMR 032E-4-0065 808873MR750W 000032E-4-0171 6033343TW 000032E-4-0173 856247MR1 000032E-4-0177 301570MR 000032E-4-0180 337289MR1 000032E-4-0185 858622HDMR1 000032E-4-0186 808485MR2 000032E-4-0200 212543MR750 032e-4-0165 713798MR1 000032E-4-0210 415CBMRP 000032E-4-0213 608CHILDMR1 000032E-4-0218 COR526967 000032E-4-0227 513636MR4 000032E-4-0234 217528MR1 000032E-4-0236 602863MR 000032E-4-0242 913322MISHMRW 000032E-4-0243 937641MR2 032E-4-0110 843815750W 000032E-4-0262 715847MR3T 000032E-4-0266 410328MR3T 000032E-4-0292 949760MR3 000032E-4-0300 954355MR750W 000032E-4-0301 843750WGEM 000032E-4-0303 207MMC3TMR 000032E-4-0319 914750W 000032E-4-0321 309655MR750W 000032E-4-0326 703478MR2DV 000032E-4-0333 702731MR750W 000032E-4-0343 386FNSBMR3T 000032E-4-0347 6142343T 000032E-4-0360 317887MR3 000032E-4-0361 865984MR750 000032E-4-0365 703573MR1DV 000032E-4-0377 608262PETMR1 032E-4-0336 506648MR4 032E-4-0399 920738MR750 032E-4-0456 206CH750MR 000032E-4-0401 MSK53MR02 032E-4-0400 503251PMR3 032E-4-0404 MR229049 032E-4-0491 269DVMR 032E-4-0535 720848MR4 032E-4-0503 480342MR3 032E-4-0542 423493MR750W 032E-4-0463 608890MR750 032E-4-0541 859334MR 000032E-4-0009 815395MR2 000032E-4-0239 937641MR2 000032E-4-0299 207MMC3TMR 000032E-4-0325 703478MR2DV 000032E-4-0339 386FNSBMR3T 040-4-0321 920738MR750
US Nationwide Distribution
2121